摘要
PURPOSE. To compare posterior capsule opacification (PCO) between 2 three,piece intraocular tenses (IOLs) both with sharp,optic edges and open loop haptics, one made of silicone and the other made of hydrophobic acrylate, 6 years after surgery. DESIGN: Randomized, controlled, patient- and examiner-masked trial with intra-patient comparison. METHODS: One hundred and six eyes in 53 patients were included initially. Forty,four eyes of 22 patients with age related bilateral cataract were available for the 6 years follow-up. Each patient had standard cataract surgery and randomly received a silicone (CeeOn 911A; AMO, Santa Ana, California, USA) or hydrophobic acrylic (AcrySof MA60BM; Alcon, Fort Worth, Texas, USA) IOL in the first eye, the other type of IOL in the fellow eye. Follow-up examinations were at 1, 3, and 6 years after surgery. The intensity of posterior capsule opacification (PCO) was assessed using digital retro-illuminated photos and the "automated quantification of after-cataract" (AQUA) software. Neodymium-yttrium aluminum-garnet-laser capsulotomies (YAG-LC) performed were noted. RESULTS: Six years after surgery, the AQUA scores were 2.3 +/- 1.4 for the silicone and 3.8 +/- 2.0 for the acrylic IOLs (P = .0016). The acrylic 101, eyes also had a significant increase in AQUA score from the 3-year to the 6-year follow-up. YAG-LC were performed in 2 silicone and 6 acrylic IOLs (P = .01). CONCLUSIONS: An angulated three-piece hydrophobic acrylic IOL had more PCO 6 years after surgery than a silicone IOL of otherwise similar design. Apart from an optic material effect, differences in haptic design and the degree of optic edge sharpness may play a role. (Am J Ophthalmol 2009;147:683-690. (C) 2009 by Elsevier Inc. All rights reserved.)
引用
收藏
页码:683 / 690
页数:8
相关论文
共 35 条
[1]   Uveal and capsular biocompatibility of hydrophilic acrylic, hydrophobic acrylic, and silicone intraocular lenses [J].
Abela-Formanek, C ;
Amon, M ;
Schild, G ;
Schauersberger, J ;
Heinze, G ;
Kruger, A .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (01) :50-61
[2]   Quantification of posterior capsule opacification with round and sharp edge intraocular lenses [J].
Auffarth, GU ;
Golescu, A ;
Becker, KA ;
Völcker, HE .
OPHTHALMOLOGY, 2003, 110 (04) :772-780
[3]   Reproducibility of standardized retroillumination photography for quantification of posterior capsule opacification [J].
Buehl, W ;
Findl, O ;
Menapace, R ;
Georgopoulos, M ;
Rainer, G ;
Wirtitsch, M ;
Siegl, H ;
Pinz, A .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) :265-270
[4]   Statistical problems caused by missing data resulting from neodymium:YAG laser capsulotomies in long-term posterior capsule opacification studies -: Problem identification and possible solutions [J].
Buehl, Wolf ;
Heinzl, Harald ;
Mittlboeck, Martina ;
Findl, Oliver .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (02) :268-273
[5]  
*EUROSTAT, 2006, POP STAT, P24
[6]   Interventions for preventing posterior capsule opacification [J].
Findl, O. ;
Buehl, W. ;
Bauer, P. ;
Sycha, T. .
COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2007, (03)
[7]   Effect of optic material on posterior capsule opacification in intraocular lenses with sharp-edge optics - Randomized clinical trial [J].
Findl, O ;
Menapace, R ;
Sacu, S ;
Buehl, W ;
Rainer, G .
OPHTHALMOLOGY, 2005, 112 (01) :67-72
[8]   Comparison of 4 methods for quantifying posterior capsule opacification [J].
Findl, O ;
Buehl, W ;
Menapace, R ;
Georgopoulos, M ;
Rainer, G ;
Siegl, H ;
Kaider, A ;
Pinz, A .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (01) :106-111
[9]   Influence of intraocular lens material on regeneratory posterior capsule opacification after neodymium: YAG laser capsulotomy [J].
Georgopoulos, M ;
Findl, O ;
Menapace, R ;
Buehl, W ;
Wirtitsch, M ;
Rainer, G .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (08) :1560-1565
[10]   MEASUREMENT OF ELASTIC RESISTING FORCES OF INTRAOCULAR HAPTIC LOOPS OF VARYING GEOMETRICAL DESIGNS AND MATERIAL COMPOSITION [J].
GUTHOFF, R ;
ABRAMO, F ;
DRAEGER, J ;
CHUMBLEY, L .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1990, 16 (05) :551-558