Optimisation of an MDCT abdominal protocol: Image quality assessment of standard vs. iterative reconstructions

被引:12
作者
Berta, L. [1 ]
Mascaro, L. [1 ]
Feroldi, P. [1 ]
Maroldi, R. [2 ]
机构
[1] Spedali Civili Brescia, Dept Imaging, Med Phys Unit, I-25123 Brescia, Italy
[2] Univ Brescia, Dept Radiol, I-25123 Brescia, Italy
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2014年 / 30卷 / 03期
关键词
Iterative reconstruction; Multi detector CT; CT protocol optimisation; Abdominal CT; Radiation dose reduction; CORONARY CT ANGIOGRAPHY; DOSE REDUCTION; ALGORITHM;
D O I
10.1016/j.ejmp.2013.07.126
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work aims to construct a method to objectively evaluate CT image quality when new clinical protocol performances must be compared with a standard reference. We compare iterative reconstruction in the image space with filtered back projection reconstruction and accurately quantify the dose reduction. The comparison strategy accounts for both physical and clinical image qualities that are evaluated using a standard metric. The quasi-ideal observer metric is also explored to verify its reportedly high correlation with perceived image quality. Water or spatial resolution phantom images are used to characterise the physical image quality using the classic metrics in the Fourier domain by calculating the modulation transfer functions and noise power spectra (NPS). The clinical-image quality is evaluated with a 4-alternative forced-choice test. The human observers are asked to detect a positive image that contains a simulated lesion in a background image. Then, the same positive images are characterised with the quasi-ideal observer metric, which calculates the non-prewhitening matched filter signal-to-noise ratio (SNRNPWMF). Iterative reconstruction strongly reduces the image noise, but the NPS are slightly shifted to lower frequencies, which gives the images a coarse graininess. Compared with the reference FBP protocol for abdomen exams, the highest dose reduction is 40% if the standard metric is used and 30% if the SNRNPWMF metric is used. The detectability test results achieve a better correlation with SNRNPWMF than with the standard metric. The identified Fourier metric is a useful descriptor of human quality perception and can be used for future protocol optimisation. (C) 2013 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:271 / 279
页数:9
相关论文
共 29 条
[21]   Iterative reconstruction methods in two different MDCT scanners: Physical metrics and 4-alternative forced-choice detectability experiments - A phantom approach [J].
Mieville, Frederic A. ;
Gudinchet, Francois ;
Brunelle, Francis ;
Bochud, Francois O. ;
Verdun, Francis R. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (01) :99-110
[22]   Adaptive statistical iterative reconstruction versus filtered back projection in the same patient: 64 channel liver CT image quality and patient radiation dose [J].
Mitsumori, Lee M. ;
Shuman, William P. ;
Busey, Janet M. ;
Kolokythas, Orpheus ;
Koprowicz, Kent M. .
EUROPEAN RADIOLOGY, 2012, 22 (01) :138-143
[23]   Iterative reconstruction of dual-source coronary CT angiography: assessment of image quality and radiation dose [J].
Park, Eun-Ah ;
Lee, Whal ;
Kim, Kwang Woo ;
Kim, Kwang Gi ;
Thomas, Allmendinger ;
Chung, Jin Wook ;
Park, Jae Hyung .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2012, 28 (07) :1775-1786
[24]   Chest computed tomography using iterative reconstruction vs filtered back projection (Part 1): evaluation of image noise reduction in 32 patients [J].
Pontana, Francois ;
Pagniez, Julien ;
Flohr, Thomas ;
Faivre, Jean-Baptiste ;
Duhamel, Alain ;
Remy, Jacques ;
Remy-Jardin, Martine .
EUROPEAN RADIOLOGY, 2011, 21 (03) :627-635
[25]   Chest computed tomography using iterative reconstruction vs filtered back projection (Part 2): image quality of low-dose CT examinations in 80 patients [J].
Pontana, Francois ;
Duhamel, Alain ;
Pagniez, Julien ;
Flohr, Thomas ;
Faivre, Jean-Baptiste ;
Hachulla, Anne-Lise ;
Remy, Jacques ;
Remy-Jardin, Martine .
EUROPEAN RADIOLOGY, 2011, 21 (03) :636-643
[26]  
Rasband WS, 1997, ImageJ
[27]   Iterative Reconstruction Algorithm for Abdominal Multidetector CT at Different Tube Voltages: Assessment of Diagnostic Accuracy, Image Quality, and Radiation Dose in a Phantom Study [J].
Schindera, Sebastian T. ;
Diedrichsen, Lars ;
Mueller, Hubert C. ;
Rusch, Oliver ;
Marin, Daniele ;
Schmidt, Bernhard ;
Raupach, Rainer ;
Vock, Peter ;
Szucs-Farkas, Zsolt .
RADIOLOGY, 2011, 260 (02) :454-462
[28]  
TAPIOVAARA M, 2006, STUKA219
[29]   Image quality and radiation dose of low dose coronary CT angiography in obese patients: Sinogram affirmed iterative reconstruction versus filtered back projection [J].
Wang, Rui ;
Schoepf, U. Joseph ;
Wu, Runze ;
Reddy, Ryan P. ;
Zhang, Chuanchen ;
Yu, Wei ;
Liu, Yi ;
Zhang, Zhaoqi .
EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (11) :3141-3145