Optimisation of an MDCT abdominal protocol: Image quality assessment of standard vs. iterative reconstructions

被引:12
作者
Berta, L. [1 ]
Mascaro, L. [1 ]
Feroldi, P. [1 ]
Maroldi, R. [2 ]
机构
[1] Spedali Civili Brescia, Dept Imaging, Med Phys Unit, I-25123 Brescia, Italy
[2] Univ Brescia, Dept Radiol, I-25123 Brescia, Italy
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2014年 / 30卷 / 03期
关键词
Iterative reconstruction; Multi detector CT; CT protocol optimisation; Abdominal CT; Radiation dose reduction; CORONARY CT ANGIOGRAPHY; DOSE REDUCTION; ALGORITHM;
D O I
10.1016/j.ejmp.2013.07.126
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work aims to construct a method to objectively evaluate CT image quality when new clinical protocol performances must be compared with a standard reference. We compare iterative reconstruction in the image space with filtered back projection reconstruction and accurately quantify the dose reduction. The comparison strategy accounts for both physical and clinical image qualities that are evaluated using a standard metric. The quasi-ideal observer metric is also explored to verify its reportedly high correlation with perceived image quality. Water or spatial resolution phantom images are used to characterise the physical image quality using the classic metrics in the Fourier domain by calculating the modulation transfer functions and noise power spectra (NPS). The clinical-image quality is evaluated with a 4-alternative forced-choice test. The human observers are asked to detect a positive image that contains a simulated lesion in a background image. Then, the same positive images are characterised with the quasi-ideal observer metric, which calculates the non-prewhitening matched filter signal-to-noise ratio (SNRNPWMF). Iterative reconstruction strongly reduces the image noise, but the NPS are slightly shifted to lower frequencies, which gives the images a coarse graininess. Compared with the reference FBP protocol for abdomen exams, the highest dose reduction is 40% if the standard metric is used and 30% if the SNRNPWMF metric is used. The detectability test results achieve a better correlation with SNRNPWMF than with the standard metric. The identified Fourier metric is a useful descriptor of human quality perception and can be used for future protocol optimisation. (C) 2013 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:271 / 279
页数:9
相关论文
共 29 条
[1]   Low-dose CT of the lung: potential value of iterative reconstructions [J].
Baumueller, Stephan ;
Winklehner, Anna ;
Karlo, Christoph ;
Goetti, Robert ;
Flohr, Thomas ;
Russi, Erich W. ;
Frauenfelder, Thomas ;
Alkadhi, Hatem .
EUROPEAN RADIOLOGY, 2012, 22 (12) :2597-2606
[2]   Iterative reconstruction methods in X-ray CT [J].
Beister, Marcel ;
Kolditz, Daniel ;
Kalender, Willi A. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02) :94-108
[3]   Application of the noise power spectrum in modern diagnostic MDCT: Part II. Noise power spectra and signal to noise [J].
Boedeker, K. L. ;
McNitt-Gray, M. F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) :4047-4061
[4]   Application of the noise power spectrum in modern diagnostic MDCT: Part I. Measurement of noise power spectra and noise equivalent quanta [J].
Boedeker, K. L. ;
Cooper, V. N. ;
McNitt-Gray, M. F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) :4027-4046
[5]  
Bruder H, 2010, REDUCTION RAD DOSE C, pS131
[6]   Combination of a Low-Tube-Voltage Technique With Hybrid Iterative Reconstruction (iDose) Algorithm at Coronary Computed Tomographic Angiography [J].
Funama, Yoshinori ;
Taguchi, Katsuyuki ;
Utsunomiya, Daisuke ;
Oda, Seitaro ;
Yanaga, Yumi ;
Yamashita, Yasuyuki ;
Awai, Kazuo .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2011, 35 (04) :480-485
[7]   CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT [J].
Gervaise, Alban ;
Osemont, Benoit ;
Lecocq, Sophie ;
Noel, Alain ;
Micard, Emilien ;
Felblinger, Jacques ;
Blum, Alain .
EUROPEAN RADIOLOGY, 2012, 22 (02) :295-301
[8]   CT iterative reconstruction in image space: A phantom study [J].
Ghetti, C. ;
Ortenzia, O. ;
Serreli, G. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02) :161-165
[9]  
Glantz SA, 2007, PRIMER BIOSTATISTICS
[10]   Assessment of an iterative reconstruction algorithm (SAFIRE) on image quality in pediatric cardiac CT datasets [J].
Han, B. Kelly ;
Grant, Katharine L. R. ;
Garberich, Ross ;
Sedlmair, Martin ;
Lindberg, Jana ;
Lesser, John R. .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2012, 6 (03) :200-204