Basic principles of 18F-fluoro-deoxyglucose positron emission tomography

被引:1
作者
Standke, R [1 ]
机构
[1] GE Med Syst Deutschland GmbH, D-42665 Solingen, Germany
关键词
positron emission tomography; positron decay; detector material; standard uptake value; hybrid system;
D O I
10.1046/j.1563-2571.2002.02040.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Positron emission tomography uses photons to receive regional information about dynamic, physiologic, and biochemical processes in the living body. A positron decay is measured indirectly by the simultaneous registration of both gamma rays created by the annihilation. The event is counted, if two directly opposite located detectors register gamma rays in coincidence. Unfortunately the detectors of a positron emission tomography system do not register only true coincident events. There are also scattered and random coincidences. Different types of positron tomographs are presented and scintillation crystals, which are in use for positron emission tomography are discussed. The 2D- and 3D-acquisition methods are described as well as preprocessing methods, such as correction for attenuation, scatter and dead time. For quantification the relative parameter standard uptake value (SUV) is explained. Finally hybrid systems, such as combined positron emission tomography/computed tomography scanners and the use of computed tomography data for attenuation correction are introduced.
引用
收藏
页码:149 / 155
页数:7
相关论文
共 17 条
  • [1] CORRECTION FOR SCATTERED RADIATION IN A RING DETECTOR POSITRON CAMERA BY INTEGRAL TRANSFORMATION OF THE PROJECTIONS
    BERGSTROM, M
    ERIKSSON, L
    BOHM, C
    BLOMQVIST, G
    LITTON, J
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1983, 7 (01) : 42 - 50
  • [2] Beyer T, 2000, PET IN CLINICAL ONCOLOGY, P101
  • [3] BURGER C, 2002, EUR J NUCL MED APR
  • [4] PET and [F-18]-FDG in oncology: A clinical update
    Conti, PS
    Lilien, DL
    Hawley, K
    Keppler, J
    Grafton, ST
    Bading, JR
    [J]. NUCLEAR MEDICINE AND BIOLOGY, 1996, 23 (06) : 717 - 735
  • [5] DAHLBOM M, 1992, J NUCL MED, V33, P1191
  • [6] Karp JS, 1998, J NUCL MED, V39, p50P
  • [7] PERFORMANCE-CHARACTERISTICS OF AN 8-RING WHOLE-BODY PET SCANNER
    KOPS, ER
    HERZOG, H
    SCHMID, A
    HOLTE, S
    FEINENDEGEN, LE
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1990, 14 (03) : 437 - 445
  • [8] Melcher CL, 2000, PET IN CLINICAL ONCOLOGY, P85
  • [9] MUEHLLEHNER G, 1986, J NUCL MED, V27, P90
  • [10] Patton JA, 1999, J NUCL MED, V40, P432