GlowFit - a new tool for thermoluminescence glow-curve deconvolution

被引:243
作者
Puchalska, M. [1 ]
Bilski, P. [1 ]
机构
[1] Polish Acad Sci, H Niewodniczanski Inst Nucl Phys, PL-31342 Krakow, Poland
关键词
thermoluminescence; first-order kinetic; deconvolution; LiF; Levenberg-Marquardt method;
D O I
10.1016/j.radmeas.2006.03.008
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
A new computer program, GlowFit, for deconvoluting first-order kinetics thermoluminescence (TL) glow-curves has been developed. A non-linear function describing a single glow-peak is fitted to experimental points using the least squares Levenberg-Marquardt method. The main advantage of GlowFit is the ability to resolve complex TL glow-curves consisting of strongly overlapping peaks, as those observed in heavily-doped LiF:Mg,Ti (MTT) detectors. This resolution is achieved mainly by setting constraints or by fixing selected parameters. The initial values of the fitted parameters are placed in the so-called pattern files. GlowFit is a Microsoft Windows-operated user-friendly program. The graphic interface enables easy intuitive manipulation of glow-peaks, at the initial stage (parameter initialization) and at the final stage (manual adjustment) of fitting peak parameters to the glow-curves. The program is freely downloadable from the web site http://www.ifj.edu.pl/dept/no5/nz58/deconvolution.htm (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:659 / 664
页数:6
相关论文
共 15 条
[1]  
Abramowitz M., 1970, HDB MATH FUNCTIONS
[2]   LiF:Mg,Ti (MTT) TL detectors optimised for high-LET radiation dosimetry [J].
Bilski, P ;
Budzanowski, A ;
Olko, P ;
Mandowska, E .
RADIATION MEASUREMENTS, 2004, 38 (4-6) :427-430
[3]   Lithium fluoride:: From LiF:Mg,Ti to LiF:Mg,Cu,P [J].
Bilski, P .
RADIATION PROTECTION DOSIMETRY, 2002, 100 (1-4) :199-206
[4]   A systematic evaluation of the dependence of glow curve structure on the concentration of dopants in LiF:Mg,Cu,P [J].
Bilski, P ;
Budzanowski, M ;
Olko, P .
RADIATION PROTECTION DOSIMETRY, 1996, 65 (1-4) :195-198
[5]  
BOS AJJ, 1993, RADIAT PROT DOSIM, V47, P473
[6]  
BOS AJJ, 1994, RADIAT PROT DOSIM, V51, P257
[7]  
BOS AJJ, 1993, IRICIEMAT
[8]  
Brandt S., 1999, DATA ANAL
[9]   Computerised glow curve analysis: A tool for routine thermoluminescence dosimetry [J].
Delgado, A ;
Ros, JMG .
RADIATION PROTECTION DOSIMETRY, 2001, 96 (1-3) :127-132
[10]  
HOROWITZ YS, 1986, RADIAT PROT DOSIM, V17, P337