Fiber optic surface plasmon resonance sensor for detection of E. coli O157:H7 based on antimicrobial peptides and AgNPs-rGO

被引:119
作者
Zhou, Chen [1 ]
Zou, Haimin [1 ,2 ]
Li, Ming [2 ]
Sun, Chengjun [1 ]
Ren, Dongxia [1 ]
Li, Yongxin [1 ,3 ]
机构
[1] Sichuan Univ, West China Sch Publ Hlth, Chengdu, Sichuan, Peoples R China
[2] Chengdu Ctr Dis Control & Prevent, Chengdu, Sichuan, Peoples R China
[3] Sichuan Univ, Coll Life Sci, Chengdu, Sichuan, Peoples R China
基金
中国博士后科学基金;
关键词
Optical fiber; Surface plasmon resonance; Escherichia coli O157:H7; Antimicrobial peptides; Silver nanoparticles-reduced graphene oxide; ESCHERICHIA-COLI; PREVALENCE; BIOSENSOR; SAMPLES; WATER; AMPLIFICATION; CATTLE; GENES; BEEF;
D O I
10.1016/j.bios.2018.06.005
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A fiber optic surface plasmon resonance (FOSPR) sensor was developed for detection of Escherichia coli O157:H7 (E. coli O157:H7) in water and juice, based on antimicrobial peptides (AMP), Magainin I, as recognition elements and silver nanoparticles-reduced graphene oxide (AgNPs-rGO) nanocomposites assisted signal amplification. The uniform AgNPs-rGO was fixed on the surface of optical fiber and covered with gold film. Not only was the SPR response greatly enhanced, but also the AgNPs was prevented from being oxidized. The FOSPR showed a sensitivity of about 1.5 times higher than that fabricated only with gold film. In the assay, Magainin I, immobilized on the surface of gold film, could specifically capture E. coli O157:H7, resulting in the wavelength shift of the SPR absorption peak. Under the optimized conditions, the SPR resonance wavelength exhibited a good linear relationship with natural logarithm of the target bacteria concentration in the range of 1.0 x 10(3) to 5.0 x 10(7) cfu/mL with the detection limit of 5.0 x 10(2) cfu/mL (S/N = 3). The FOSPR sensor showed good specificity for E. coli O157:H7 detection compared to other bacteria similar to the target bacterial species. Furthermore, the FOSPR sensor was successfully applied to the detection of E. coli O157:H7 in water, fruit and vegetable juice with the satisfactory recoveries of 88-110%. This assay for E. coli O157:H7 detection possesses high sensitivity, good selectivity, reproducibility and stability. In addition, the AMP based SPR biosensing methodology could be extended to detect a wide variety of foodborne pathogens. Therefore, the versatile method might become a potential alternative tool in food analysis and early clinical diagnosis.
引用
收藏
页码:347 / 353
页数:7
相关论文
共 35 条
[1]   A rapid and highly specific immunofluorescence method to detect Escherichia coli O157:H7 in infected meat samples [J].
Balakrishnan, Baskar ;
Barizuddin, Syed ;
Wuliji, Tumen ;
El-Dweik, Majed .
INTERNATIONAL JOURNAL OF FOOD MICROBIOLOGY, 2016, 231 :54-62
[2]   Highly sensitive detection of pathogen Escherichia coli O157:H7 by electrochemical impedance spectroscopy [J].
Barreiros dos Santos, M. ;
Agusil, J. P. ;
Prieto-Simon, B. ;
Sporer, C. ;
Teixeira, V. ;
Samitier, J. .
BIOSENSORS & BIOELECTRONICS, 2013, 45 :174-180
[3]   PEPTIDE ANTIBIOTICS AND THEIR ROLE IN INNATE IMMUNITY [J].
BOMAN, HG .
ANNUAL REVIEW OF IMMUNOLOGY, 1995, 13 :61-92
[4]   The Roles of Antimicrobial Peptides in Innate Host Defense [J].
Diamond, Gill ;
Beckloff, Nicholas ;
Weinberg, Aaron ;
Kisich, Kevin O. .
CURRENT PHARMACEUTICAL DESIGN, 2009, 15 (21) :2377-2392
[5]   Detection of Escherichia coli O157 and Escherichia coli O157:H7 by the immunomagnetic separation technique and stx1 and stx2 genes by multiplex PCR in slaughtered cattle in Samsun Province, Turkey [J].
Inat, Gokhan ;
Siriken, Belgin .
JOURNAL OF VETERINARY SCIENCE, 2010, 11 (04) :321-326
[6]   Size control of gold nanocrystals in citrate reduction: The third role of citrate [J].
Ji, Xiaohui ;
Song, Xiangning ;
Li, Jun ;
Bai, Yubai ;
Yang, Wensheng ;
Peng, Xiaogang .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2007, 129 (45) :13939-13948
[7]   Visual endpoint detection of Escherichia coli O157:H7 using isothermal Genome Exponential Amplification Reaction (GEAR) assay and malachite green [J].
Jothikumar, Prithiviraj ;
Narayanan, Jothikumar ;
Hill, Vincent R. .
JOURNAL OF MICROBIOLOGICAL METHODS, 2014, 98 :122-127
[8]   Simultaneous and sensitive determination of ascorbic acid, dopamine, uric acid, and tryptophan with silver nanoparticles-decorated reduced graphene oxide modified electrode [J].
Kaur, Balwinder ;
Pandiyan, Thangarasu ;
Satpati, Biswarup ;
Srivastava, Rajendra .
COLLOIDS AND SURFACES B-BIOINTERFACES, 2013, 111 :97-106
[9]   Antimicrobial peptide-based array for Escherichia coli and Salmonella screening [J].
Kulagina, Nadezhda V. ;
Shaffer, Kara M. ;
Anderson, George P. ;
Ligler, Frances S. ;
Taitt, Chris R. .
ANALYTICA CHIMICA ACTA, 2006, 575 (01) :9-15
[10]   Antimicrobial peptides for detection of bacteria in biosensor assays [J].
Kulagina, NV ;
Lassman, ME ;
Ligler, FS ;
Taitt, CR .
ANALYTICAL CHEMISTRY, 2005, 77 (19) :6504-6508