Integrated Circuits for Volumetric Ultrasound Imaging With 2-D CMUT Arrays

被引:83
|
作者
Bhuyan, Anshuman [1 ]
Choe, Jung Woo [1 ]
Lee, Byung Chul [1 ]
Wygant, Ira O. [2 ]
Nikoozadeh, Amin [1 ]
Oralkan, Omer [3 ]
Khuri-Yakub, Butrus T. [1 ]
机构
[1] Stanford Univ, Edward L Ginzton Lab, Stanford, CA 94305 USA
[2] Texas Instruments Inc, Sunnyvale, CA 94089 USA
[3] N Carolina State Univ, Dept Elect Engn, Raleigh, NC 27606 USA
关键词
2D array; capacitive micromachined ultrasonic transducer (CMUT); flip-chip bonding; integrated circuits; phased array imaging; ultrasound; volumetric imaging; TRANSDUCERS; VOLTAGE; FETUS;
D O I
10.1109/TBCAS.2014.2298197
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Real-time volumetric ultrasound imaging systems require transmit and receive circuitry to generate ultrasound beams and process received echo signals. The complexity of building such a system is high due to requirement of the front-end electronics needing to be very close to the transducer. A large number of elements also need to be interfaced to the back-end system and image processing of a large dataset could affect the imaging volume rate. In this work, we present a 3-D imaging system using capacitive micromachined ultrasonic transducer (CMUT) technology that addresses many of the challenges in building such a system. We demonstrate two approaches in integrating the transducer and the front-end electronics. The transducer is a 5-MHz CMUT array with an 8 mm x 8 mm aperture size. The aperture consists of 1024 elements (32 x 32) with an element pitch of 250 mu m. An integrated circuit (IC) consists of a transmit beamformer and receive circuitry to improve the noise performance of the overall system. The assembly was interfaced with an FPGA and a back-end system (comprising of a data acquisition system and PC). The FPGA provided the digital I/O signals for the IC and the back-end system was used to process the received RF echo data (from the IC) and reconstruct the volume image using a phased array imaging approach. Imaging experiments were performed using wire and spring targets, a ventricle model and a human prostrate. Real-time volumetric images were captured at 5 volumes per second and are presented in this paper.
引用
收藏
页码:796 / 804
页数:9
相关论文
共 50 条
  • [21] Integration of a Dual-Mode Catheter for Ultrasound Image Guidance and HIFU Ablation using a 2-D CMUT Array
    Jang, Ji Hoon
    Chang, Chienliu
    Rasmussen, Morten Fischer
    Moini, Azadeh
    Brenner, Kevin
    Stephens, Douglas N.
    Oralkan, Omer
    Khuri-Yakub, Butrus
    2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [22] 2D airborne ultrasound piezotransducer arrays for corneal imaging
    Aminot, Antoine
    Shirkovskiy, Pavel
    Dorme, Christian
    Legros, Mathieu
    Dufait, Remi
    Fink, Mathias
    Ing, Ros Kiri
    2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 800 - 802
  • [23] Versatile Low-Cost Volumetric 3-D Ultrasound Platform for Existing Clinical 2-D Systems
    Morgan, Matthew R.
    Broder, Joshua S.
    Dahl, Jeremy J.
    Herickhoff, Carl D.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (10) : 2248 - 2256
  • [24] Volumetric photoacoustic/ultrasound imaging using 2D matrix array transducer scanner
    Kim, Wangyu
    Choi, Wonseok
    Ahn, Joongho
    Lee, Changyeop
    Kim, Chulhong
    PHOTONS PLUS ULTRASOUND: IMAGING AND SENSING 2022, 2022, 11960
  • [25] Computer implementation in the reconstruction of 2-D flow velocity fields in ultrasound doppler color imaging
    Fei, DY
    Fu, CT
    Liu, DD
    COMPUTERS IN BIOLOGY AND MEDICINE, 1995, 25 (06) : 495 - 503
  • [26] ANALYSIS OF 2-D ULTRASOUND CARDIAC STRAIN IMAGING USING JOINT PROBABILITY DENSITY FUNCTIONS
    Ma, Chi
    Varghese, Tomy
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2014, 40 (06) : 1118 - 1132
  • [27] Effect of Respirophasic Displacement of the Inferior Vena Cava on Size Measurement in 2-D Ultrasound Imaging
    Policastro, Piero
    Ermini, Leonardo
    Civera, Stefania
    Albani, Stefano
    Musumeci, Giuseppe
    Roatta, Silvestro
    Mesin, Luca
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2024, 50 (12) : 1785 - 1792
  • [28] Investigation of the beam pattern of an integrated 2D CMUT spiral array element
    La Mura, Monica
    Lamberti, Patrizia
    Savoia, Alessandro S.
    2022 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IEEE IUS), 2022,
  • [29] Transtracheal 2-D ultrasound for identification of esophageal intubation
    Milling, Truman John
    Jones, Matthew
    Khan, Tara
    Tad-y, Darlene
    Melniker, Lawrence A.
    Bove, Joseph
    Yarmush, Jody
    SchianodiCola, Joseph
    JOURNAL OF EMERGENCY MEDICINE, 2007, 32 (04) : 409 - 414
  • [30] IMPROVING INTERPRETABILITY OF 2-D ULTRASOUND OF THE LUMBAR SPINE
    Porto, Lucas
    Rohling, Robert
    2020 IEEE 17TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI 2020), 2020, : 717 - 721