A system for quality assurance in radiosurgery

被引:3
作者
Mack, A [1 ]
Scheib, SG [1 ]
Rieker, M [1 ]
Weltz, D [1 ]
Mack, G [1 ]
Czempiel, H [1 ]
Kreiner, HJ [1 ]
Boettcher, HD [1 ]
Seifert, V [1 ]
机构
[1] Gamma Knife Zentrum, DE-60528 Frankfurt, Germany
来源
RADIOSUGERY, VOL 5 | 2004年 / 5卷
关键词
D O I
10.1159/000078123
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
A quality assurance system has been developed for stereotactic radiosurgery and radiotherapy, to fulfill three goals: first to evaluate the geometrical error within imaging, second to simulate the complete chain of a treatment beginning with the imaging and ending with the application of dose, and third to verify calculated dose distributions in three dimensions with high spatial resolution and accuracy. A water-filled cylindrical imaging phantom has been manufactured in which equally spaced rods are placed. It can be mounted to a stereotactic frame in different orientations to be imaged using CT and MRI. The distances of the rod positions are analyzed and a reference grid can be positioned to calculate displacement vectors according to which e.g. the head coil can be adjusted. The spectral emission of the light and the filters together with the efficiency of the red CCD channel were matched and balanced with the absorption spectra of the radiochromic film used with the adapted drum scanner. Standardized software has been designed for evaluating the film data, for calibration, for correction of temperature and darkening and for test routines. A spherical phantom with shells and inserts of different tissue equivalent materials has been developed to simulate complex treatment situations within the human skull. Copyright (C) 2004 S. Karger AG, Basel.
引用
收藏
页码:236 / 246
页数:11
相关论文
共 50 条
  • [31] Report of AAPM TG 135: Quality assurance for robotic radiosurgery
    Dieterich, Sonja
    Cavedon, Carlo
    Chuang, Cynthia F.
    Cohen, Alan B.
    Garrett, Jeffrey A.
    Lee, Charles L.
    Lowenstein, Jessica R.
    d'Souza, Maximian F.
    Taylor, David D., Jr.
    Wu, Xiaodong
    Yu, Cheng
    MEDICAL PHYSICS, 2011, 38 (06) : 2914 - 2936
  • [32] Dosimetric Performance of the Plastic Scintillation Detector for Radiosurgery Quality Assurance
    Gagnon, J.
    Theriault, D.
    Guillot, M.
    Archambault, L.
    Beddar, S.
    Gingras, L.
    Beaulieu, L.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [33] Implementation of a New UPMC Gamma Knife Radiosurgery Quality Assurance Registry
    Niranjan, Ajay
    Layne, Denise
    Briercheck, Mark
    Trofimova, Svetlana
    Monaco, Edward A., III
    Kano, Hideyuki
    Lunsford, L. Dade
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2017, 95 (01) : 49 - 59
  • [34] THE LOST METASTASES: DEEP LEARNING'S POTENTIAL IN RADIOSURGERY QUALITY ASSURANCE
    Bang, Charmin
    Chartrand, Gabriel
    Pawlowski, Sophie
    Emiliani, Ramon
    Markel, Daniel
    Bahig, Houda
    Samak, Alexandre
    Rajakesari, Selvan
    Lavoie, Jeremi
    Ducharme, Simon
    Roberge, David
    NEURO-ONCOLOGY, 2022, 24 : 50 - 51
  • [35] An Inexpensive and Convenient Phantom for Quality Assurance in Image Guidance Based Radiosurgery
    Soisson, E.
    MEDICAL PHYSICS, 2012, 39 (07) : 4633 - 4634
  • [36] Radioluminescence imaging feasibility for robotic radiosurgery field size quality assurance
    Shakarami, Zahra
    Broggi, Sara
    Del Vecchio, Antonella
    Fiorino, Claudio
    Spinelli, Antonello E.
    MEDICAL PHYSICS, 2022, 49 (10) : 6588 - 6598
  • [37] Automated Quality Assurance Test for MLC-Based Stereotactic Radiosurgery
    Mamalui-Hunter, M.
    Drzymala, R.
    Low, D.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [38] Stereotactic radiosurgery in India: A nationwide survey of technology and quality assurance practices
    Patwe, Parimal T.
    Deshpande, Sudesh
    Chaudhari, Suresh
    Mahajan, Gajanan R.
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2024, 20 (03) : 1013 - 1019
  • [39] A quality assurance program for stereotactic radiosurgery using a miniature multileaf collimator
    Kim, CS
    Schella, J
    MEDICAL PHYSICS, 2003, 30 (06) : 1343 - 1343
  • [40] NOVEL METHOD FOR QUALITY ASSURANCE OF STEREOTAXIC LINEAR ACCELERATORS USED FOR RADIOSURGERY
    SAYEG, JA
    MCLAUGHLIN, WL
    LEWIS, DF
    RADIOLOGY, 1992, 182 (02) : 611 - 611