In vivo skeletal responses to porous-surfaced implants subjected to small induced motions

被引:341
|
作者
Jasty, M
Bragdon, C
Burke, D
OConnor, D
Lowenstein, J
Harris, WH
机构
[1] MASSACHUSETTS GEN HOSP, DEPT ORTHOPAED SURG, HIP & IMPLANT SURG UNIT, BOSTON, MA 02114 USA
[2] HARVARD UNIV, SCH MED, BOSTON, MA 02114 USA
关键词
D O I
10.2106/00004623-199705000-00010
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Cylindrical porous-coated implants were placed in the distal femoral metaphyses of twenty dogs and were subjected to zero, twenty, forty, or 150 micrometers of oscillatory motion for eight hours each day for six weeks with use of a specially designed loading apparatus, The in vivo skeletal responses to the different magnitudes of relative motion were evaluated, Histological analysis demonstrated growth of bone into the porous coatings of all of the implants, including those that had been subjected to 150 micrometers of motion, However, the ingrown bone was in continuity with the surrounding bone only in the groups of implants that had not been subjected to motion or that had been subjected to twenty micrometers of motion; in contrast, the implants that had been subjected to forty micrometers of motion were surrounded in part by trabecular bone but also in part by fibrocartilage and fibrous tissue, and those that had been subjected to 150 micrometers of motion were surrounded by dense fibrous tissue, Trabecular microfractures were identified around three of the five implants that had been subjected to forty micrometers of motion and around four of the five that had been subjected to 150 micrometers of motion, suggesting that the ingrown bone had failed at the interface because of the large movements. The architecture of the surrounding trabecular bone also was altered by the micromotion of the implant. The implants that had stable ingrowth of bone were surrounded by a zone of trabecular atrophy, whereas those that had unstable ingrowth of bone were surrounded by a zone of trabecular hypertrophy, The trabeculae surrounding the fibrocartilage or fibrous tissue that had formed around the implants that had been subjected to forty or 150 micrometers of motion had been organized into a shell of dense bone tangen-tial to the implant (that is, a neocortex outside the non-osseous tissue). CLINICAL RELEVANCE: The findings of the present study quantitate the in vivo patterns of bone in-growth and remodeling that occur in association with different magnitudes of micromovement of porous-coated implants. Small movements (zero and twenty micrometers) are compatible with stable ingrowth of bone and atrophy of the surrounding trabecular bone, whereas larger movements (forty and 150 micrometers) result in less stable or unstable ingrowth of bone, the formation of fibrocartilage or fibrous tissue around the implant, and hypertrophy of the surrounding trabecular bone, This study not only quantified the magnitudes of relative micromotion that cause these different skeletal responses but also may help in the interpretation of radiographs of patients who have a porous-coated prosthesis.
引用
收藏
页码:707 / 714
页数:8
相关论文
共 43 条
  • [21] Managing the posterior mandible of partially edentulous patients with short, porous-surfaced dental implants: Early data from a clinical trial
    Deporter, D
    Pilliar, RM
    Todescan, R
    Watson, P
    Pharoah, M
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2001, 16 (05) : 653 - 658
  • [22] A Retrospective Study of Sintered Porous-Surfaced Dental Implants in Restoring the Edentulous Posterior Mandible: Up to 9 Years of Functioning
    Sohn, Dong-Seok
    Kim, Woo-Sung
    Lee, Won-Hyuk
    Jung, Heui-Seung
    Shin, Im-Hee
    IMPLANT DENTISTRY, 2010, 19 (05) : 409 - +
  • [23] Calcium phosphate sol-gel-derived thin films on porous-surfaced implants for enhanced osteoconductivity. Part II: Short-term in vivo studies
    Gan, L
    Wang, H
    Tache, A
    Valiquette, N
    Deporter, D
    Pilliar, R
    BIOMATERIALS, 2004, 25 (22) : 5313 - 5321
  • [24] Effect of surface chemistry on the rate of osseointegration of sintered porous-surfaced Ti-6Al-4V implants
    Taché, A
    Gan, L
    Deporter, D
    Pillar, RM
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2004, 19 (01) : 19 - 29
  • [25] An assessment of crown-to-root ratios with short sintered porous-surfaced implants supporting prostheses in partially edentulous patients
    Rokni, S
    Todescan, R
    Watson, P
    Pharoah, M
    Adegbembo, AO
    Deporter, D
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2005, 20 (01) : 69 - 76
  • [26] Mechanical properties of fully porous and porous-surfaced Ti-6Al-4V implants fabricated by electro-discharge-sintering
    Jo, Y. J.
    Lee, C. M.
    Jang, H. S.
    Lee, N. S.
    Suk, J.-H.
    Lee, W. H.
    JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2007, 194 (1-3) : 121 - 125
  • [27] The effect of sol-gel-formed calcium phosphate coatings on bone ingrowth and osteoconductivity of porous-surfaced Ti alloy implants
    Nguyen, HQ
    Deporter, DA
    Pilliar, RM
    Valiquette, N
    Yakubovich, R
    BIOMATERIALS, 2004, 25 (05) : 865 - 876
  • [28] Effect of calcium phosphate surface coating on bone ingrowth onto porous-surfaced titanium alloy implants in rabbit tibiae - Discussion
    Block, MS
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2002, 60 (04) : 426 - 426
  • [29] Peri-implant conditions around sintered porous-surfaced (SPS) implants. A 36-month prospective cohort study
    Malchiodi, Luciano
    Ghensi, Paolo
    Cucchi, Alessandro
    Pieroni, Stefano
    Bertossi, Dario
    CLINICAL ORAL IMPLANTS RESEARCH, 2015, 26 (02) : 212 - 219
  • [30] Performance of titanium alloy sintered porous-surfaced (SPS) implants supporting mandibular overdentures during a 20-year prospective study
    Deporter, Douglas
    Pharoah, Michael
    Yeh, Simon
    Todescan, Reynaldo
    Atenafu, Eshetu G.
    CLINICAL ORAL IMPLANTS RESEARCH, 2014, 25 (02) : E189 - E195