Ray tracing for intraocular lens calculation

被引:107
|
作者
Preussner, PR
Wahl, J
Lahdo, H
Dick, B
Findl, O
机构
[1] Univ Mainz, Augenklin, D-6500 Mainz, Germany
[2] Univ Vienna, Augenklin, Vienna, Austria
来源
关键词
D O I
10.1016/S0886-3350(01)01346-3
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To improve accuracy in intraocular lens (IOL) calculations and clarify the effect of various errors. Setting: University eye hospitals, Mainz, Germany, and Vienna, Austria. Methods: A numerical ray-tracing calculation has been developed for the pseudophakic eye. Individual rays are calculated and then undergo refractions on all surfaces of the IOL and cornea. The calculations do not use approximations; ie, the refractions are calculated exactly using Snell's law. Rays can be calculated for any distance from the optical axis and for other parameter variations. The effects of aspheric surfaces can also be investigated. Instead of IOL powers, manufacturers' IOL data (radii, refractive index, thickness) are used in the calculations for different IOL types. The resulting optical quality is visualized by using Landolt rings superimposed on the grid of retinal receptors. Results: Intraocular lens design, corneal asphericity, and specific spherical aberration influence the visual quality of the pseudophakic eye significantly. The IOL refractive power is an ambiguous parameter that cannot characterize the visual outcome sufficiently accurately for an IOL implanted at a given position. The effects can be calculated only in numerical ray tracing, not in Gaussian optics. The accuracy of numerical ray tracing is independent of axial length. Therefore, very long or very short eyes gain the most from the higher accuracy of this approach. For average-size eyes, however, the results are the same as with SRK calculations. Conclusion: Calculations in Gaussian optics should be replaced by state-of-the-art numerical methods, which can be run on any standard personal computer. (C) 2002 ASCRS and ESCRS.
引用
收藏
页码:1412 / 1419
页数:8
相关论文
共 50 条
  • [21] Ray-tracing intraocular lens power calculation using anterior segment optical coherence tomography measurements
    Minami, Keiichiro
    Kataoka, Yasushi
    Matsunaga, Jiro
    Ohtani, Shinichiro
    Honbou, Masato
    Miyata, Kazunori
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (10): : 1758 - 1763
  • [22] Accuracy of Constant C for Ray Tracing: Assisted Intraocular Lens Power Calculation in Normal Ocular Axial Eyes
    Chang, Ping-Jun
    Zhang, Fan
    Wang, Jia-Jun
    Zhao, Yun-E
    OPHTHALMIC RESEARCH, 2021, 64 (01) : 85 - 90
  • [23] Intraocular lens power calculation in virgin eyes: Accuracy of the Barrett Universal II formula and a Ray tracing software
    Fernandez-Roses, Joaquim
    Lamarca, Jose
    Pinero, David P.
    Barraquer, Rafael, I
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2022, 32 (05) : 2954 - 2960
  • [24] Astigmatism induced by intraocular lens tilt evaluated via ray tracing
    Weikert, Mitchell P.
    Golla, Abhinav
    Wang, Li
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2018, 44 (06): : 745 - 749
  • [25] Comparison of intraocular lens power calculation formulas with and without total keratometry and ray tracing in patients with previous myopic SMILE
    Elotmani, Anouar Meziane
    Messerschmidt-Roth, Anke
    Nehme, Alexandra
    Mueller, Hans-Helge
    Sekundo, Walter
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2023, 49 (05): : 467 - 473
  • [26] Intraocular lens power calculation using standard formulas and ray tracing after DMEK in patients with Fuchs endothelial dystrophy
    Alnawaiseh, Maged
    Zumhagen, Lars
    Rosentreter, Andre
    Eter, Nicole
    BMC OPHTHALMOLOGY, 2017, 17
  • [27] Intraocular lens power calculation using standard formulas and ray tracing after DMEK in patients with Fuchs endothelial dystrophy
    Maged Alnawaiseh
    Lars Zumhagen
    André Rosentreter
    Nicole Eter
    BMC Ophthalmology, 17
  • [28] Using the lens haptic plane concept and thick-lens ray tracing to calculate intraocular lens power
    Norrby, S
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (05): : 1000 - 1005
  • [29] Pilot evaluation of refractive prediction errors associated with a new method for ray-tracing-based intraocular lens power calculation
    Hirnschall, Nino
    Buehren, Tobias
    Trost, Micheal
    Findl, Oliver
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (06): : 738 - 744
  • [30] Comparison of ray-tracing method and thin-lens formula in intraocular lens power calculations
    Jin, Haiying
    Rabsilber, Tanja
    Ehmer, Angela
    Borkenstein, Andreas F.
    Limberger, Il-Joo
    Guo, Haike
    Auffarth, Gerd U.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (04): : 650 - 662