Rapid and accurate measurement of transverse relaxation times using a single shot multi-echo echo-planar imaging sequence

被引:11
作者
Tyler, DJ [1 ]
Moore, RJ [1 ]
Marciani, L [1 ]
Gowland, PA [1 ]
机构
[1] Univ Nottingham, Sch Phys & Astron, Magnet Resonance Ctr, Nottingham NG7 2RD, England
关键词
transverse relaxation; quantitation; rapid measurement; EPI;
D O I
10.1016/j.mri.2004.01.069
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Methods for making rapid and accurate measurements and maps of the transverse relaxation time from a single free induction decay (FID) are proposed. The methods use a multi-echo sequence in combination with B, insensitive (hyperbolic secant or BIREF2b) refocusing pulses and rapid echo-planar imaging techniques. The results were calibrated against a single spin echo echo-planar imaging sequence using a phantom containing a range of CuSO4 concentrations. The mean percentage absolute difference between the multi-echo and single-echo results was 3% for the multi-echo sequence using the hyperbolic secant refocusing pulse, and 7% for the multi-echo sequence using the BIREF2b refocusing pulse, compared to 13% for a multi-echo sequence using a nonselective sinc refocusing pulse. The use of the sequences in vivo has been demonstrated in studies of gastric function, i.e., the measurement of gastric dilution and monitoring of formation of a raft of alginate polysaccharide within the stomach. (C) 2004 Elsevier Inc. All rights reserved.
引用
收藏
页码:1031 / 1037
页数:7
相关论文
共 24 条
[1]  
Bevington R., 1969, DATA REDUCTION ERROR
[2]   EFFECTS OF DIFFUSION ON FREE PRECESSION IN NUCLEAR MAGNETIC RESONANCE EXPERIMENTS [J].
CARR, HY ;
PURCELL, EM .
PHYSICAL REVIEW, 1954, 94 (03) :630-638
[3]   VARIABLE-RATE SELECTIVE EXCITATION [J].
CONOLLY, S ;
NISHIMURA, D ;
MACOVSKI, A ;
GLOVER, G .
JOURNAL OF MAGNETIC RESONANCE, 1988, 78 (03) :440-458
[4]   A REDUCED POWER SELECTIVE ADIABATIC SPIN-ECHO PULSE SEQUENCE [J].
CONOLLY, S ;
GLOVER, G ;
NISHIMURA, D ;
MACOVSKI, A .
MAGNETIC RESONANCE IN MEDICINE, 1991, 18 (01) :28-38
[5]   ERRORS IN T2 ESTIMATION USING MULTISLICE MULTIPLE-ECHO IMAGING [J].
CRAWLEY, AP ;
HENKELMAN, RM .
MAGNETIC RESONANCE IN MEDICINE, 1987, 4 (01) :34-47
[6]  
Does MD, 2000, NMR BIOMED, V13, P1, DOI 10.1002/(SICI)1099-1492(200002)13:1<1::AID-NBM603>3.0.CO
[7]  
2-E
[8]  
DUIJN JH, 1984, P 3 SOC MAG RES NEW, P197
[9]   In vivo relaxation time measurements in the human placenta using echo planar imaging at 0.5 T [J].
Gowland, PA ;
Freeman, A ;
Issa, B ;
Boulby, P ;
Duncan, KR ;
Moore, RJ ;
Baker, PN ;
Bowtell, RW ;
Johnson, IR ;
Worthington, BS .
MAGNETIC RESONANCE IMAGING, 1998, 16 (03) :241-247
[10]   MULTIPLE-SPIN-ECHO IMAGING WITH A 2D FOURIER METHOD [J].
GRAUMANN, R ;
OPPELT, A ;
STETTER, E .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (05) :707-721