Experimental investigation of the change of the electron saturation current of a dc-heated emissive probe

被引:14
作者
Marek, A.
Pickova, I.
Kudrna, P.
Tichy, M.
Apetrei, R. P.
Olenici, S. B.
Gstrein, R.
Schrittwieser, R.
Ionita, C.
机构
[1] Charles Univ Prague, Fac Math & Phys, CR-18000 Prague, Czech Republic
[2] Alexandru Ioan Cuza Univ, Fac Phys, RO-700506 Iasi, Romania
[3] Univ Innsbruck, Inst Ion Phys & Appl Phys, A-6020 Innsbruck, Austria
关键词
emissive probe; probe diagnostics; cylindrical magnetron; DP machine;
D O I
10.1007/s10582-006-0306-0
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
We report on experimental investigations of the change of the electron saturation current of a de-heated emissive probe with the probe heating current. According to the simple theory of the emissive probe, the electron saturation current should not be affected by emission. However, in many experiments a variation of the electron saturation current with the emission current was observed. We consider two possible reasons for such variations: (a) the influence of the space charge around the probe shaft, (b) the change of the work function of the probe surface material due to heating. We tried to find sufficient experimental evidence for supporting one or the other (or both) of these two explanations. We used two different types of plasma to validate the results: a cylindrical magnetron plasma and the non-magnetized plasma of a DP machine. From our experiments follows that the electron saturation current of the emissive probe is affected by the space charge effect as well as it depends on the probe wire material.
引用
收藏
页码:B932 / B937
页数:6
相关论文
共 11 条
[1]   Emissive probe measurements of plasma potential fluctuations in the edge plasma regions of tokamaks [J].
Balan, P ;
Schrittwieser, R ;
Ionita, C ;
Cabral, JA ;
Figueiredo, HFC ;
Fernandes, H ;
Varandas, C ;
Adámek, J ;
Hron, M ;
Stöckel, J ;
Martines, E ;
Tichy, M ;
Van Oost, G .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (03) :1583-1587
[2]  
BRADLYE JW, 2000, PLASMA SOURCES SCI T, V13, P189
[3]  
Dimitriu DG, 2004, ACTA PHYS SLOVACA, V54, P89
[4]   2-D experimental study of the plasma parameter variations or the magnetically sustained DC discharge in cylindrical symmetry in argon [J].
Holík, N ;
Bilyk, O ;
Marek, A ;
Kudrna, P ;
Behnke, JF ;
Tichy, M .
CONTRIBUTIONS TO PLASMA PHYSICS, 2004, 44 (7-8) :613-618
[5]   SPACE POTENTIAL MEASUREMENTS WITH A CONTINUOUSLY EMITTING PROBE [J].
MRAVLAG, E ;
KRUMM, P .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (08) :2164-2170
[6]   Measurements with an emissive probe in the CASTOR tokamak [J].
Schrittwieser, R ;
Adámek, J ;
Balan, P ;
Hron, M ;
Ionita, C ;
Jakubka, K ;
Kryska, L ;
Martines, E ;
Stöckel, J ;
Tichy, M ;
Van Oost, G .
PLASMA PHYSICS AND CONTROLLED FUSION, 2002, 44 (05) :567-578
[7]  
Schrittwieser R, 2001, CONTRIB PLASM PHYS, V41, P494, DOI 10.1002/1521-3986(200109)41:5<494::AID-CTPP494>3.0.CO
[8]  
2-X
[9]   A new simple emissive probe [J].
Siebenforcher, A ;
Schrittwieser, R .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1996, 67 (03) :849-850
[10]   An emissive probe with a rhenium filament for measuring plasma potential in a radio frequency oxygen plasma [J].
Wilson, EH ;
Jeong, J ;
Hershkowitz, N .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (05) :2033-2037