CT Metal Artifact Reduction in the Spine: Can an Iterative Reconstruction Technique Improve Visualization?

被引:55
|
作者
Kotsenas, A. L. [1 ]
Michalak, G. J. [1 ]
DeLone, D. R. [1 ]
Diehn, F. E. [1 ]
Grant, K. [2 ]
Halaweish, A. F. [2 ]
Krauss, A. [3 ]
Raupach, R. [3 ]
Schmidt, B. [3 ]
McCollough, C. H. [1 ]
Fletcher, J. G. [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[2] Siemens Med Solut, Malvern, PA USA
[3] Siemens Healthcare, Forchheim, Germany
关键词
DUAL-ENERGY CT; COMPUTED-TOMOGRAPHY; VIVO EVALUATION; IMPLANTS;
D O I
10.3174/ajnr.A4416
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: Metal-related artifacts from spine instrumentation can obscure relevant anatomy and pathology. We evaluated the ability of CT images reconstructed with and without iterative metal artifact reduction to visualize critical anatomic structures in postoperative spines and assessed the potential for implementation into clinical practice. MATERIALS AND METHODS: We archived CT projection data in patients with instrumented spinal fusion. CT images were reconstructed by using weighted filtered back-projection and iterative metal artifact reduction. Two neuroradiologists evaluated images in the region of spinal hardware and assigned a score for the visualization of critical anatomic structures by using soft-tissue and bone windows (critical structures totally obscured, n = 0; anatomic recognition with high diagnostic confidence, n = 5). Using bone windows, we measured the length of the most pronounced linear artifacts. For each patient, neuroradiologists made recommendations regarding the optimal use of iterative metal artifact reduction and its impact on diagnostic confidence. RESULTS: Sixty-eight patients met the inclusion criteria. Visualization of critical soft-tissue anatomic structures was significantly improved by using iterative metal artifact reduction compared with weighted filtered back-projection (median,1 +/- 1.5 versus 3 +/- 1.3, P <.001), with improvement in the worst visualized anatomic structure in 88% (60/68) of patients. There was not significant improvement in visualization of critical osseous structures. Linear metal artifacts were reduced from 29 to 11 mm (P <.001). In 87% of patients, neuroradiologists recommended reconstructing iterative metal artifact reduction images instead of weighted filtered back-projection images, with definite improvement in diagnostic confidence in 32% (22/68). CONCLUSIONS: Iterative metal artifact reduction improves visualization of critical soft-tissue structures in patients with spinal hardware. Routine generation of these images in addition to routine weighted filtered back-projection is recommended.
引用
收藏
页码:2184 / 2190
页数:7
相关论文
共 50 条
  • [21] Clinical evaluation of the iterative metal artifact reduction algorithm for CT simulation in radiotherapy
    Axente, Marian
    Paidi, Ajay
    Von Eyben, Rie
    Zeng, Chuan
    Bani-Hashemi, Ali
    Krauss, Andreas
    Hristov, Dimitre
    MEDICAL PHYSICS, 2015, 42 (03) : 1170 - 1183
  • [22] Weighted Total Variation Constrained Reconstruction for Reduction of Metal Artifact in CT
    Zhang, Yanbo
    Mou, Xuanqin
    Yan, Hao
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2630 - 2634
  • [23] Reduction of Metal Artifact in Single Photon-Counting Computed Tomography by Spectral-Driven Iterative Reconstruction Technique
    Nasirudin, Radin A.
    Mei, Kai
    Panchev, Petar
    Fehringer, Andreas
    Pfeiffer, Franz
    Rummeny, Ernst J.
    Fiebich, Martin
    Noel, Peter B.
    PLOS ONE, 2015, 10 (05):
  • [24] Development of a CT Metal Artifact Reduction Technique Based On Deep Learning
    Lee, J.
    Cho, H.
    Kim, H.
    Chae, H.
    Ye, S.
    MEDICAL PHYSICS, 2018, 45 (06) : E500 - E500
  • [25] Automated-Adaptive Iterative Metal Artifact Reduction: Quantitative Evaluation of Streak Artifact Reduction When Reconstructing CT Images in Presence of Metal
    Mistry, N.
    Duba, I.
    Halaweish, A.
    Schmidt, B.
    Hofmann, C.
    MEDICAL PHYSICS, 2018, 45 (06) : E637 - E637
  • [26] REDUCTION OF BLADDER ARTIFACT IN SPECT BY AN ITERATIVE RECONSTRUCTION ALGORITHM
    REULAND, P
    KAISER, W
    MULLERSCHAUENBURG, W
    FEINE, U
    NUC COMPACT-EUROPEAN-AMERICAN COMMUNICATIONS IN NUCLEAR MEDICINE, 1991, 22 (01): : 23 - 24
  • [27] Combination of Iterative Metal Artifact Reduction and Virtual Monoenergetic Reconstruction Using Split-Filter Dual-Energy CT in Patients With Dental Artifact on Head and Neck CT
    Schmidt, Alexander Marcel Alfred
    Grunz, Jan-Peter
    Petritsch, Bernhard
    Gruschwitz, Philipp
    Knarr, Jonas
    Huflage, Henner
    Bley, Thorsten Alexander
    Kosmala, Aleksander
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2022, 218 (04) : 716 - 727
  • [28] Radiation Dose Reduction in Temporal Bone CT with Iterative Reconstruction Technique
    Niu, Y. T.
    Mehta, D.
    Zhang, Z. R.
    Zhang, Y. X.
    Liu, Y. F.
    Kang, T. L.
    Xian, J. F.
    Wang, Z. C.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (06) : 1020 - 1026
  • [29] Diagnostic Value of Model-Based Iterative Reconstruction Combined with a Metal Artifact Reduction Algorithm during CT of the Oral Cavity
    Kubo, Y.
    Ito, K.
    Sone, M.
    Nagasawa, H.
    Onishi, Y.
    Umakoshi, N.
    Hasegawa, T.
    Akimoto, T.
    Kusumoto, M.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2020, 41 (11) : 2132 - 2138
  • [30] Application of Iterative Metal Artifact Reduction Algorithm to CT Urography for Patients With Hip Prostheses
    Trabzonlu, Tugce Agirlar
    Terrazas, Martha
    Mozaffary, Amirhossein
    Velichko, Yuri S.
    Yaghmai, Vahid
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2020, 214 (01) : 137 - 143