Radiation Dose Reduction in the Cardiac Catheterization Laboratory Utilizing a Novel Protocol

被引:58
|
作者
Wassef, Anthony W. A. [1 ]
Hiebert, Brett [1 ]
Ravandi, Amir [1 ]
Ducas, John [1 ]
Minhas, Kunal [1 ]
Vo, Minh [1 ]
Kass, Malek [1 ]
Parmar, Gurpreet [1 ]
Hussain, Farrukh [1 ]
机构
[1] Univ Manitoba, Dept Med, Div Cardiol, Winnipeg, MB R2H 2A6, Canada
关键词
diagnostic coronary angiography; frame rate; percutaneous coronary intervention; radiation; x-ray; CARDIOLOGY; QUALITY; PATIENT; CANCER; SAFETY;
D O I
10.1016/j.jcin.2013.11.022
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives This study reports the results a novel radiation reduction protocol (RRP) system for coronary angiography and interventional procedures and the determinants of radiation dose. Background The cardiac catheterization laboratory is an important source of radiation and should be kept in good working order with dose-reduction and monitoring capabilities. Methods All diagnostic coronary angiograms and percutaneous coronary interventions from a single catheterization laboratory were analyzed 2 months before and after RRP implementation. The primary outcome was the relative dose reduction at the interventional reference point. Separate analyses were done for conventional 15 frames/s (FPS) and at reduced 7.5 FPS post-RRP groups. Results A total of 605 patients underwent coronary angiography (309 before RRP and 296 after RRP), with 129 (42%) and 122 (41%) undergoing percutaneous coronary interventions before and after RRP, respectively. With RRP, a 48% dose reduction (1.07 +/- 0.05 Gy vs. 0.56 +/- 0.03 Gy, p < 0.0001) was obtained, 35% with 15 FPS RRP (0.70 +/- 0.05 Gy, p < 0.0001) and 62% with 7.5 FPS RRP (0.41 +/- 0.03 Gy, p < 0.001). Similar dose reductions for diagnostic angiograms and percutaneous coronary interventions were noted. There was no change in the number of stents placed or vessels intervened on. Increased dose was associated with male sex, radial approach, increasing body mass index, cine runs, and frame rates. Using a multivariable model, a 48% relative risk with RRP (p < 0.001), 44% with 15 FPS RRP and 68% with 7.5 FPS RRP was obtained. Conclusions We demonstrate a highly significant 48.5% adjusted radiation dose reduction using a novel algorithm, which needs strong consideration among interventional cardiology practice. (C) 2014 by the American College of Cardiology Foundation
引用
收藏
页码:550 / 557
页数:8
相关论文
共 50 条
  • [21] Primary operator radiation dose in the cardiac catheter laboratory
    Crowhurst, James A.
    Whitby, Mark
    Aroney, Nicholas
    Dautov, Rustem
    Walters, Darren
    Raffel, Owen
    BRITISH JOURNAL OF RADIOLOGY, 2020, 93 (1112)
  • [22] Efficacy of Low-Dose Compared With Standard-Dose Radiation for Cardiac Catheterization and Intervention (KAR RAD Study)
    Kar, Subrata
    Teleb, Mohamed
    Albaghdadi, Aymen
    Ibrahim, Ahmed
    Mukherjee, Debabrata
    JOURNAL OF INVASIVE CARDIOLOGY, 2019, 31 (06) : 187 - 194
  • [23] Advancements in Cardiac Catheterization Safety: Novel Radiation Protection Approaches Redefining Occupational Health
    Shaghaghi, Zahra
    Javid, Roozbeh Narimani
    Alvandi, Maryam
    CURRENT CARDIOLOGY REVIEWS, 2025, 21 (01)
  • [24] Effective Reduction of Radiation Exposure during Cardiac Catheterization
    Gutierrez-Barrios, Alejandro
    Camacho-Galan, Hugo
    Medina-Camacho, Francisco
    Canadas-Pruano, Dolores
    Jimenez-Moreno, Antonio
    Calle-Perez, German
    Vazquez-Garcia, Rafael
    TEXAS HEART INSTITUTE JOURNAL, 2019, 46 (03): : 167 - 171
  • [25] Optimizing Radiation Safety in the Cardiac Catheterization Laboratory: A Practical Approach
    Christopoulos, Georgios
    Makke, Lorenza
    Christakopoulos, Georgios
    Kotsia, Anna
    Rangan, Bavana V.
    Roesle, Michele
    Haagen, Donald
    Kumbhani, Dharam J.
    Chambers, Charles E.
    Kapadia, Samir
    Mahmud, Ehtisham
    Banerjee, Subhash
    Brilakis, Emmanouil S.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2016, 87 (02) : 291 - 301
  • [26] The Cardiac Catheterization Laboratory
    Lopez, Pedro
    RADIOLOGIC TECHNOLOGY, 2023, 94 (05) : 372 - 378
  • [27] Patient Radiation Exposure in a Modern, Large-Volume, Pediatric Cardiac Catheterization Laboratory
    Glatz, Andrew C.
    Patel, Akash
    Zhu, Xiaowei
    Dori, Yoav
    Hanna, Brian D.
    Gillespie, Matthew J.
    Rome, Jonathan J.
    PEDIATRIC CARDIOLOGY, 2014, 35 (05) : 870 - 878
  • [28] Important Strategies to Reduce Occupational Radiation Exposure in the Cardiac Catheterization Laboratory No Lower Limit
    Hirshfeld, John W.
    Fiorilli, Paul N.
    Silvestry, Frank E.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2018, 71 (11) : 1255 - 1258
  • [29] Reduction in Radiation Dose in a Pediatric Cardiac Catheterization Lab Using the Philips AlluraClarity X-ray System
    Patrick M. Sullivan
    David Harrison
    Sarah Badran
    Cheryl M. Takao
    Frank F. Ing
    Pediatric Cardiology, 2017, 38 : 1583 - 1591
  • [30] The contemporary role of protamine in the cardiac catheterization laboratory
    Danek, Barbara A.
    Kearney, Kathleen E.
    Chung, Christine J. J.
    Steinberg, Zachary
    Lombardi, William L.
    McCabe, James M.
    Azzalini, Lorenzo
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2023, 102 (01) : 111 - 120