Results of endovascular repair of infrarenal aortic aneurysms using the Endurant stent graft

被引:30
|
作者
Zandvoort, Herman J. A. [1 ]
Goncalves, Frederico Bastos [2 ]
Verhagen, Hence J. M. [2 ]
Werson, Debora A. B. [3 ]
Moll, Frans L. [1 ]
de Vries, Jean-Paul P. M. [3 ]
van Herwaarden, Joost A. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Vasc Surg, NL-3508 GA Utrecht, Netherlands
[2] Erasmus MC, Dept Vasc Surg, Rotterdam, Netherlands
[3] St Antonius Hosp, Dept Vasc Surg, Nieuwegein, Netherlands
关键词
MULTICENTER; ENDOGRAFT; REGISTRY; TRIAL;
D O I
10.1016/j.jvs.2013.12.031
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: Recent reports showed that the Endurant stent graft (Medtronic Cardiovascular, Santa Rosa, Calif) is safe and effective for endovascular repair of abdominal aortic aneurysms (AAAs). However, due to its relatively recent introduction, only short-term follow-up data are available. This study presents the 4-year results using this device. Methods: All clinical data, including detailed anatomic information of the first 100 consecutive patients treated with the Endurant stent graft for an infrarenal AAA in three Dutch high-volume hospitals, were prospectively collected. Computed tomography angiography was routinely performed before the procedure, <= 1 month, and at 1 year post-endovascular aneurysm repair. Thereafter, the imaging modality during yearly follow-up was individualized (duplex ultrasound imaging or computed tomography angiography). Patients were classified as within or outside the instructions for use (IFU) for analysis. Study end points were primary clinical success, overall and AAA-related mortality, and sac morphology changes and endoleak during follow-up. Estimates were obtained using Kaplan-Meier plots. Results: The study included 100 consecutive patients (88 men) with a median age of 74 years (interquartile range [IQR], 67-79 years) and median AAA diameter of 58 mm (IQR, 55-65 mm), between December 2007 and March 2009. Twenty patients (20%) were treated outside the IFU (18, outside proximal neck IFU, one outside iliac IFU, and one outside both IFUs). Median follow-up was 48 months (IQR, 36-53 months), and no patients were lost. One contained rupture was observed after 1.5 months due to graft infection. No patients had graft migration. Two type Ia endoleaks, 5 type Ib endoleaks, and 15 type II endoleaks were found. Primary clinical success was 97%, 90%, 84% and 77% at 1, 2, 3, and 4 years, respectively. Primary clinical success was comparable for patients treated within or outside IFU (P = .20), although both patients outside iliac IFU needed a secondary iliac intervention. Over time, maximum aneurysm diameter decreased >= 5 mm, remained stable, and increased >= 5 mm in 58%, 32%, and 10% of the patients, respectively. All-cause mortality was 20% at 4 years, with a 3% AAA-related mortality. Conclusions: The 4-year follow-up data of the Endurant stent graft for AAA treatment shows its use results in a low AAA-related mortality with adequate prevention of rupture or aneurysm growth. Although patients with very challenging anatomy were treated in our series, primary clinical success rates were comparable for patients treated within and outside the IFU. However, both patients outside the iliac IFU needed a secondary iliac intervention. The knowledge of the present results may aid in improving outcomes in the future.
引用
收藏
页码:1195 / 1202
页数:8
相关论文
共 50 条
  • [21] Active Proximal Sealing in the Endovascular Repair of Abdominal Aortic Aneurysms: Early Results With a New Stent-Graft
    Melas, Nikolaos
    Stavridis, Kyriakos
    Saratzis, Athanasios
    Lazarides, John
    Gitas, Christos
    Saratzis, Nikolaos
    JOURNAL OF ENDOVASCULAR THERAPY, 2015, 22 (02) : 174 - 178
  • [22] Safety and effectiveness of the INCRAFT AAA Stent Graft for endovascular repair of abdominal aortic aneurysms
    Torsello, Giovanni
    Scheinert, Dierk
    Brunkwall, Jan S.
    Chiesa, Roberto
    Coppi, Gioachino
    Pratesi, Carlo
    JOURNAL OF VASCULAR SURGERY, 2015, 61 (01) : 1 - 8
  • [23] Limb Graft Occlusion Following Endovascular Aneurysm Repair for Infrarenal Abdominal Aortic Aneurysm with the Zenith Alpha, Excluder, and Endurant Devices: a Multicentre Cohort Study
    Bogdanovic, Marko
    Stackelberg, Otto
    Lindstrom, David
    Ersryd, Samuel
    Andersson, Manne
    Roos, Hakan
    Siika, Antti
    Jonsson, Magnus
    Roy, Joy
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2021, 62 (04) : 532 - 539
  • [24] Midterm outcomes in patients undergoing endovascular repair of thoracic aortic aneurysms and penetrating atherosclerotic ulcers using the RelayPlus stent graft
    Malas, Mahmoud
    Locham, Satinderjit
    Hughes, Chad
    Bacharach, Michael
    Brinster, Derek
    McKinsey, James
    Mannava, Krishna
    Wu, James
    Rahimi, Saum
    Sharafuddin, Melham
    JOURNAL OF VASCULAR SURGERY, 2021, 73 (02) : 459 - 465
  • [25] Long-term results after endovascular repair of abdominal aortic aneurysms with the stentor and vanguard stent-graft
    Engellau, L
    Albrechtsson, U
    Norgren, L
    Larsson, EM
    ACTA RADIOLOGICA, 2004, 45 (03) : 275 - 283
  • [26] Midterm Results of Endovascular Aneurysm Repair Using the Endurant Stent-Graft According to the Instructions for Use vs. Off-Label Conditions
    Troisi, Nicola
    Torsello, Giovanni
    Weiss, Kristin
    Donas, Konstantinos P.
    Michelagnoli, Stefano
    Austermann, Martin
    JOURNAL OF ENDOVASCULAR THERAPY, 2014, 21 (06) : 841 - 847
  • [27] In vivo crushing of an aortic stent enables endovascular repair of a large infrarenal aortic pseudoaneurysm
    Dorros, G
    Jaff, MR
    Parikh, A
    Sehgal, R
    Thota, V
    Ramireddy, K
    Carballo, RE
    JOURNAL OF ENDOVASCULAR SURGERY, 1998, 5 (04): : 359 - 364
  • [28] Preliminary Results of Aorfix™ Stent Graft to Treat Infrarenal Abdominal Aortic Aneurysms with Severe Proximal Aortic Neck Angulation
    Volpe, Pietro
    Massara, Mafalda
    Alberti, Antonino
    Marino, Santo
    Volpe, Alberto
    Mazzaccaro, Daniela
    Stegher, Silvia
    Nano, Giovanni
    ANNALS OF VASCULAR SURGERY, 2017, 45 : 193 - 198
  • [29] Two-year outcome of the Endurant stent graft for endovascular abdominal aortic repair in Japanese patients: incidence of endoleak and aneurysm sac shrinkage
    Nishibe, Toshiya
    Iwahashi, Toru
    Kamiya, Kentaro
    Toguchi, Kayo
    Maruno, Keita
    Fujiyoshi, Toshiki
    Muromachi, Yukio
    Suzuki, Shun
    Koizumi, Jun
    Ogino, Hitoshi
    INTERNATIONAL ANGIOLOGY, 2017, 36 (03) : 237 - 242
  • [30] Five-year results of endovascular abdominal aortic aneurysm repair with the Ovation abdominal stent graft
    Swerdlow, Nicholas J.
    Lyden, Sean P.
    Verhagen, Hence J. M.
    Schermerhorn, Marc L.
    JOURNAL OF VASCULAR SURGERY, 2020, 71 (05) : 1528 - +