Transcatheter closure of multiple atrial septal defects - Initial results and value of two- and three-dimensional transoesophageal echocardiography

被引:111
|
作者
Cao, QL
Radtke, W
Berger, F
Zhu, W
Hijazi, ZM
机构
[1] Univ Chicago, Childrens Hosp, Dept Pediat, Chicago, IL 60637 USA
[2] Tufts Univ, Sch Med, New England Med Ctr, Floating Hosp Children,Dept Pediat,Div Cardiol, Boston, MA 02111 USA
[3] Med Univ S Carolina, Dept Pediat, Charleston, SC 29425 USA
[4] Deutsch Herzzentrum Berlin, Berlin, Germany
关键词
multiple atrial septal defects; Amplatzer septal occulder; echocardiography;
D O I
10.1053/euhj.1999.1909
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims To examine the feasibility of transcatheter closure of multiple atrial septal defects using two Amplatzer devices simultaneously and to describe the importance and the role of two- and three-dimensional transoesophageal echocardiography in the selection and closure of such defects. Methods Twenty-two patients with more than one atrial septal defect underwent an attempt at transcatheter closure of their atrial septal defects at a mean +/- SD age of 30.8 +/- 18.6 years (range 3.7-65.9 years) and mean weight of 56.6 +/- 25.5 kg (range 12.9-99 kg) using two Amplatzer devices implanted simultaneously via two separate delivery systems. During catheterization, two dimensional transoesophageal echocardiography was performed in all but one patient, during and after transcatheter closure, while three dimensional transoesophageal echocardiography was performed in six patients before and after transcatheter closure. Results Forty-four devices were deployed in all patients to close 45 defects tone patient with three defects closed by two devices). Two dimensional transoesophageal echocardiography was helpful in selection and in guiding correct deployment of the devices. The mean size of the larger defect, as measured by transoesophageal echocardiography was 12.8 +/- 5.9 mm and the mean size of the smaller defect was 6.6 +/- 3.0 mm. The mean size of the larger devices was 15 +/- 7.5 mm, and 8.4 +/- 3.7 mm for the smaller. Three-dimensional transoesophageal echocardiography provided superior imaging and demonstrated the number, shape and the surrounding structures of the atrial septal defects in one single view. The median fluoroscopy time was 28.7 min. Device embolization with successful catheter retrieval occurred in one patient. Forty-four devices were evaluated by colour Doppler transoesophageal echocardiography immediately after the catheterization with a successful closure rate of 97.7%. On follow-up colour Doppler transthoracic echocardiography demonstrated successful closure in 97.5% at 3 months. Conclusions The use of more than one Amplatzer septal occluder to close multiple atrial septal defects is safe and effective. The use of two- and three-dimensional transoesophageal echocardiography provided useful information for transcatheter closure of multiple atrial septal defects using two devices. Three-dimensional transoesophageal echocardiography enhanced our ability to image and understand the spatial relationship of the atrial septal defect anatomy. (Eur Heart J 2000; 21: 941-947) (C) 2000 The European Society of Cardiology.
引用
收藏
页码:941 / 947
页数:7
相关论文
共 50 条
  • [1] Real-Time Three-Dimensional Transesophageal Echocardiography Is Useful for Percutaneous Closure of Multiple Secundum Atrial Septal Defects
    Zhang, Chun
    Li, Zhian
    Xu, Jin
    HELLENIC JOURNAL OF CARDIOLOGY, 2014, 55 (06) : 486 - 491
  • [2] Transcatheter closure of multiple atrial septal defects in an adolescent
    Sungur, Metin
    Celiker, Alpay
    Ozkutlu, Suheyla
    Karagoz, Tevfik
    Baysal, Kemal
    PEDIATRICS INTERNATIONAL, 2009, 51 (05) : 756 - 758
  • [3] Clinical value of stereoscopic three-dimensional echocardiography in assessment of atrial septal defects: Feasibility and efficiency
    Jing Wang
    Xinfang Wang
    Mingxing Xie
    Lin He
    Qing Lv
    Lei Wang
    Journal of Huazhong University of Science and Technology [Medical Sciences], 2009, 29 : 791 - 794
  • [4] Clinical Value of Stereoscopic Three-dimensional Echocardiography in Assessment of Atrial Septal Defects: Feasibility and Efficiency
    王静
    王新房
    谢明星
    贺林
    吕清
    王蕾
    Current Medical Science, 2009, (06) : 791 - 794
  • [5] Clinical value of stereoscopic three-dimensional echocardiography in assessment of atrial septal defects: Feasibility and efficiency
    Wang, Jing
    Wang, Xinfang
    Xie, Mingxing
    He, Lin
    Lv, Qing
    Wang, Lei
    JOURNAL OF HUAZHONG UNIVERSITY OF SCIENCE AND TECHNOLOGY-MEDICAL SCIENCES, 2009, 29 (06) : 791 - 794
  • [6] Two- and three-dimensional transesophageal echocardiography in patient selection and assessment of atrial septal defect closure by the new DAS-angel wings device - Initial clinical experience
    Magni, G
    Hijazi, ZM
    Pandian, NG
    Delabays, A
    Sugeng, L
    Laskari, C
    Marx, GR
    CIRCULATION, 1997, 96 (06) : 1722 - 1728
  • [7] Dynamic morphology of the secundum atrial septal defect evaluated by three dimensional transoesophageal echocardiography
    Maeno, YV
    Benson, LN
    McLaughlin, PR
    Boutin, C
    HEART, 2000, 83 (06) : 673 - 677
  • [8] Personalized Three-Dimensional Printing and Echoguided Procedure Facilitate Single Device Closure for Multiple Atrial Septal Defects
    Li, Ping
    Fang, Fang
    Qiu, Xu
    Xu, Nan
    Wang, Yan
    Ouyang, Wen-Bin
    Zhang, Feng-Wen
    Hu, Hai-Bo
    Pan, Xiang-Bin
    JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2020, 2020
  • [9] Three-dimensional echocardiographic assessment of atrial septal defects
    German, Charles
    Nanda, Navin C.
    ANNALS OF CARDIAC ANAESTHESIA, 2015, 18 (01) : 69 - 73
  • [10] Kissing balloons and kissing devices: transcatheter closure of multiple atrial septal defects on hypermobile atrial septum with two devices
    Kucukdurmaz, Zekeriya
    Karapinar, Hekim
    Oflaz, Mehmet Burhan
    Gul, Ibrahim
    Yilmaz, Ahmet
    POSTEPY W KARDIOLOGII INTERWENCYJNEJ, 2012, 8 (04): : 357 - 359