Enhanced efficiency in helical tomotherapy quality assurance using a custom-designed water-equivalent phantom

被引:9
作者
Goddu, S. Murty [1 ]
Mutic, Sasa [1 ]
Pechenaya, Olga L. [1 ]
Chaudhari, Summer R. [1 ]
Garcia-Ramirez, Jose [1 ]
Rangaraj, Dharanipathy [1 ]
Klein, Eric E. [1 ]
Yang, Deshan [1 ]
Grigsby, James [1 ]
Low, Daniel A. [1 ]
机构
[1] Washington Univ, Sch Med, Dept Radiat Oncol, St Louis, MO 63110 USA
关键词
RADIOTHERAPY; OUTPUT;
D O I
10.1088/0031-9155/54/19/001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Tomotherapy is an image-guided, intensity-modulated radiation therapy system that delivers highly conformal dose distributions in a helical fashion. This system is also capable of acquiring megavoltage computed-tomography images and registering them to the planning kVCT images for accurate target localization. Quality assurance (QA) of this device is time intensive, but can be expedited by improved QA tools and procedures. A custom- designed phantom was fabricated to improve the efficiency of daily QA of our Tomotherapy machine. The phantom incorporates ionization chamber measurement points, plugs of different densities and slide-out film cartridges. The QA procedure was designed to verify in less than 30 min the vital components of the tomotherapy system: static beam quality and output, image quality, correctness of image registration and energy of the helical dose delivery. Machine output, percent depth dose and off-axis factors are simultaneously evaluated using a static 5 x 40 cm(2) open field. A single phantom scan is used to evaluate image quality and registration accuracy. The phantom can also be used for patient plan-specific QA. The QA results over a period of 6 months are reported in this paper. The QA process was found to be simple, efficient and capable of simultaneously verifying several important parameters.
引用
收藏
页码:5663 / 5674
页数:12
相关论文
共 14 条
[1]   Clinical helical tomotherapy commissioning dosimetry [J].
Balog, J ;
Olivera, G ;
Kapatoes, J .
MEDICAL PHYSICS, 2003, 30 (12) :3097-3106
[2]   Benchmarking beam alignment for a clinical helical tomotherapy device [J].
Balog, J ;
Mackie, TR ;
Pearson, D ;
Hui, S ;
Paliwal, B ;
Jeraj, R .
MEDICAL PHYSICS, 2003, 30 (06) :1118-1127
[3]   Helical tomotherapy dynamic quality assurance [J].
Balog, John ;
Holmes, Tim ;
Vaden, Richard .
MEDICAL PHYSICS, 2006, 33 (10) :3939-3950
[4]   Characterization of the output for helical delivery of intensity modulated slit beams [J].
Balog, JP ;
Mackie, TR ;
Reckwerdt, P ;
Glass, M ;
Angelos, L .
MEDICAL PHYSICS, 1999, 26 (01) :55-64
[5]   Automatic registration of megavoltage to kilovoltage CT images in helical tomotherapy:: An evaluation of the setup verification process for the special case of a rigid head phantom [J].
Boswell, Sarah ;
Tome, Wolfgang ;
Jeraj, Robert ;
Jaradat, Hazirn ;
Mackie, T. Rock .
MEDICAL PHYSICS, 2006, 33 (11) :4395-4404
[6]   Quality assurance of a helical tomotherapy machine [J].
Fenwick, JD ;
Tomé, WA ;
Jaradat, HA ;
Hui, SK ;
James, JA ;
Balog, JP ;
DeSouza, CN ;
Lucas, DB ;
Olivera, GH ;
Mackie, TR ;
Paliwal, BR .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (13) :2933-2953
[7]   COMPREHENSIVE QA FOR RADIATION ONCOLOGY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK-GROUP-40 [J].
KUTCHER, GJ ;
COIA, L ;
GILLIN, M ;
HANSON, WF ;
LEIBEL, S ;
MORTON, RJ ;
PALTA, JR ;
PURDY, JA ;
REINSTEIN, LE ;
SVENSSON, GK ;
WELLER, M ;
WINGFIELD, L .
MEDICAL PHYSICS, 1994, 21 (04) :581-618
[8]  
LOW D, 2003, MED PHYS, V30, P9
[9]   Tomotherapy [J].
Mackie, TR ;
Balog, J ;
Ruchala, K ;
Shepard, D ;
Aldridge, S ;
Fitchard, E ;
Reckwerdt, P ;
Olivera, G ;
McNutt, T ;
Mehta, M .
SEMINARS IN RADIATION ONCOLOGY, 1999, 9 (01) :108-117
[10]   Image guidance for precise conformal radiotherapy [J].
Mackie, TR ;
Kapatoes, J ;
Ruchala, K ;
Lu, WG ;
Wu, C ;
Olivera, G ;
Forrest, L ;
Tome, W ;
Welsh, J ;
Jeraj, R ;
Harari, P ;
Reckwerdt, P ;
Paliwal, B ;
Ritter, M ;
Keller, H ;
Fowler, J ;
Mehta, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01) :89-105