Diffusive solute transport in hollow fiber dialyzers is not affected by variable feed viscosity

被引:3
|
作者
Schneditz, Daniel [1 ,2 ]
Sauseng, Notburga [1 ]
机构
[1] Med Univ Graz, Otto Loewi Res Ctr, Div Physiol, Graz, Austria
[2] Med Univ Graz, Otto Loewi Res Ctr, Div Physiol, Neue Stiftingtal str 6 V, A-8010 Graz, Austria
关键词
Hemodialysis; Transport; Clearance; Viscosity; Diffusivity; Dialyzer; INTERNAL FILTRATION; MASS-TRANSFER; BLOOD-FLOW; UREA; HEMODIALYZERS; OPTIMIZATION; PERFORMANCE; CLEARANCE; EXCHANGE; REMOVAL;
D O I
10.1016/j.bbe.2022.09.003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Dialyzer clearance (K) for hemodialysis is usually predicted from the mass transfer area product (K0A) provided in manufacturer data sheets without accounting for elevated feed-viscosity when treating blood. The boundary layer model for mass transport across hollow fiber membranes, however, predicts an increase in mass transfer resistance (1/K0) and a decrease in K with increasing feed-viscosity.The effect of increased feed-side viscosity relative to baseline crystalloid viscosity on small solute K and 1/K0 was therefore examined in commercial high- (HF) and low-flux (LF) dialyzers in lab-bench studies using standard dialysis equipment in the normal operating range. Homogeneous colloid solutions and bovine plasma were used to simulate the range of relative viscosities (grel) and oncotic pressures expected under in-vivo conditions. Internal filtration (IF) was quantified by a mathematical model to obtain diffusive transport characteristics (K', 1/K00).An up to 5-fold increase in grel caused a small increase in K and a small decrease in 1/K0 in HF, but not in LF dialyzers. After correction for a small convective contribution by IF, K' and 1/K00 remained constant in both LF and HF dialyzers.Diffusive transport characteristics of commercial HF and LF dialyzers are independent of variable feed-side viscosity. This suggests an insignificant contribution of the feed-side boundary layer resistance in dialyzers optimized for operation in hemodialysis. Increasing the feed-side viscosity, however, increases the convective component of dialyzer solute transport because of IF. Diffusive dialyzer clearance predicted from the dialyzer K0A is independent of elevated feed-viscosity. (c) 2022 The Author(s). Published by Elsevier B.V. on behalf of Nalecz Institute of Biocybernetics and Biomedical Engineering of the Polish Academy of Sciences. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
引用
收藏
页码:1112 / 1122
页数:11
相关论文
共 6 条
  • [1] Determinants of Hemodialysis Performance:Modeling Fluid and Solute Transport in Hollow-Fiber Dialyzers
    Yu, Jian
    Chitalia, Vipul C.
    Akintewe, Olukemi O.
    Edwards, Aurelie
    Wong, Joyce Y.
    REGENERATIVE ENGINEERING AND TRANSLATIONAL MEDICINE, 2021, 7 (03) : 291 - 300
  • [2] Determinants of Hemodialysis Performance:Modeling Fluid and Solute Transport in Hollow-Fiber Dialyzers
    Jian Yu
    Vipul C. Chitalia
    Olukemi O. Akintewe
    Aurelie Edwards
    Joyce Y. Wong
    Regenerative Engineering and Translational Medicine, 2021, 7 : 291 - 300
  • [3] Diffusive transport across unconfined hollow fiber membranes
    Johnson, DW
    Semmens, MJ
    Gulliver, JS
    JOURNAL OF MEMBRANE SCIENCE, 1997, 128 (01) : 67 - 81
  • [4] A novel model of solute transport in a hollow-fiber bioartificial pancreas based on a finite element method
    Dulong, JL
    Legallais, C
    Darquy, S
    Reach, G
    BIOTECHNOLOGY AND BIOENGINEERING, 2002, 78 (05) : 576 - 582
  • [5] Analysis of Solute Transport by Diffusion and Convection in Hollow-fiber Hemodialyzer Using the Finite Element Method
    Kanamori, Toshiyuki
    Mizoguchi, Kensaku
    KAGAKU KOGAKU RONBUNSHU, 2011, 37 (02) : 91 - 95
  • [6] Oscillations with a long periodical time observed in solute transport by diffusion combined with convection through a single hollow-fiber membrane
    Kanamori, T
    Ohmori, T
    Yamaguchi, T
    Shinbo, T
    JOURNAL OF MEMBRANE SCIENCE, 2001, 184 (02) : 287 - 292