Comparisons of intraocular pressure measurements: Goldmann applanation tonometry, noncontact tonometry, Tono-Pen tonometry, and dynamic contour tonometry
被引:34
作者:
Hsu, S-Y
[1
,2
,3
]
Sheu, M-M
[1
,3
]
Hsu, A-H
[6
]
Wu, K-Y
[4
,5
]
Yeh, J-I
[7
]
Tien, J-N
[8
]
Tsai, R-K
[1
,3
]
机构:
[1] Tzu Chi Univ, Dept Ophthalmol, Buddhist Tzu Chi Gen Hosp, Hualien 970, Taiwan
[2] Tzu Chi Univ, Inst Med Sci, Hualien 970, Taiwan
[3] Tzu Chi Univ, Dept Ophthalmol & Visual Sci, Inst Med, Coll Med, Hualien 970, Taiwan
[4] Kaohsiung Med Univ, Dept Ophthalmol, Chung Ho Mem Hosp, Kaohsiung, Taiwan
[5] Kaohsiung Med Univ, Fac Med, Coll Med, Kaohsiung, Taiwan
[6] Buddhist Tzu Chi Gen Hosp, Dept Internal Med, Hualien, Taiwan
[7] Buddhist Tzu Chi Gen Hosp, Dept Family Med, Hualien, Taiwan
[8] Tzu Chi Univ, Branch Publ Hlth, Inst Med Sci, Coll Med, Hualien 970, Taiwan
来源:
关键词:
corneal curvature;
corneal thickness;
lens thickness;
ocular pulse amplitude;
tonometry;
CENTRAL CORNEAL THICKNESS;
OCULAR RESPONSE ANALYZER;
BIOMECHANICAL PROPERTIES;
PULSE AMPLITUDE;
AGE;
CURVATURE;
GLAUCOMA;
PATIENT;
EYES;
D O I:
10.1038/eye.2009.77
中图分类号:
R77 [眼科学];
学科分类号:
100212 ;
摘要:
Aims To compare intraocular pressure (IOP) readings between Tono-Pen tonometry and GAT, between noncontact tonometry (NCT) and GAT, and between dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT). The correlation between IOP reading and possible confounder was identified. Methods This observational cross-sectional study enrolled sixty-two healthy subjects. All IOP and ocular pulse amplitude (OPA) measurements were taken by a single ophthalmologist; mean keratometric power (MK), central corneal thickness (CCT), and lens thickness (LT) were measured by a single experienced technician. Results Stepwise multiple regression analysis indicated that GAT (P=0.017) and DCT (P=0.002) readings correlated positively with MK; GAT, NCT, and Tono-Pen readings correlated positively with CCT (P<0.05); NCT (P=0.035), and DCT (P=0.016) readings correlated negatively with LT; GAT (P=0.006) and Tono-Pen (P=0.009) readings correlated positively with OPA. Conclusions The K, CCT, LT, and OPA are confounders in tonometry readings. Eye (2009) 23, 1582-1588; doi:10.1038/eye.2009.77; published online 1 May 2009