An on-line pre-concentration system for determination of cadmium in drinking water using FAAS

被引:37
|
作者
dos Santos, Walter N. L. [1 ]
Costa, Jorge L. O.
Araujo, Rennan G. O.
de Jesus, Djane S.
Costa, Antonio C. S.
机构
[1] Univ Fed Bahia, Inst Quim, BR-40170290 Salvador, BA, Brazil
[2] Ctr Fed Educ Tecnol Bahia, BR-40300010 Salvador, BA, Brazil
关键词
cadmium determination; drinking water; FAAS; pre-concentration;
D O I
10.1016/j.jhazmat.2006.03.046
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In the present paper, a minicolumn of polyurethane foam loaded with 4-(2-pyridylazo)-resorcinol (PAR) is proposed as pre-concentration system for cadmium determination in drinking water samples by flame atomic absorption spectrometry. The optimization step was performed using two-level full factorial design and Doehlert matrix, involving the variables: sampling flow rate, elution concentration, buffer concentration and pH. Using the established experimental conditions in the optimization step of: pH 8.2, sampling flow rate 8.5 mL min(-1), buffer concentration 0.05 mol L-1 and elution concentration of 1.0 mol L-1, this system allows the determination of cadmium with detection limit (LD) (3 sigma/S) of 20.0ng L-1 and quantification limit (LQ) (10 sigma/S) of 64ngL(-1), precision expressed as relative standard deviation (R.S.D.) of 5.0 and 4.7% for cadmium concentration of 5.0 and 40.0 mu g L-1, respectively, and a pre-concentration factor of 158 for a sample volume of 20.0mL. The accuracy was confirmed by cadmium determination in the standard reference material, NIST SRM 1643d trace elements in natural water. This procedure was applied for cadmium determination in drinking water samples collected from Salvador City, Bahia, Brazil. For five samples analyzed, the achieved concentrations varied from 0.31 to 0.86 mu g L-1. (c) 2006 Published by Elsevier B.V.
引用
收藏
页码:1357 / 1361
页数:5
相关论文
共 50 条
  • [21] Liquid chromatographic analysis of milk phospholipids with on-line pre-concentration
    Fagan, P
    Wijesundera, C
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1054 (1-2) : 241 - 249
  • [22] Determination of benzoylureas in ground water samples by fully automated on-line pre-concentration and liquid chromatography-fluorescence detection
    García, MDG
    Galera, MM
    Martínez, DB
    Gallego, JG
    JOURNAL OF CHROMATOGRAPHY A, 2006, 1103 (02) : 271 - 277
  • [23] Determination of Hg in seawater by inductively coupled plasma mass spectrometry after on-line pre-concentration
    Seibert, EL
    Dressler, VL
    Pozebon, D
    Curtius, AJ
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2001, 56 (10) : 1963 - 1971
  • [24] Determination of cadmium and zinc in vegetables with online FAAS after simultaneous pre-concentration with 1,5-diphenylthiocarbazone immobilised on naphthalene
    Chandio, Zaheer A.
    Talpur, Farah N.
    Khan, Humaira
    Afridi, H. I.
    Khaskheli, Ghulam Quadir
    FOOD ADDITIVES AND CONTAMINANTS PART A-CHEMISTRY ANALYSIS CONTROL EXPOSURE & RISK ASSESSMENT, 2013, 30 (01): : 110 - 115
  • [25] Room Temperature Ionic Liquid-based Microextraction for Pre-concentration of Cadmium and Copper from Biological Samples and Determination by FAAS
    Kandhro, Ghulam A.
    Soylak, Mustafa
    Kazi, Tasneem Gul
    Yilmaz, Erkan
    Afridi, Hassan I.
    ATOMIC SPECTROSCOPY, 2012, 33 (05) : 166 - 172
  • [26] AN AUTOMATIC FLOW SYSTEM OF RAPID ON-LINE DIGESTION AND PRE-CONCENTRATION FOR IN-FIELD DETERMINATION OF TRACE TOTAL MERCURY IN SEAWATER
    Sun, Lumin
    Lin, Shanshan
    Feng, Lifeng
    Yuan, Dongxing
    ANALYTICAL LETTERS, 2012, 45 (10) : 1321 - 1331
  • [27] PRE-CONCENTRATION OF TRACE AMOUNT OF CADMIUM IN WATER BY FERRITE PROCESS
    KANZAKI, T
    TONOIKE, H
    KATSURA, T
    BUNSEKI KAGAKU, 1982, 31 (06) : E207 - E210
  • [28] Determination of Fe, Zn and Cu in ambient air by Combining pre-concentration Methods and FAAS
    Yaman, M.
    Erel, E.
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH, 2013, 7 (04) : 989 - 994
  • [29] Flow analysis by using solenoid valves for As(III) determination in natural waters by an on-line separation and pre-concentration system coupled to a tungsten coil atomizer
    Neira, JY
    Mendoza, J
    Bruhn, CG
    QUIMICA NOVA, 2005, 28 (02): : 217 - 223
  • [30] On-line coupling of derivatization with pre-concentration to determine trace levels of methotrexate
    Samy Emara
    Tsutomu Masujima
    Walaa Zarad
    Maha Kamal
    Ramzia EL-Bagary
    JournalofPharmaceuticalAnalysis, 2013, 3 (01) : 28 - 35