Highly wearable cuff-less blood pressure and heart rate monitoring with single-arm electrocardiogram and photoplethysmogram signals

被引:91
作者
Zhang, Qingxue [1 ]
Zhou, Dian [1 ,2 ]
Zeng, Xuan [2 ]
机构
[1] Univ Texas Dallas, Dept Elect Engn, 800 W Campbell Rd, Richardson, TX 75080 USA
[2] Fudan Univ, Dept Microelect, 220 Handan Rd, Shanghai 200433, Peoples R China
基金
中国国家自然科学基金;
关键词
Wearable computers; Blood pressure; Heart rate; Photoplethysmogram; Electrocardiography; Pulse transit time; Machine learning; RATE-VARIABILITY; ECG;
D O I
10.1186/s12938-017-0317-z
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Long-term continuous systolic blood pressure (SBP) and heart rate (HR) monitors are of tremendous value to medical (cardiovascular, circulatory and cerebrovascular management), wellness (emotional and stress tracking) and fitness (performance monitoring) applications, but face several major impediments, such as poor wearability, lack of widely accepted robust SBP models and insufficient proofing of the generalization ability of calibrated models. Methods: This paper proposes a wearable cuff-less electrocardiography (ECG) and photoplethysmogram (PPG)-based SBP and HR monitoring system and many efforts are made focusing on above challenges. Firstly, both ECG/PPG sensors are integrated into a single-arm band to provide a super wearability. A highly convenient but challenging single-lead configuration is proposed for weak single-arm-ECG acquisition, instead of placing the electrodes on the chest, or two wrists. Secondly, to identify heartbeats and estimate HR from the motion artifacts-sensitive weak arm-ECG, a machine learning-enabled framework is applied. Then ECG-PPG heartbeat pairs are determined for pulse transit time (PTT) measurement. Thirdly, a PTT&HR-SBP model is applied for SBP estimation, which is also compared with many PTT-SBP models to demonstrate the necessity to introduce HR information in model establishment. Fourthly, the fitted SBP models are further evaluated on the unseen data to illustrate the generalization ability. A customized hardware prototype was established and a dataset collected from ten volunteers was acquired to evaluate the proof-of-concept system. Results: The semi-customized prototype successfully acquired from the left upper arm the PPG signal, and the weak ECG signal, the amplitude of which is only around 10% of that of the chest-ECG. The HR estimation has a mean absolute error (MAE) and a root mean square error (RMSE) of only 0.21 and 1.20 beats per min, respectively. Through the comparative analysis, the PTT&HR-SBP models significantly outperform the PTT-SBP models. The testing performance is 1.63 +/- 4.44, 3.68, 4.71 mmHg in terms of mean error +/- standard deviation, MAE and RMSE, respectively, indicating a good generalization ability on the unseen fresh data. Conclusions: The proposed proof-of-concept system is highly wearable, and its robustness is thoroughly evaluated on different modeling strategies and also the unseen data, which are expected to contribute to long-term pervasive hypertension, heart health and fitness management.
引用
收藏
页数:20
相关论文
共 25 条
[11]   Analysis of heart rate variability in posttraumatic stress disorder patients in response to a trauma-related reminder [J].
Cohen, H ;
Kotler, M ;
Matar, MA ;
Kaplan, Z ;
Loewenthal, U ;
Miodownik, H ;
Cassuto, Y .
BIOLOGICAL PSYCHIATRY, 1998, 44 (10) :1054-1059
[12]  
Dotsinsky I., 2010, Bioautomation, V14, P129
[13]   Tremor suppression in ECG [J].
Dotsinsky, Ivan A. ;
Mihov, Georgy S. .
BIOMEDICAL ENGINEERING ONLINE, 2008, 7 (1)
[14]  
Fuke S, 2013, IEEE ENG MED BIO, P6107, DOI 10.1109/EMBC.2013.6610946
[15]   Feasibility of cuff-free measurement of systolic and diastolic arterial blood pressure [J].
Mase, Michela ;
Mattei, Walter ;
Cucino, Roberta ;
Faes, Luca ;
Nollo, Giandomenico .
JOURNAL OF ELECTROCARDIOLOGY, 2011, 44 (02) :201-207
[16]  
Mazaheri S., 2014, Smart Instrumentation, Measurement and Applications (ICSIMA), 2014 IEEE International Conference on, P1
[17]  
Mottaghi S., 2012, International Journal of Bioscience, Biochemistry and Bioinformatics, V2, P394, DOI DOI 10.7763/IJBBB.2012.V2.140
[18]   Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and Practice [J].
Mukkamala, Ramakrishna ;
Hahn, Jin-Oh ;
Inan, Omer T. ;
Mestha, Lalit K. ;
Kim, Chang-Sei ;
Toreyin, Hakan ;
Kyal, Survi .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (08) :1879-1901
[19]   Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressure [J].
Payne, RA ;
Symeonides, CN ;
Webb, DJ ;
Maxwell, SRJ .
JOURNAL OF APPLIED PHYSIOLOGY, 2006, 100 (01) :136-141
[20]   A Rapid and Accurate Extraction Procedure for Analysing Free Amino Acids in Meat Samples by GC-MS [J].
Perez-Palacios, Trinidad ;
Barroso, Miguel A. ;
Ruiz, Jorge ;
Antequera, Teresa .
INTERNATIONAL JOURNAL OF ANALYTICAL CHEMISTRY, 2015, 2015