Highly wearable cuff-less blood pressure and heart rate monitoring with single-arm electrocardiogram and photoplethysmogram signals

被引:91
作者
Zhang, Qingxue [1 ]
Zhou, Dian [1 ,2 ]
Zeng, Xuan [2 ]
机构
[1] Univ Texas Dallas, Dept Elect Engn, 800 W Campbell Rd, Richardson, TX 75080 USA
[2] Fudan Univ, Dept Microelect, 220 Handan Rd, Shanghai 200433, Peoples R China
基金
中国国家自然科学基金;
关键词
Wearable computers; Blood pressure; Heart rate; Photoplethysmogram; Electrocardiography; Pulse transit time; Machine learning; RATE-VARIABILITY; ECG;
D O I
10.1186/s12938-017-0317-z
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Long-term continuous systolic blood pressure (SBP) and heart rate (HR) monitors are of tremendous value to medical (cardiovascular, circulatory and cerebrovascular management), wellness (emotional and stress tracking) and fitness (performance monitoring) applications, but face several major impediments, such as poor wearability, lack of widely accepted robust SBP models and insufficient proofing of the generalization ability of calibrated models. Methods: This paper proposes a wearable cuff-less electrocardiography (ECG) and photoplethysmogram (PPG)-based SBP and HR monitoring system and many efforts are made focusing on above challenges. Firstly, both ECG/PPG sensors are integrated into a single-arm band to provide a super wearability. A highly convenient but challenging single-lead configuration is proposed for weak single-arm-ECG acquisition, instead of placing the electrodes on the chest, or two wrists. Secondly, to identify heartbeats and estimate HR from the motion artifacts-sensitive weak arm-ECG, a machine learning-enabled framework is applied. Then ECG-PPG heartbeat pairs are determined for pulse transit time (PTT) measurement. Thirdly, a PTT&HR-SBP model is applied for SBP estimation, which is also compared with many PTT-SBP models to demonstrate the necessity to introduce HR information in model establishment. Fourthly, the fitted SBP models are further evaluated on the unseen data to illustrate the generalization ability. A customized hardware prototype was established and a dataset collected from ten volunteers was acquired to evaluate the proof-of-concept system. Results: The semi-customized prototype successfully acquired from the left upper arm the PPG signal, and the weak ECG signal, the amplitude of which is only around 10% of that of the chest-ECG. The HR estimation has a mean absolute error (MAE) and a root mean square error (RMSE) of only 0.21 and 1.20 beats per min, respectively. Through the comparative analysis, the PTT&HR-SBP models significantly outperform the PTT-SBP models. The testing performance is 1.63 +/- 4.44, 3.68, 4.71 mmHg in terms of mean error +/- standard deviation, MAE and RMSE, respectively, indicating a good generalization ability on the unseen fresh data. Conclusions: The proposed proof-of-concept system is highly wearable, and its robustness is thoroughly evaluated on different modeling strategies and also the unseen data, which are expected to contribute to long-term pervasive hypertension, heart health and fitness management.
引用
收藏
页数:20
相关论文
共 25 条
[1]  
AAMI, 2007, SP101992 AAMI ANSIAA
[2]  
Altman DG, 2002, CLIN CHEM, V48, P801
[3]   Heart rate variability as an index of regulated emotional responding [J].
Appelhans, Bradley M. ;
Luecken, Linda J. .
REVIEW OF GENERAL PSYCHOLOGY, 2006, 10 (03) :229-240
[4]   Daily worry is related to low heart rate variability during waking and the subsequent nocturnal sleep period [J].
Brosschot, Jos F. ;
Van Dijk, Eduard ;
Thayer, Julian F. .
INTERNATIONAL JOURNAL OF PSYCHOPHYSIOLOGY, 2007, 63 (01) :39-47
[5]   A survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time [J].
Buxi, Dilpreet ;
Redoute, Jean-Michel ;
Yuce, Mehmet Rasit .
PHYSIOLOGICAL MEASUREMENT, 2015, 36 (03) :R1-R26
[6]  
Cattivelli FS, 2009, SIXTH INTERNATIONAL WORKSHOP ON WEARABLE AND IMPLANTABLE BODY SENSOR NETWORKS, PROCEEDINGS, P114, DOI [10.1109/BSN.2009.35, 10.1109/P3644.34]
[7]   LIBSVM: A Library for Support Vector Machines [J].
Chang, Chih-Chung ;
Lin, Chih-Jen .
ACM TRANSACTIONS ON INTELLIGENT SYSTEMS AND TECHNOLOGY, 2011, 2 (03)
[8]   Continuous estimation of systolic blood pressure using the pulse arrival time and intermittent calibration [J].
Chen, W ;
Kobayashi, T ;
Ichikawa, S ;
Takeuchi, Y ;
Togawa, T .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2000, 38 (05) :569-574
[9]  
Chi Yu Mike, 2010, IEEE Rev Biomed Eng, V3, P106, DOI 10.1109/RBME.2010.2084078
[10]   Signal quality indices and data fusion for determining clinical acceptability of electrocardiograms [J].
Clifford, G. D. ;
Behar, J. ;
Li, Q. ;
Rezek, I. .
PHYSIOLOGICAL MEASUREMENT, 2012, 33 (09) :1419-1433