Micromotion of implant-abutment interfaces (IAI) after loading: correlation of finite element analysis with in vitro performances

被引:15
|
作者
Li, Zhi [1 ,2 ]
Gao, Shanshan [1 ]
Chen, Hongyu [1 ]
Ma, Ruiyang [1 ]
Wu, Tingting [1 ]
Yu, Haiyang [1 ]
机构
[1] Sichuan Univ, State Key Lab Oral Dis, Natl Clin Res Ctr Oral Dis, West China Hosp Stomatol, Chengdu, Sichuan, Peoples R China
[2] Aerosp Ctr Hosp, Beijing, Peoples R China
基金
中国国家自然科学基金;
关键词
Dental implants; Finite element analysis; Microleakage; Micromotion; Locked condition; TITANIUM ABUTMENTS; BACTERIAL LEAKAGE; DENTAL IMPLANT; REMOVAL FORCE; CONNECTIONS; ZIRCONIA; COLONIZATION; MICROLEAKAGE; PARAMETERS; COMPONENTS;
D O I
10.1007/s11517-018-1937-6
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Micromotion between IAI affects long-term survival rate of dental implants. The use of practical implants for mechanical test is costly. Finite element analysis (FEA) could test the micron level deformation changes, but whether it reflects the in vitro mechanical performances remains unknown. This study aims to investigate the correlation between IAI micromotion of FEA and in vitro performances. The two-step-two-component FEA method was used to test the relative deformation between IAI for three implant designs (M1, M2, and M3) during torque loading and cyclic oblique loading. The micromotion was divided into directions that perpendicular to (x-axis) and parallel to (y-axis) IAI. In vitro experiments on the micromotion relevant performance of IAI microleakage (tested by toluidine blue releasing with a spectrometer) and IAI locked condition (tested by abutment removal force tests after detaching the central screws) were also conducted for the identical implant systems (G1, G2, and G3). One-way ANOVA and Pearson's correlation tests were performed for data analysis. FEA illustrated that the three implant systems performed different micromotion patterns. Significant differences were found in the IAI microleakage and removal force among the groups. Positive correlations were found between FEA and in vitro outcomes. Therefore, the two-step-two-component FEA method is an appropriate method to evaluate the IAI micromotion after loading.
引用
收藏
页码:1133 / 1144
页数:12
相关论文
共 50 条
  • [1] Micromotion of implant-abutment interfaces (IAI) after loading: correlation of finite element analysis with in vitro performances
    Zhi Li
    Shanshan Gao
    Hongyu Chen
    Ruiyang Ma
    Tingting Wu
    Haiyang Yu
    Medical & Biological Engineering & Computing, 2019, 57 : 1133 - 1144
  • [2] Effect of Cyclic Loading on Micromotion at the Implant-Abutment Interface
    Karl, Matthias
    Taylor, Thomas D.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2016, 31 (06) : 1292 - 1302
  • [3] Longitudinal Observation of Micromotion upon Loading of Implant-Abutment Connection
    Yamashita, Kohei
    Kataoka, Yu
    Munakata, Motohiro
    Yamaguchi, Kikue
    Hayashi, Myu
    Baba, Daisuke
    BIOENGINEERING-BASEL, 2024, 11 (06):
  • [4] Correlation Between Micromotion and Gap Formation at the Implant-Abutment Interface
    Grobecker-Karl, Tanja
    Karl, Matthias
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2017, 30 (02) : 150 - 152
  • [5] Implant-Abutment Misfit after Cyclic Loading: An In Vitro Experimental Study
    de Vasconcelos, John Eversong Lucena
    De Matos, Jefferson David Melo
    Queiroz, Daher Antonio
    Lopes, Guilherme da Rocha Scalzer
    De Lacerda, Bruna Caroline Goncalves Vasconcelos
    Bottino, Marco Antonio
    Turssi, Cecilia Pedroso
    Basting, Roberta Tarkany
    do Amaral, Flavia Lucisano Botelho
    Franca, Fabiana Mantovani Gomes
    MATERIALS, 2022, 15 (15)
  • [6] Nonlinear finite element analysis of three implant-abutment interface designs
    ChunBo Tang SiYu Liu GuoXing Zhou JinHua Yu GuangDong Zhang YiDong Bao and QiuJu Wang Dental Research Institute Nanjing Medical University Nanjing China and College of Mechanical and Electrical Engineering Nanjing University of Aeronautics and Astronautics Nanjing China
    International Journal of Oral Science, 2012, 4 (02) : 101 - 108
  • [7] Nonlinear finite element analysis of three implant-abutment interface designs
    Tang, Chun-Bo
    Liu, Si-Yu
    Zhou, Guo-Xing
    Yu, Jin-Hua
    Zhang, Guang-Dong
    Bao, Yi-Dong
    Wang, Qiu-Ju
    INTERNATIONAL JOURNAL OF ORAL SCIENCE, 2012, 4 (02) : 101 - 108
  • [9] Efficacy of a sealing material on the mechanical behaviors for various implant abutment interfaces (IAI) after fatigue loading
    Li, Z.
    Gao, S. S.
    Tan, X.
    Min, J.
    Yu, H. Y.
    BASIC & CLINICAL PHARMACOLOGY & TOXICOLOGY, 2018, 123 : 17 - 18
  • [10] Crestal and Subcrestal Placement of Morse Cone Implant-Abutment Connection Implants: An In Vitro Finite Element Analysis (FEA) Study
    Comuzzi, Luca
    Ceddia, Mario
    Di Pietro, Natalia
    Inchingolo, Francesco
    Inchingolo, Angelo Michele
    Romasco, Tea
    Tumedei, Margherita
    Specchiulli, Alessandro
    Piattelli, Adriano
    Trentadue, Bartolomeo
    BIOMEDICINES, 2023, 11 (11)