Microchannel plate operation at high count rates: New results

被引:43
作者
Tremsin, AS
Pearson, JF
Fraser, GW
Feller, WB
White, P
机构
[1] UNIV LEICESTER,DEPT PHYS & ASTRON,XRAY ASTRON GRP,LEICESTER LE1 7RH,LEICS,ENGLAND
[2] NOVA SCI INC,STURBRIDGE,MA 01566
基金
美国国家科学基金会;
关键词
D O I
10.1016/0168-9002(96)00482-2
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe an experimental study of two distinct microchannel plate (MCP) detector configurations with extended dynamic range. First, we report on measurements made with single MCPs of very low resistance (less than or equal to 500 k Omega) bonded to a Peltier cooler. Stable, reproducible operation for a rear-cooled MCP with 10 mu m diameter channels of length-to-diameter ratio 60:1 was achieved for biases up to 1320 V. At this point, the MCP strip current was 5.1 mA and the rate of internal heat generation, 0.78 Wcm(-2) A maximum count rate of 10(8) cm(-2)s(-1), limited by the available intensity of UV illumination, was observed, coupled to a dark noise count rate of less than 1 cm(-2)s(-1). Second, we describe a substantial (up to 20-fold) improvement in count rate capability for two-stage MCP multipliers incorporating plates with gold, rather than nichrome, electrodes. We also present, in support of the experimental study, a thermal model of bonded, conductively-cooled MCPs which we use to determine critical parameters for stable detector operation.
引用
收藏
页码:139 / 151
页数:13
相关论文
共 50 条
  • [41] Improvement of microchannel plate technology for the new generation of image converters
    Kanchiev, ZI
    Kutasov, VA
    Petrovskii, GT
    JOURNAL OF OPTICAL TECHNOLOGY, 2000, 67 (10) : 907 - 911
  • [42] Suppression of driving laser in high harmonic generation with a microchannel plate
    Zhang, Qi
    Zhao, Kun
    Li, Jie
    Chini, Michael
    Cheng, Yan
    Wu, Yi
    Cunningham, Eric
    Chang, Zenghu
    OPTICS LETTERS, 2014, 39 (12) : 3670 - 3673
  • [43] A high resolution imaging microchannel plate detector for EUV spectrometry
    Bannister, NP
    Lapington, JS
    Barstow, MA
    Fraser, GW
    Sanderson, BS
    Tandy, JA
    Pearson, JF
    Spragg, JE
    X-RAY AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY XI, 2000, 4140 : 199 - 210
  • [44] A new approach to modeling current amplification in the channel of a Microchannel plate
    Berkin, A. B.
    Vasil'ev, V. V.
    TECHNICAL PHYSICS LETTERS, 2007, 33 (08) : 664 - 666
  • [45] A simple power supply and control unit for pulsed operation of a microchannel plate imaging detector
    J Upadhyay
    H R Bundel
    R Chandra
    J A Chakera
    C P Navathe
    P D Gupta
    Sadhana, 1999, 24 : 235 - 243
  • [46] A new medium for determining the total plate count in food
    Smith, CF
    Townsend, DE
    JOURNAL OF FOOD PROTECTION, 1999, 62 (12) : 1404 - 1410
  • [47] Control Issues of High Pole Count Machines at High Speed Operation
    Yim, Jung-Sik
    Sul, Seung-Ki
    Bae, Bon-Ho
    Patel, Nitin
    Hiti, Silva
    CONFERENCE RECORD OF THE 2007 IEEE INDUSTRY APPLICATIONS CONFERENCE FORTY-SECOND IAS ANNUAL MEETING, VOLS. 1-5, 2007, : 2045 - +
  • [48] AVALANCHE TRANSISTOR PULSER FOR FAST-GATED OPERATION OF MICROCHANNEL PLATE IMAGE-INTENSIFIERS
    LUNDY, A
    PARKER, JR
    LUNSFORD, JS
    MARTIN, AD
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1978, 25 (01) : 591 - 597
  • [49] LOW-NOISE PREAMPLIFIER FOR HIGH COUNT RATES
    KURZ, R
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1972, NS19 (03) : 418 - &
  • [50] Non-linearity of APDs at high count rates
    Gran, M
    Pike, ER
    Pailharey, E
    LIGHT SCATTERING AND PHOTON CORRELATION SPECTROSCOPY, 1997, 40 : 117 - 127