Dose verification for liver target volumes undergoing respiratory motion

被引:1
|
作者
Dyce, Emma [1 ]
Cutajar, Dean [2 ]
Metcalfe, Peter [2 ]
Downes, Simon [1 ]
机构
[1] Prince Wales Hosp, Nelune Comprehens Canc Ctr, Sydney, NSW, Australia
[2] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW, Australia
关键词
Liver; 4D; IGRT; Respiratory motion; IMRT; OPTIMIZATION; DELIVERY; IMPACT; SBRT;
D O I
10.1007/s13246-019-00737-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Respiratory motion has a significant impact on dose delivered to abdominal targets during radiotherapy treatment. Accurate treatment of liver tumours adjacent to the diaphragm is complicated by large respiratory movement, as well as differing tissue densities at the lung-liver interface. This study aims to evaluate the accuracy of dose delivered to superior liver tumours using passive respiratory monitoring, in the absence of gating technology, for a range of treatment techniques. An in-house respiratory phantom was designed and constructed to simulate the lung and liver anatomy. The phantom consisted of adjacent slabs of lung and liver equivalent materials and a cam drive system to emulate respiratory motion. A CC04 ionisation chamber and Gafchromic EBT3 film were used to perform point dose and dose plane measurements respectively. Plans were calculated using an Elekta Monaco treatment planning system (TPS) on exhale phase study sets for conformal, volume modulated arc therapy (VMAT) and intensity modulated radiation therapy (IMRT) techniques, with breathing rates of 8, 14 and 23bpm. Analysis confirmed the conformal delivery protocol currently used for this site within the department is suitable. The experiments also determined that VMAT is a viable alternative technique for treatment of superior liver lesions undergoing respiratory motion and was superior to IMRT. Furthermore, the measurements highlighted the need for respiratory management in these cases. Displacements due to respiration exceeding planned margins could result in reduced coverage of the clinical target volume and much higher doses to the lung than anticipated.
引用
收藏
页码:619 / 626
页数:8
相关论文
共 50 条
  • [11] Development of a liver respiratory motion simulator to investigate magnetic tracking for abdominal interventions
    Cleary, K
    Banovac, F
    Levy, E
    Tanaka, D
    MEDICAL IMAGING 2002: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2002, 4681 : 25 - 29
  • [12] Segmental analysis of respiratory liver motion in patients with and without a history of abdominal surgery
    Shimizu, Yasuhiro
    Takamatsu, Shigeyuki
    Yamamoto, Kazutaka
    Maeda, Yoshikazu
    Sasaki, Makoto
    Tamamura, Hiroyasu
    Bou, Sayuri
    Kumano, Tomoyasu
    Gabata, Toshifumi
    JAPANESE JOURNAL OF RADIOLOGY, 2018, 36 (08) : 511 - 518
  • [13] Segmental analysis of respiratory liver motion in patients with and without a history of abdominal surgery
    Yasuhiro Shimizu
    Shigeyuki Takamatsu
    Kazutaka Yamamoto
    Yoshikazu Maeda
    Makoto Sasaki
    Hiroyasu Tamamura
    Sayuri Bou
    Tomoyasu Kumano
    Toshifumi Gabata
    Japanese Journal of Radiology, 2018, 36 : 511 - 518
  • [14] Kilovoltage intrafraction motion monitoring and target dose reconstruction for stereotactic volumetric modulated arc therapy of tumors in the liver
    Poulsen, Per Rugaard
    Worm, Esben S.
    Petersen, Jorgen B. B.
    Grau, Cai
    Fledelius, Walther
    Hoyer, Morten
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 (03) : 424 - 430
  • [15] Impact of motion velocity on four-dimensional target volumes: A phantom study
    Nakamura, Mitsuhiro
    Narita, Yuichiro
    Sawada, Akira
    Matsugi, Kiyotomo
    Nakata, Manabu
    Matsuo, Yukinori
    Mizowaki, Takashi
    Hiraoka, Masahiro
    MEDICAL PHYSICS, 2009, 36 (05) : 1610 - 1617
  • [16] Dosimetric and radiobiological impact of dose fractionation on respiratory motion induced IMRT delivery errors: A volumetric dose measurement study
    Duan, J
    Shen, S
    Fiveash, JB
    Popple, RA
    Brezovich, IA
    MEDICAL PHYSICS, 2006, 33 (05) : 1380 - 1387
  • [17] Study on relation between location of target volume and respiratory motion
    Wang Suli
    Zheng Feng
    Feng Beibei
    Proceedings of the First International Symposium on Test Automation & Instrumentation, Vols 1 - 3, 2006, : 213 - 215
  • [18] Respiratory motion control for stereotactic and robotic liver interventions
    Widmann, Gerlig
    Schullian, Peter
    Haidu, Marion
    Wiedermann, Franz J.
    Bale, Reto
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2010, 6 (03): : 343 - 349
  • [19] Respiratory motion analysis: Towards gated augmentation of the liver
    Olbrich, B
    Traub, J
    Wiesner, S
    Wichert, A
    Feussner, H
    Navab, N
    CARS 2005: Computer Assisted Radiology and Surgery, 2005, 1281 : 248 - 253
  • [20] Total dose, fraction dose and respiratory motion management impact adrenal SBRT outcome
    Haisraely, Ory
    Weiss, Ilana
    Jaffe, Marcia
    Appel, Sarit
    Person-Kaidar, Orit
    Symon, Zvi
    Ben-Ayun, Maoz
    Dubinski, Sergi
    Lawrence, Yaacov
    CLINICAL AND TRANSLATIONAL RADIATION ONCOLOGY, 2024, 47