A method to assess image quality for Low-dose PET: analysis of SNR, CNR, bias and image noise

被引:72
|
作者
Yan, Jianhua [1 ,2 ,3 ]
Schaefferkoetter, Josh [3 ,4 ]
Conti, Maurizio [5 ]
Townsend, David [3 ,4 ]
机构
[1] Shanxi Med Univ, Hosp 1, Dept Nucl Med, 85 Jiefang S Rd, Taiyuan 030001, Shanxi, Peoples R China
[2] Shanxi Med Univ, Mol Imaging Precis Med Collaborat Innovat Ctr, 85 Jiefang S Rd, Taiyuan 030001, Shanxi, Peoples R China
[3] A STAR NUS, Ctr Translat Med, Clin Imaging Res Ctr, 14 Med Dr B1-01, Singapore 17599, Singapore
[4] Natl Univ Singapore Hosp, Dept Diagnost Radiol, Main Bldg 5,Lower Kent Ridge Rd, Singapore 119074, Singapore
[5] Siemens Healthcare Mol Imaging, 810 Innovat Dr, Knoxville, TN 37932 USA
来源
CANCER IMAGING | 2016年 / 16卷
基金
中国国家自然科学基金;
关键词
Low dose; PET/MR; PET/CT; Lung; Image quality; LUNG-CANCER; COMPUTED-TOMOGRAPHY; CT; RECONSTRUCTION; PERFORMANCE; MANAGEMENT; SCANNER;
D O I
10.1186/s40644-016-0086-0
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Lowering injected dose will have an effect on PET image quality. In this article, we aim to investigate this effect in terms of signal-to-noise ratio (SNR) in the liver, contrast-to-noise ratio (CNR) in the lesion, bias and ensemble image noise. Methods: We present here our method and preliminary results using tuberculosis (TB) cases. Sixteen patients who underwent F-18-FDG PET/MR scans covering the whole lung and portion of the liver were selected for the study. Reduced doses were simulated by randomly discarding events in the PET list mode data stream, and ten realizations at each simulated dose were generated and reconstructed. The volumes of interest (VOI) were delineated on the image reconstructed from the original full statistics data for each patient. Four thresholds (20, 40, 60 and 80 % of SUVmax) were used to quantify the effect of the threshold on CNR at the different count level. Image metrics were calculated for each VOI. This experiment allowed us to quantify the loss of SNR and CNR as a function of the counts in the scan, in turn related to dose injected. Reproducibility of mean and maximum standardized uptake value (SUVmean and SUVmax) measurement in the lesions was studied as standard deviation across 10 realizations. Results: At 5 x 10(6) counts in the scan, the average SNR in the liver in the observed samples is about 3, and the CNR is reduced to 60 % of the full statistics value. The CNR in the lesion and SNR in the liver decreased with reducing count data. The variation of CNR across the four thresholds does not significantly change until the count level of 5 x 106. After correcting the factor related to subject's weight, the square of the SNR in the liver was found to have a very good linear relationship with detected counts. Some quantitative bias appears with count reduction. At the count level of 5 x 10(6), bias and noise in terms of SUVmean and SUVmax are up to 10 and 20 %, respectively. To keep both bias and noise less than 10 %, 5 x 10(6) counts and 20 x 10(6) counts were required for SUVmean and SUVmax, respectively. Conclusions: Initial results with the given data of 16 patients diagnosed as TB demonstrated that 5 x 10(6) counts in the scan could be sufficient to yield good images in terms of SNR, CNR, bias and noise. In the future, more work needs to be done to validate the proposed method with a larger population and lung cancer patient data.
引用
收藏
页数:12
相关论文
共 50 条
  • [31] A new method to evaluate the PET image quality based on the image noise and reproducibility of recovery coefficients
    Akamatsu, Go
    Taniguchi, Takafumi
    Kidera, Daisuke
    Kihara, Ken
    Mikasa, Shohei
    Komiya, Isao
    Sasaki, Masayuki
    JOURNAL OF NUCLEAR MEDICINE, 2013, 54
  • [32] A Low-Dose CT Image Denoising Method Based on Generative Adversarial Network and Noise Level Estimation
    Zhang Xiong
    Yang Linlin
    Hong, Shangguan
    Han Zefang
    Han Xinglong
    Wang Anhong
    Cui Xueying
    JOURNAL OF ELECTRONICS & INFORMATION TECHNOLOGY, 2021, 43 (08) : 2404 - 2413
  • [33] A Hybrid Neural Network Architecture to improve low-dose PET Image Reconstruction
    Kaviani, S.
    Sanaat, A.
    Mokri, M.
    Cohalan, C.
    Carrier, J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (SUPPL 1) : S100 - S100
  • [34] Study of low-dose PET image recovery using supervised learning with CycleGAN
    Zhao, Kui
    Zhou, Long
    Gao, Size
    Wang, Xiaozhuang
    Wang, Yaofa
    Zhao, Xin
    Wang, Huatao
    Liu, Kanfeng
    Zhu, Yunqi
    Ye, Hongwei
    PLOS ONE, 2020, 15 (09):
  • [35] Image Quality Measurement for Low-Dose Human Lung CT Scans
    Bognar, Gergo
    2015 38TH INTERNATIONAL CONFERENCE ON TELECOMMUNICATIONS AND SIGNAL PROCESSING (TSP), 2015, : 587 - 591
  • [36] Effects of sparse sampling schemes on image quality in low-dose CT
    Abbas, Sajid
    Lee, Taewon
    Shin, Sukyoung
    Lee, Rena
    Cho, Seungryong
    MEDICAL PHYSICS, 2013, 40 (11)
  • [37] Low-dose CT image restoration based on noise prior regression network
    Yan Jin
    Zhiwei Jiang
    Mengjia Huang
    Zhizhong Xue
    The Visual Computer, 2023, 39 : 459 - 471
  • [38] QUANTUM NOISE-REDUCTION OF LOW-DOSE IMAGE RECORDED ON IMAGING PLATE
    ISHIKAWA, A
    SUDA, H
    FUKAMI, A
    OIKAWA, T
    ARAI, Y
    JOURNAL OF ELECTRON MICROSCOPY, 1990, 39 (04): : 324 - 324
  • [39] Low-dose CT image restoration based on noise prior regression network
    Jin, Yan
    Jiang, Zhiwei
    Huang, Mengjia
    Xue, Zhizhong
    VISUAL COMPUTER, 2023, 39 (01): : 459 - 471
  • [40] ADVERSARIAL EM FOR PARTIALLY-SUPERVISED IMAGE-QUALITY ENHANCEMENT: APPLICATION TO LOW-DOSE PET IMAGING
    Sharma, Vatsala
    Awate, Suyash P.
    2024 IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING, ICIP, 2024, : 4007 - 4013