Performance of 14 rubella IgG immunoassays on samples with low positive or negative haemagglutination inhibition results

被引:19
作者
Huzly, Daniela [1 ]
Hanselmann, Ingeborg [1 ]
Neumann-Haefelin, Dieter [1 ]
Panning, Marcus [1 ]
机构
[1] Univ Med Ctr Freiburg, Inst Virol, Freiburg, Germany
关键词
Rubella; CRS; Assay comparison; Seroepidemiologic studies; LINKED-IMMUNOSORBENT-ASSAY; IMMUNOGLOBULIN-G; INTERNATIONAL UNITS; RADIAL HEMOLYSIS; HEPATITIS-B; VIRUS; ANTIBODY; PERSISTENCE; RESPONSES; IMMUNITY;
D O I
10.1016/j.jcv.2015.11.022
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Background: Rubella IgG testing is routinely done in prenatal care and seroepidemiological studies. Recently concern was raised that seropositivity rates were decreasing questioning vaccination policies. Manufacturers of rubella IgG assays and authors of seroepidemiological studies use different cut-offs for the definition of seropositivity. As rubella virus circulation is reduced since many years, seronegativity rates might be overestimated using an inappropriate cut-off. Objectives: Using different cut-off definitions we compared fourteen current rubella IgG immunoassays for sensitivity and qualitative result concordance in samples with low positive or negative haemagglutination inhibition (HI) titre. Study design: 150 clinical samples from patients and health care workers were included in the study. All samples were measured in 14 different rubella IgG immunoassays. Seropositivity was defined using recombinant rubella IgG immunoblot as reference standard. Results: The concordance of qualitative results using the manufacturers cut-off definitions was 56.4% if grey-zone results were analysed separately and 69.8% if grey-zone results were defined as positive. Using universal cut-offs of 10 IU/ml or 15 IU/ml the concordance was 70% and 61.4% respectively. Using the different cut-off definitions up to 71 out of the 124 immunoblot-positive samples tested negative in the immunoassays. The mean coefficient of variation (CV) of quantitative results in positive samples was 51% (range 19-113%). Conclusions: Determination of rubella immunity by measurement of rubella-IgG in a population with high vaccination coverage with current assays leads to a high number of false negative results. The value of routine rubella antibody testing in countries with high vaccination coverage should be discussed. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:13 / 18
页数:6
相关论文
共 35 条
[11]   Antenatal screening for hepatitis B and antibodies to Toxoplasma gondii and rubella virus:: Evaluation of two commercial immunoassay systems [J].
Diepersloot, RJA ;
Dunnewold-Hoekstra, H ;
Kruit-Den Hollander, J ;
Vlaspolder, F .
CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, 2001, 8 (04) :785-787
[12]   MULTICENTER EVALUATION OF 5 COMMERCIAL RUBELLA-VIRUS IMMUNOGLOBULIN-G KITS WHICH REPORT IN INTERNATIONAL UNITS PER MILLILITER [J].
DIMECH, W ;
BETTOLI, A ;
ECKERT, D ;
FRANCIS, B ;
HAMBLIN, J ;
KERR, T ;
RYAN, C ;
SKURRIE, I .
JOURNAL OF CLINICAL MICROBIOLOGY, 1992, 30 (03) :633-641
[13]   Evaluation of eight anti-rubella virus immunoglobulin G immunoassays that report results in international units per milliliter [J].
Dimech, Wayne ;
Panagiotopoulos, Lena ;
Francis, Barbara ;
Laven, Nicholas ;
Marler, Joan ;
Dickeson, David ;
Panayotou, Tony ;
Wilson, Kim ;
Wootten, Robyn ;
Dax, Elizabeth M. .
JOURNAL OF CLINICAL MICROBIOLOGY, 2008, 46 (06) :1955-1960
[14]   Investigation into Low-Level Anti-Rubella Virus IgG Results Reported by Commercial Immunoassays [J].
Dimech, Wayne ;
Arachchi, Nilukshi ;
Cai, Jingjing ;
Sahin, Terri ;
Wilson, Kim .
CLINICAL AND VACCINE IMMUNOLOGY, 2013, 20 (02) :255-261
[15]   Performance of the Elecsys Rubella IgG Assay in the Diagnostic Laboratory Setting for Assessment of Immune Status [J].
Enders, Martin ;
Bartelt, Uwe ;
Knotek, Frank ;
Bunn, Kristina ;
Strobel, Sirpa ;
Dietz, Klaus ;
Enders, Gisela .
CLINICAL AND VACCINE IMMUNOLOGY, 2013, 20 (03) :420-426
[16]   A Qualitative and Quantitative Comparison of Two Rubella Virus-Specific IgG Antibody Immunoassays [J].
Greenwood, Nicholas P. ;
Ovsyannikova, Inna G. ;
Vierkant, Robert A. ;
O'Byrne, Megan M. ;
Poland, Gregory A. .
VIRAL IMMUNOLOGY, 2010, 23 (04) :353-357
[17]   Rubella monitoring in pregnancy as a means for evaluating a possible reemergence of rubella [J].
Hutton, Jill ;
Rowan, Paul ;
Greisinger, Anthony ;
Mouzoon, Melanie .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2014, 211 (05)
[18]   Comparison of nine commercially available assays for quantification of antibody response to hepatitis B virus surface antigen [J].
Huzly, D. ;
Schenk, T. ;
Jilg, W. ;
Neumann-Haefelin, D. .
JOURNAL OF CLINICAL MICROBIOLOGY, 2008, 46 (04) :1298-1306
[19]   RUBELLA ANTIBODY MEASURED BY RADIAL HEMOLYSIS - CHARACTERISTICS AND PERFORMANCE OF A SIMPLE SCREENING METHOD FOR USE IN DIAGNOSTIC LABORATORIES [J].
KURTZ, JB ;
MORTIMER, PP ;
MORTIMER, PR ;
MORGANCAPNER, P ;
SHAFI, MS ;
WHITE, GBB .
JOURNAL OF HYGIENE, 1980, 84 (02) :213-&
[20]   Persistence and Titer Changes of Rubella Virus Antibodies in Primiparous Women Who Had Been Vaccinated with Strain RA 27/3 in Junior High School [J].
Lin, Ching-Chiang ;
Yang, Chun-Yuh ;
Shih, Yung-Luen ;
Huang, Yang-Yang ;
Yang, Tsung-Han ;
Liang, Jin-Yuan ;
Chang, Chu-Fen ;
Hsieh, Hsiu-Shu ;
Huang, Yeou-Lih .
CLINICAL AND VACCINE IMMUNOLOGY, 2012, 19 (01) :1-4