Piezocomposite and CMUT Arrays Assessment Through In Vitro Imaging Performances

被引:26
|
作者
Legros, M. [1 ]
Meynier, C. [1 ,2 ]
Dufait, R. [1 ]
Ferin, G. [1 ]
Tranquart, F. [2 ]
机构
[1] Vermon SA, 180 Rue Gen Renault, Tours, France
[2] INSERM U619, CHRU Bretonneau, Tours, France
来源
2008 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4 AND APPENDIX | 2008年
关键词
Quantitative image assessment; CMUT; piezocomposite based transducer;
D O I
10.1109/ULTSYM.2008.0275
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Piezoelectric and Capacitive Micromachined Ultrasound Transducers (CMUT) are usually measured and compared in regards to acoustic and electro-acoustic performances. This paper is focused on the imaging performances of such transducers and propose a quantitative imaging assessment of B-mode images. In this purpose, fully integrated CMUT and piezocomposite-based probes were manufactured Transducers were designed with close features (geometries, center frequency, interconnect and packaging) and plug on a clinical ultrasound system with a research interface. Major imaging performances (Speckle to Noise Ratio, Resolutions and contrast) of the probes are presented. Despite an environment dedicated to piezoelectric transducers, CMUT probe exhibits comparable image quality as compared to a state-of-the-an piezocomposite probe. Indeed, close resolutions are calculated, field of view is improved in phased-array imaging, and contrast is significantly improved.
引用
收藏
页码:1142 / +
页数:2
相关论文
共 11 条
  • [11] Two-dimensional capacitive micromachined ultrasonic transducer (CMUT) arrays for a miniature integrated volumetric ultrasonic imaging system
    Zhuang, X
    Wygant, IO
    Yeh, DT
    Nikoozadeh, A
    Oralkan, O
    Ergun, AS
    Cheng, CH
    Huang, Y
    Yaralioglu, GG
    Khuri-Yakub, BT