Filtering mechanical chest compression artefacts from out-of-hospital cardiac arrest data

被引:15
作者
Aramendi, E. [1 ]
Irusta, U. [1 ]
Ayala, U. [1 ]
Naas, H. [2 ,3 ,4 ]
Kramer-Johansen, J. [2 ,3 ,4 ]
Eftestol, T. [5 ]
机构
[1] Univ Basque Country UPV EHU, Commun Engn Dept, Bilbao 48013, Spain
[2] Oslo Univ Hosp, Natl Advisory Unit Prehosp Emergency Med NAKOS, N-0424 Oslo, Norway
[3] Oslo Univ Hosp, Dept Anaesthesiol, N-0424 Oslo, Norway
[4] Univ Oslo, N-0424 Oslo, Norway
[5] Univ Stavanger, Fac Sci & Technol, Dept Elect Engn & Comp Sci, N-4036 Stavanger, Norway
关键词
Cardiopulmonary resuscitation; Rhythm analysis; Mechanical chest compression; Cardiac arrest; EXTERNAL DEFIBRILLATOR ALGORITHM; CARDIOPULMONARY-RESUSCITATION; CPR ARTIFACTS; CIRCULATION; SPECIFICITY; SENSITIVITY; QUALITY; SUPPRESSION; RHYTHMS; FORCE;
D O I
10.1016/j.resuscitation.2015.10.012
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Aim: Filtering techniques to remove manual compression artefacts from the ECG have not been incorporated to defibrillators to diagnose the rhythm during cardiopulmonary resuscitation. Mechanical and manual compression artefacts may be very different. The aim of this study is to characterize the compression artefact caused by the LUCAS 2 device and to evaluate whether filtering the LUCAS 2 artefact results in an accurate rhythm analysis. Methods: A dataset of 1045 segments were obtained from 230 out-of-hospital cardiac arrest (OHCA) patients after LUCAS 2 activation. Rhythms were 201 shockable, 270 asystole and 574 organized. Segments during asystole were used to characterize the artefact in time and frequency domains. Three filtering methods, a comb filter and two adaptive filters, were used to remove the mechanical compression artefact. The filtered ECG was then diagnosed with a shock decision algorithm from a defibrillator. Results: When compared to the manual compression artefact, the LUCAS 2 artefact presented a similar amplitude (1.2 mV, p-value 0.26), fixed frequency (101.7 min(-1)), more harmonic components, smaller spectral dispersion, and a more regular waveform (p-val <3 x10(-7)). The sensitivity (SE) and specificity (SP) before filtering the LUCAS 2 artefact were 52.8% (90% low CI, 46.0%) and 81.5% (79.0%), respectively. For the best filter, SE and SP after filtering were 97.9% (95.7%) and 84.1% (82.0%), respectively. Optimal filters require more harmonics and smaller bandwidths than for manual compressions. Conclusion: Filtering resulted in a large increase in SE and small increase in SP. Despite differences in artefact characteristics between manual and mechanical compressions, filtering the LUCAS 2 compression artefact results in SE/SP values comparable to those obtained for manual compression artefacts. The SP is still below the 95% recommended by the American Heart Association. (C) 2015 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:41 / 47
页数:7
相关论文
共 35 条
[11]   Mechanical chest compression for out of hospital cardiac arrest: Systematic review and meta-analysis [J].
Gates, Simon ;
Quinn, Tom ;
Deakin, Charles D. ;
Blair, Laura ;
Couper, Keith ;
Perkins, Gavin D. .
RESUSCITATION, 2015, 94 :91-97
[12]   Continuous mechanical chest compressions during cardiac arrest to facilitate restoration of coronary circulation with percutaneous coronary intervention [J].
Grogaard, Haakon Kiil ;
Wik, Lars ;
Eriksen, Morten ;
Brekke, Magne ;
Sunde, Kjetil .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 50 (11) :1093-1094
[13]   The R Package geepack for Generalized Estimating Equations [J].
Halekoh, U ;
Hojsgaard, S ;
Yan, J .
JOURNAL OF STATISTICAL SOFTWARE, 2006, 15 (02) :1-11
[14]   A high-temporal resolution algorithm to discriminate shockable from nonshockable rhythms in adults and children [J].
Irusta, Unai ;
Ruiz, Jesus ;
Aramendi, Elisabete ;
Ruiz de Gauna, Sofia ;
Ayala, Unai ;
Alonso, Erik .
RESUSCITATION, 2012, 83 (09) :1090-1097
[15]   A Least Mean-Square Filter for the Estimation of the Cardiopulmonary Resuscitation Artifact Based on the Frequency of the Compressions [J].
Irusta, Unai ;
Ruiz, Jesus ;
Ruiz de Gauna, Sofia ;
Eftestol, Trygve ;
Kramer-Johansen, Jo .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2009, 56 (04) :1052-1062
[16]  
Kerber RE, 1997, CIRCULATION, V95, P1677
[17]   Shock Advisory System for Heart Rhythm Analysis During Cardiopulmonary Resuscitation Using a Single ECG Input of Automated External Defibrillators [J].
Krasteva, Vessela ;
Jekova, Irena ;
Dotsinsky, Ivan ;
Didon, Jean-Philippe .
ANNALS OF BIOMEDICAL ENGINEERING, 2010, 38 (04) :1326-1336
[18]   Reducing CPR artefacts in ventricular fibrillation in vitro [J].
Langhelle, A ;
Eftestol, T ;
Myklebust, H ;
Eriksen, M ;
Holten, BT ;
Steen, PA .
RESUSCITATION, 2001, 48 (03) :279-291
[19]   Identifying potentially shockable rhythms without interrupting cardiopulmonary resuscitation [J].
Li, Yongqin ;
Bisera, Joe ;
Geheb, Fredrick ;
Tang, Wanchun ;
Weil, Max Harry .
CRITICAL CARE MEDICINE, 2008, 36 (01) :198-203
[20]   Techniques for artefact filtering from chest compression corrupted ECG signals: Good, but not enough [J].
Li, Yongqin ;
Tang, Wanchun .
RESUSCITATION, 2009, 80 (11) :1219-1220