An Integrated Instrumentation System for Velocity, Concentration and Mass Flow Rate Measurement of Solid Particles Based on Electrostatic and Capacitance Sensors

被引:24
作者
Li, Jian [1 ]
Kong, Ming [2 ]
Xu, Chuanlong [1 ]
Wang, Shimin [1 ]
Fan, Ying [3 ]
机构
[1] Southeast Univ, Sch Energy & Environm, Key Lab Energy Thermal Convers & Control, Minist Educ, Nanjing 210096, Jiangsu, Peoples R China
[2] China Jiliang Univ, Coll Metrol & Measurement Engn, Hangzhou 310018, Zhejiang, Peoples R China
[3] Lanzhou Univ Technol, Sch Energy & Power Engn, Lanzhou 730050, Peoples R China
基金
中国国家自然科学基金;
关键词
integrated instrumentation system; electrostatic sensor; capacitance sensor; particle velocity; concentration; solid mass flow rate; PNEUMATICALLY CONVEYED PARTICLES; PULVERIZED FUEL; TOMOGRAPHY;
D O I
10.3390/s151229843
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The online and continuous measurement of velocity, concentration and mass flow rate of pneumatically conveyed solid particles for the high-efficiency utilization of energy and raw materials has become increasingly significant. In this paper, an integrated instrumentation system for the velocity, concentration and mass flow rate measurement of dense phase pneumatically conveyed solid particles based on electrostatic and capacitance sensorsis developed. The electrostatic sensors are used for particle mean velocity measurement in combination with the cross-correlation technique, while the capacitance sensor with helical surface-plate electrodes, which has relatively homogeneous sensitivity distribution, is employed for the measurement of particle concentration and its capacitance is measured by an electrostatic-immune AC-based circuit. The solid mass flow rate can be further calculated from the measured velocity and concentration. The developed instrumentation system for velocity and concentration measurement is verified and calibrated on a pulley rig and through static experiments, respectively. Finally the system is evaluated with glass beads on a gravity-fed rig. The experimental results demonstrate that the system is capable of the accurate solid mass flow rate measurement, and the relative error is within -3%-8% for glass bead mass flow rates ranging from 0.13 kg/s to 0.9 kg/s.
引用
收藏
页码:31023 / 31035
页数:13
相关论文
共 29 条
[1]   THE USE OF CAPACITANCE SENSORS FOR PHASE PERCENTAGE DETERMINATION IN MULTIPHASE PIPELINES [J].
ABOUELWAFA, MSA ;
KENDALL, EJM .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1980, 29 (01) :24-27
[2]   Flow regime determination in horizontal pneumatic transport of fine powders using non-intrusive acoustic probes [J].
Albion, Katherine ;
Briens, Lauren ;
Briens, Cedric ;
Berruti, Franco .
POWDER TECHNOLOGY, 2007, 172 (03) :157-166
[3]  
Arko A., 1999, 1st World Congress on Industrial Process Tomography, P140
[4]   A parallel-beam radiometric instrumentation system for the mass flow measurement of pneumatically conveyed solids [J].
Barratt, IR ;
Yan, Y ;
Byrne, B .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2001, 12 (09) :1515-1528
[5]   In-line measurement of pneumatically conveyed particles by a light transmission fluctuation method [J].
Cai, XS ;
Li, JF ;
Xin, OY ;
Zhao, ZJ ;
Su, MX .
FLOW MEASUREMENT AND INSTRUMENTATION, 2005, 16 (05) :315-320
[6]   An instrumentation system using combined sensing strategies for online mass flow rate measurement and particle sizing [J].
Carter, RM ;
Yan, Y .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2005, 54 (04) :1433-1437
[7]   Void fraction measurements in gas-liquid flows under 1-g and mu-g conditions using capacitance sensors [J].
Elkow, KJ ;
Rezkallah, KS .
INTERNATIONAL JOURNAL OF MULTIPHASE FLOW, 1997, 23 (05) :815-829
[8]   Electrostatic nonintrusive method for measuring the electric charge, mass flow rate, and velocity of particulates in the two-phase gas-solid pipe flows - Its only or as many as 50 years of historical evolution [J].
Gajewski, Juliusz B. .
IEEE TRANSACTIONS ON INDUSTRY APPLICATIONS, 2008, 44 (05) :1418-1430
[9]   A novel capacitive system for the concentration measurement of pneumatically conveyed pulverized fuel at power stations [J].
Hu, HL ;
Xu, TM ;
Hui, SE ;
Zhou, QL .
FLOW MEASUREMENT AND INSTRUMENTATION, 2006, 17 (02) :87-92
[10]   Capacitance sensor for void fraction measurement in water/steam flows [J].
Jaworek, A ;
Krupa, A ;
Trela, M .
FLOW MEASUREMENT AND INSTRUMENTATION, 2004, 15 (5-6) :317-324