Microcolumn ion chromatography of inorganic anions using bovine serum albumin stationary phase with indirect photometric detection

被引:12
|
作者
Munaf, E [1 ]
Zein, R [1 ]
Takeuchi, T [1 ]
Miwa, T [1 ]
机构
[1] ANDALAS UNIV,FAC MATH & NAT SCI,ANALYT ENVIRONM CHEM LAB,PADANG 25163,INDONESIA
关键词
ion chromatography; indirect detection; bovine serum albumin; inorganic anions;
D O I
10.1007/BF02271000
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Microcolumn ion chromatography of inorganic anions has been studied using bovine serum albumin immobilized on silica gel as a stationary phase. Several mobile phase solutions were examined, involving sodium iodide, potassium hydrogen phthalate, 2,6-anthraquinone disulfonate (2,6-AQDS) and sodium salicylate. 2,6-AQDS achieved better separation of the analytes tested. Chloride, nitrate, iodide, thiocyanate and sulphate could be separated within 8 min. Detection limits were in the range of 0.9-2.9 mu M, corresponding to mass detection limits of 0.18-0.59 pmol. The system was applied to the determination of inorganic anions in environmental water and biological samples.