摘要
Objective. The objective of this study was to verify if accurate patient-to-image registration for precision navigation in maxillofacial surgery is possible based on cone beam computed tomography (CBCT) image data. Study design. A maxillary registration template was placed on a standard plastic skull phantom that was equipped with a custom made model of the maxilla and with target markers. Imaging was performed with a CBCT device (Newtom 9000 Digital Volume Tomograph (DVT), QR s.r.l., Verona, Italy) and a computed tomography (CT) scanner (Somatom 4, Siemens, Forchheim, Germany). Using an infrared navigation system (Polaris, NDI, Waterloo, Ontario), multiple pair-point registration of both image data sets and the phantom were performed. The target registration error (TRE) was evaluated. Results. A total of 243 registrations were performed for either image data set. The spatial distribution of TRE on the skull showed increasing inaccuracy with growing distance from the registration markers. The average target registration error was 1.50 +/- 0.82 mm with CBCT and 1.57 +/- 0.84 mm with CT image data and did not differ significantly. Error distribution correlated strongly between CT- and CBCT-based registration. Conclusions. The overall registration accuracy based on CBCT image data was similar to CT. The strong correlation of the geometric distribution of TRE between CT-and CBCT-based measurements proves that CBCT can be equivalent to CT in image-guided maxillofacial surgery. (Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2009;107:e41-e48)
引用
收藏
页码:E41 / E48
页数:8
相关论文
共 37 条
  • [1] PIONEERS OF STEREOTAXIC NEUROSURGERY
    ALRODHAN, NRF
    KELLY, PJ
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1992, 58 (1-4) : 60 - 66
  • [2] Using a flat-panel detector in high resolution cone beam CT for dental imaging
    Baba, R
    Ueda, K
    Okabe, M
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2004, 33 (05) : 285 - 290
  • [3] Comparison of flat-panel detector and image-intensifier detector for cone-beam CT
    Baba, R
    Konno, Y
    Ueda, K
    Ikeda, S
    [J]. COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2002, 26 (03) : 153 - 158
  • [4] Relevance of head motion in dental cone-beam CT scanner images depending on patient positioning
    Bontempi, Marco
    Bettuzzi, Matteo
    Casali, Franco
    Pasini, Alessandro
    Rossi, Alberto
    Ariu, Marilu
    [J]. INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2008, 3 (3-4) : 249 - 255
  • [5] Comparison of different 3D navigation systems by a clinical "user"
    Cartellieri, M
    Kremser, J
    Vorbeck, F
    [J]. EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2001, 258 (01) : 38 - 41
  • [6] Endoscopic sinus surgery using intraoperative computed tomography imaging for updating a three-dimensional navigation system
    Cartellieri, M
    Vorbeck, F
    [J]. LARYNGOSCOPE, 2000, 110 (02) : 292 - 296
  • [7] Caversaccio M, 1999, Comput Aided Surg, V4, P328, DOI 10.3109/10929089909148186
  • [8] Do flat detector cardiac X-ray systems convey advantages over image-intensifier-based systems? Study comparing X-ray dose and image quality
    Davies, Andrew G.
    Cowen, Arnold R.
    Kengyelics, Stephen M.
    Moore, Janet
    Sivananthan, Mohan U.
    [J]. EUROPEAN RADIOLOGY, 2007, 17 (07) : 1787 - 1794
  • [9] Ecke U, 2003, HNO, V51, P386, DOI 10.1007/s00106-002-0738-1
  • [10] Image-to-patient registration techniques in head surgery
    Eggers, G.
    Muehling, J.
    Marmulla, R.
    [J]. INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2006, 35 (12) : 1081 - 1095