共 34 条
MRI Findings Associated with Recalled Modular Femoral Neck Rejuvenate and ABG Implants
被引:15
作者:
Walsh, Christopher P.
[1
]
Hubbard, James C.
[2
]
Nessler, Joseph P.
[3
]
Markel, David C.
[1
,4
,5
]
机构:
[1] Providence Hosp, Orthopaed Surg Residency Program, Detroit Med Ctr, Detroit, MI USA
[2] Wayne State Univ, Sch Med, Detroit, MI USA
[3] St Cloud Orthopaed, St Cloud, MN USA
[4] Providence Hosp & Med Ctr, Southfield, MI 48075 USA
[5] CORE Inst, Southfield, MI USA
关键词:
chromium;
cobalt;
revision total hip arthroplasty;
modular neck hip implants;
adverse local tissue reactions;
metal artifact reduction sequence magnetic resonance imaging;
TOTAL HIP-ARTHROPLASTY;
LEG LENGTH DISCREPANCY;
FRETTING WEAR;
CORROSION;
COMPONENT;
TISSUES;
DESIGN;
MOTION;
RANGE;
TAPER;
D O I:
10.1016/j.arth.2015.05.021
中图分类号:
R826.8 [整形外科学];
R782.2 [口腔颌面部整形外科学];
R726.2 [小儿整形外科学];
R62 [整形外科学(修复外科学)];
学科分类号:
摘要:
MARS-MRI is suggested for the diagnosis of adverse local tissue reactions (ALTR) in patients with recalled femoral stems with modular necks, but there has been no major study looking at MARS-MRI findings in this population. A retrospective review was performed on 312 patients who received a modular neck hip implant between October 2007 and February 2012. 62% of patients had intra-articular effusions, with 27% containing debris. Extra-capsular effusions were present in 35% of hips. 54% had synovitis and 5.4% had osteolysis. Tendinopathy and tendon disruption was present in the gluteus medius (58%/12%), hamstring (56%/12%), gluteus minimus (38%/7.7%) and iliopsoas (7.1%/4.8%). Abnormal MARS-MRI findings are associated with modular neck femoral components and can suggest underlying ALTR. MARS-MRI abnormalities merit serious consideration in this population. (C) 2015 Elsevier Inc. All rights reserved.