A new device to limit extravasation during contrast-enhanced CT

被引:10
作者
Powell, CC [1 ]
Li, JM [1 ]
Rodino, L [1 ]
Anderson, FA [1 ]
机构
[1] Univ Massachusetts, Med Ctr, Dept Surg, Worcester, MA 01655 USA
关键词
D O I
10.2214/ajr.174.2.1740315
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The extravasation detection accessory (EDA) is designed for use during contrast-enhanced CT studies performed with a power injector. The EDA detects the changes in soft-tissue impedance that occur with enhanced extravasation and halts the further infusion of contrast material via a feedback circuit to the injector. We tested the sensitivity of this device in a model of contrast extravasation. MATERIALS AND METHODS. Study subjects had an extravasation of 5% dextrose in water (nonionic contrast equivalent) in one arm and 0.9% sodium chloride solution (ionic contrast equivalent) in the other. An EDA was placed over the site of infusion and connected to a power injector. Injections were performed at 0.25 ml/sec (n = 40), 2.5 ml/sec (n = 62), or 5 ml/sec (n = 20). RESULTS. At infusion rates of 2.5 and 5 ml/sec, the device halted the injector in every subject after an average volume of 12.5 +/- 1.6 ml was delivered. At 0.25 ml/sec, the device failed to halt the injector in 11 of 20 events. After reprogramming the algorithm, 10 more subjects were tested at the lowest injection rate. The device halted 18 of 20 extravasation events with an average volume of 3.7 +/- 0.5 ml. CONCLUSION. In our model of contrast extravasation, the EDA halted a power injector with reliability and reproducibility before a large volume of contrast material was delivered. The sensitivity of the device approached, but did not reach, 100%. This device may serve to diminish the morbidity of extravasation events.
引用
收藏
页码:315 / 318
页数:4
相关论文
共 16 条
  • [1] IMPEDANCE PLETHYSMOGRAPHY IN THE DIAGNOSIS OF ARTERIAL AND VENOUS DISEASE
    ANDERSON, FA
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 1984, 12 (01) : 79 - 102
  • [2] [Anonymous], 1940, THE J
  • [3] SEVERE EXTRAVASATION INJURY - AN AVOIDABLE IATROGENIC DISASTER
    BURD, DAR
    SANTIS, G
    MILWARD, TM
    [J]. BRITISH MEDICAL JOURNAL, 1985, 290 (6481) : 1579 - 1580
  • [4] EXTRAVASATION OF NONIONIC RADIOLOGIC CONTRAST-MEDIA - EFFICACY OF CONSERVATIVE TREATMENT
    COHAN, RH
    DUNNICK, NR
    LEDER, RA
    BAKER, ME
    [J]. RADIOLOGY, 1990, 176 (01) : 65 - 67
  • [5] COHAN RH, 1996, RADIOLOGY, V199, P697
  • [6] ENSMINGER AH, 1994, FOODS NUTR ENCY, P222
  • [7] ESPEJO MGA, 1989, CRIT CARE MED, V17, P360
  • [8] CORRELATION OF WHOLE-BODY IMPEDANCE WITH TOTAL BODY WATER VOLUME
    HOFFER, EC
    MEADOR, CK
    SIMPSON, DC
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1969, 27 (04) : 531 - &
  • [9] EXPERIMENTAL TISSUE-DAMAGE AFTER SUBCUTANEOUS INJECTION OF WATER-SOLUBLE CONTRAST-MEDIA
    KIM, SH
    PARK, JH
    KIM, YI
    KIM, CW
    HAN, MC
    [J]. INVESTIGATIVE RADIOLOGY, 1990, 25 (06) : 678 - 685
  • [10] EXTRAVASATIONS OF RADIOGRAPHIC CONTRAST MATERIAL IN THE UPPER EXTREMITY
    LOTH, TS
    JONES, DEC
    [J]. JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1988, 13A (03): : 395 - 398