A simple differential steady-state method to measure the thermal conductivity of solid bulk materials with high accuracy

被引:48
作者
Kraemer, D. [1 ]
Chen, G. [1 ]
机构
[1] MIT, Dept Mech Engn, Cambridge, MA 02139 USA
关键词
DIFFUSIVITY;
D O I
10.1063/1.4865111
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Accurate measurements of thermal conductivity are of great importance for materials research and development. Steady-state methods determine thermal conductivity directly from the proportionality between heat flow and an applied temperature difference (Fourier Law). Although theoretically simple, in practice, achieving high accuracies with steady-state methods is challenging and requires rather complex experimental setups due to temperature sensor uncertainties and parasitic heat loss. We developed a simple differential steady-state method in which the sample is mounted between an electric heater and a temperature-controlled heat sink. Our method calibrates for parasitic heat losses from the electric heater during the measurement by maintaining a constant heater temperature close to the environmental temperature while varying the heat sink temperature. This enables a large signal-to-noise ratio which permits accurate measurements of samples with small thermal conductance values without an additional heater calibration measurement or sophisticated heater guards to eliminate parasitic heater losses. Additionally, the differential nature of the method largely eliminates the uncertainties of the temperature sensors, permitting measurements with small temperature differences, which is advantageous for samples with high thermal conductance values and/or with strongly temperature-dependent thermal conductivities. In order to accelerate measurements of more than one sample, the proposed method allows for measuring several samples consecutively at each temperature measurement point without adding significant error. We demonstrate the method by performing thermal conductivity measurements on commercial bulk thermoelectric Bi2Te3 samples in the temperature range of 30-150 degrees C with an error below 3%. (C) 2014 AIP Publishing LLC.
引用
收藏
页数:6
相关论文
共 27 条
[1]  
Angstrom A.J., 1862, ANN PHYS-BERLIN, V190, P513, DOI 10.1002/andp.18621901202
[2]  
[Anonymous], 2022, IEC
[3]  
[Anonymous], 1982, 605842 IEC
[4]  
[Anonymous], THERMAL CONDUCTIVITY
[5]  
[Anonymous], 2013, C177 ASTM
[6]   Historical Evolution of the Transient Hot-Wire Technique [J].
Assael, Marc J. ;
Antoniadis, Konstantinos D. ;
Wakeham, William A. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2010, 31 (06) :1051-1072
[7]   A METHOD FOR MEASURING THE THERMAL CONDUCTIVITY OF SMALL SAMPLES OF POORLY CONDUCTING MATERIALS SUCH AS OPTICAL CRYSTALS [J].
BALLARD, SS ;
MCCARTHY, KA ;
DAVIS, WC .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1950, 21 (11) :905-907
[8]  
Carslaw J.C., 1959, CONDUCTION HEAT SOLI, V2nd
[9]   THERMAL CONDUCTIVITY OF BERYLLIUM OXIDE FROM 40-DEGREE-C TO 750-DEGREE-C [J].
DITMARS, DA ;
GINNINGS, DC .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1957, 59 (02) :93-99
[10]  
Flynn, 1969, THERMAL CONDUCTIVITY, V1, P241