Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment

被引:22
作者
He, Tian [1 ]
Qian, Xiaojun [1 ]
Zhai, Renyou [1 ]
Yang, Zongtao [1 ]
机构
[1] Capital Med Univ, Beijing Chaoyang Hosp, Dept Radiol, Beijing, Peoples R China
关键词
CT-quantitative; beam hardening artifacts; dual-energy spectral CT; RENAL CYST PSEUDOENHANCEMENT; CONTRAST-ENHANCED CT; VARIABILITY; REDUCTION; PITFALLS;
D O I
10.1097/RCT.0000000000000287
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To compare computed tomography (CT) number measurement consistency under different beam hardening conditions in phantom experiment between dual-energy spectral CT and conventional CT imaging. Materials and Methods A phantom with 8 cells in periphery region and 1 cell in central region were used. The 8 conditioning tubes in the periphery region were filled with 1 of the 3 iodine solutions to simulate different beam hardening conditions: 0 for no beam hardening (NBH), 20 mg/mL for weak beam hardening (WBH) and 50 mg/mL for severe beam hardening (SBH) condition. Test tube filled with 0, 0.1, 0.5, 1, 2, 5, 10, 20, and 50 mg/mL iodine solution was placed in the central cell alternately. The phantom was scanned with conventional CT mode with 80, 100, 120, and 140 kVp and dual energy spectral CT mode. For spectral CT, 11 monochromatic image sets from 40 to 140 keV with interval of 10 keV were reconstructed. The CT number shift caused by beam hardening was evaluated by measuring the CT number difference (CT) with and without beam hardening, with the following formulas: CTWBH = |CTWBH - CTNBH| and CTSBH = |CTSBH - CTNBH|. Data were compared with 1-way analysis of variance. Results Under both WBH and SBH conditions, the CT number shifts in all monochromatic image sets were less than those for polychromatic images (all P < 0.001). Under WBH condition, the maximum CT number shift was less than 6 Hounsfield units for monochromatic spectral CT images of all energy levels; under SBH condition, only monochromatic images at 70 keV and 80 keV had CT number shift less than 6 HU. Conclusion Dual energy spectral CT imaging provided more accurate CT number measurement than conventional CT under various beam hardening conditions. The optimal keV level for monochromatic spectral CT images with the most accurate CT number measurement depends on the severities of beam hardening condition.
引用
收藏
页码:981 / 985
页数:5
相关论文
共 16 条
[1]   Pseudoenhancement of simulated renal cysts in a phantom using different multidetector CT scanners [J].
Abdulla, C ;
Kalra, MK ;
Saini, S ;
Maher, MM ;
Ahmad, A ;
Halpern, E ;
Silverman, SG .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (06) :1473-1476
[2]   Renal cysts: Is attenuation artifactually increased on contrast-enhanced CT images? [J].
Bae, KT ;
Heiken, JP ;
Siegel, CL ;
Bennett, HF .
RADIOLOGY, 2000, 216 (03) :792-796
[3]   Renal cyst pseudoenhancement: Evaluation with an anthropomorphic body CT phantom [J].
Birnbaum, BA ;
Maki, DD ;
Chakraborty, DP ;
Jacobs, JE ;
Babb, JS .
RADIOLOGY, 2002, 225 (01) :83-90
[4]   Multi-detector row CT attenuation measurements: Assessment of intra- and interscanner variability with an anthropomorphic body CT phantom [J].
Birnbaum, Bernard A. ;
Hindman, Nicole ;
Lee, Julie ;
Babb, James S. .
RADIOLOGY, 2007, 242 (01) :109-119
[5]   Evaluation of pseudoenhancement of renal cysts during contrast-enhanced CT [J].
Coulam, CH ;
Sheafor, DH ;
Leder, RA ;
Paulson, EK ;
DeLong, DM ;
Nelson, RC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (02) :493-498
[6]   Statistical image reconstruction for polyenergetic X-ray computed tomography [J].
Elbakri, IA ;
Fessler, JA .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (02) :89-99
[7]   Pitfalls in renal mass evaluation and how to avoid them [J].
Israel, Gary M. ;
Bosniak, Morton A. .
RADIOGRAPHICS, 2008, 28 (05) :1325-1338
[8]   THE UNRELIABILITY OF CT-NUMBERS AS ABSOLUTE VALUES [J].
LEVI, C ;
GRAY, JE ;
MCCULLOUGH, EC ;
HATTERY, RR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1982, 139 (03) :443-447
[9]   Renal cyst pseudoenhancement: Beam-hardening effects on CT numbers [J].
Maki, DD ;
Birnbaum, BA ;
Chakraborty, DP ;
Jacobs, JE ;
Carvalho, BM ;
Herman, GT .
RADIOLOGY, 1999, 213 (02) :468-472
[10]   Precision of the measurement of CT numbers: comparison of dual-energy CT spectral imaging with fast kVp switching and conventional CT with phantoms [J].
Matsuda, Izuru ;
Akahane, Masaaki ;
Sato, Jiro ;
Katsura, Masaki ;
Kiryu, Shigeru ;
Yoshioka, Naoki ;
Kunimatsu, Akira ;
Ino, Kenji ;
Ohtomo, Kuni .
JAPANESE JOURNAL OF RADIOLOGY, 2012, 30 (01) :34-39