Brachytherapy dose-volume histogram commissioning with multiple planning systems

被引:9
作者
Gossman, Michael S. [1 ,2 ]
Hancock, Samuel S. [3 ]
Kudchadker, Rajat J. [4 ]
Lundahl, Paul R. [5 ]
Cao, Minsong [6 ]
Melhus, Christopher S. [7 ]
机构
[1] Tri State Reg Canc Ctr, Dept Radiat Oncol, Ashland, KY 41101 USA
[2] Regulat Direct Med Phys, Flatwoods, KY USA
[3] Southeast Missouri Hosp, Dept Radiat Oncol, Cape Girardeau, MO USA
[4] UT MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX USA
[5] Riverside Methodist Hosp, Dept Radiat Oncol, Columbus, OH 43214 USA
[6] UCLA Sch Med, Dept Radiat Oncol, Los Angeles, CA USA
[7] Tufts Univ, Sch Med, Dept Radiat Oncol, Boston, MA 02111 USA
关键词
brachytherapy; dose; DVH; histogram; volume; RADIATION-THERAPY COMMITTEE; CLINICAL IMPLEMENTATION; QUALITY-ASSURANCE; AAPM; DOSIMETRY; RECOMMENDATIONS; UPDATE;
D O I
10.1120/jacmp.v15i2.4620
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The first quality assurance process for validating dose-volume histogram data involving brachytherapy procedures in radiation therapy is presented. The process is demonstrated using both low dose-rate and high dose-rate radionuclide sources. A rectangular cuboid was contoured in five commercially available brachytherapy treatment planning systems. A single radioactive source commissioned for QA testing was positioned coplanar and concentric with one end. Using the brachytherapy dosimetry formalism defined in the AAPM Task Group 43 report series, calculations were performed to estimate dose deposition in partial volumes of the cuboid structure. The point-source approximation was used for a I-125 source and the line-source approximation was used for a Ir-192 source in simulated permanent and temporary implants, respectively. Hand-calculated, dose-volume results were compared to TPS-generated, dose-volume histogram (DVH) data to ascertain acceptance. The average disagreement observed between hand calculations and the treatment planning system DVH was less than 1% for the five treatment planning systems and less than 5% for 1 cm <= r <= 5 cm. A reproducible method for verifying the accuracy of volumetric statistics from a radiation therapy TPS can be employed. The process satisfies QA requirements for TPS commissioning, upgrading, and annual testing. We suggest that investigations be performed if the DVH % Vol(TPS) "actual variance" calculations differ by more than 5% at any specific radial distance with respect to % Vol(TG-43), or if the "average variance" DVH % Vol(TPS) calculations differ by more than 2% over all radial distances with respect to % Vol(TG-43).
引用
收藏
页码:110 / 120
页数:11
相关论文
共 50 条
[31]   Compute Unified Device Architecture-Based Parallel Dose-Volume Histogram Computation [J].
Wang, Yangping ;
Deng, Chong ;
Li, Lian ;
Dang, Jianwu .
JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2015, 5 (04) :833-840
[32]   A Varian DynaLog file-based procedure for patient dose-volume histogram-based IMRT QA [J].
Calvo-Ortega, Juan F. ;
Teke, Tony ;
Moragues, Sandra ;
Pozo, Miquel ;
Casals, Joan .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02) :100-109
[33]   Stereotactic Radiosurgery for Multiple Brain Metastases: A Dose-Volume Study [J].
Kairn, Tanya ;
Zolfaghari, Somayeh ;
Papworth, Daniel ;
West, Mark ;
Schlect, David ;
Crowe, Scott .
WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 3, 2019, 68 (03) :443-446
[34]   Comparison of dose-volume histograms and dose-wall histograms of the rectum of patients treated with intracavitary brachytherapy [J].
Olszewska, AM ;
Saarnak, AE ;
de Boer, RW ;
van Bunningen, BNFM ;
Steggerda, MJ .
RADIOTHERAPY AND ONCOLOGY, 2001, 61 (01) :83-85
[35]   Quality assurance procedures based on dosimetric, gamma analysis as a fast reliable tool for commissioning brachytherapy treatment planning systems [J].
Zwierzchowski, Grzegorz ;
Bieleda, Grzegorz ;
Skowronek, Janusz .
RADIOLOGY AND ONCOLOGY, 2017, 51 (04) :469-474
[36]   GEC-ESTRO guidelines for commissioning and quality assurance of TG-43 based brachytherapy treatment planning systems [J].
De Brabandere, Marisol ;
Beaulieu, Luc ;
Dejean, Catherine ;
Dempsey, Claire ;
Lee, Chris ;
Mason, Josh ;
Rivard, Mark J. ;
Perez-Calatayud, Jose ;
Smith, Ryan ;
Steenhuijsen, Jacco ;
Walter, Renate ;
Workman, Geraldine ;
Zuchora, Anysja ;
Siebert, Frank-Andre .
RADIOTHERAPY AND ONCOLOGY, 2025, 210
[37]   Multiple direction needle-path planning and inverse dose optimization for robotic low-dose rate brachytherapy [J].
Aumueller, Philipp ;
Rothfuss, Andreas ;
Polednik, Martin ;
Abo-Madyan, Yasser ;
Ehmann, Michael ;
Giordano, Frank A. ;
Clausen, Sven .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2022, 32 (02) :173-187
[38]   A user-oriented procedure for the commissioning and quality assurance testing of treatment planning system dosimetry in high-dose-rate brachytherapy [J].
Peppa, Vasiliki ;
Pantelis, Evaggelos ;
Pappas, Eleftherios ;
Lahanas, Vasileios ;
Loukas, Constantinos ;
Papagiannis, Panagiotis .
BRACHYTHERAPY, 2016, 15 (02) :252-262
[39]   Dose evaluation of organs at risk (OAR) cervical cancer using dose volume histogram (DVH) on brachytherapy [J].
Wibowo, R. Arif ;
Haris, Bambang ;
Islamiyah, Inganatul .
INTERNATIONAL CONFERENCE ON PHYSICAL INSTRUMENTATION AND ADVANCED MATERIALS, 2017, 853
[40]   Evolution of brachytherapy treatment planning to deterministic radiation transport for calculation of cardiac dose [J].
Sandwall, Peter A. ;
Feng, Yuntao ;
Platt, Michael ;
Lamba, Michael ;
Mahalingam, Sudha .
MEDICAL DOSIMETRY, 2018, 43 (02) :150-158