A recovery coefficient method for partial volume correction of PET images

被引:96
作者
Srinivas, Shyam M. [2 ]
Dhurairaj, Thiruvenkatasamy [3 ]
Basu, Sandip [1 ,4 ]
Bural, Gonca [5 ]
Surti, Suleman [1 ]
Alavi, Abass [1 ]
机构
[1] Hosp Univ Penn, Div Nucl Med, Dept Radiol, Philadelphia, PA 19104 USA
[2] Cleveland Clin, Dept Nucl Med, Cleveland, OH 44106 USA
[3] Univ Minnesota, Dept Radiol, Minneapolis, MN 55455 USA
[4] Tata Mem Hosp, Radiat Med Ctr BARC, Bombay 400012, Maharashtra, India
[5] Brigham & Womens Hosp, Dept Radiol, Boston, MA 02115 USA
关键词
Partial volume correction; Positron emission tomography; Hot sphere/warm background phantom; Recovery coefficient; POSITRON-EMISSION-TOMOGRAPHY; LUNG-CANCER; SIZE;
D O I
10.1007/s12149-009-0241-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Correction of the "partial volume effect" has been an area of great interest in the recent times in quantitative PET imaging and has been mainly studied with count recovery models based upon phantoms that incorporate hot spheres in a cold background. The goal of this research study was to establish a similar model that is closer to a biological imaging environment, namely hot spheres/lesions in a warm background and to apply this model in a small cohort of patients. A NEMA phantom with six spheres (diameters 1-3.7 cm) was filled with (18)FDG to give sphere:background activity ratios of 8:1, 6:1, and 4:1 for three different acquisitions on a Philips Allegro scanner. The hot sphere SUVmax and the background average SUV were measured for calculation of recovery coefficients (RCs). Using the RCs, the lesion diameters, and the lesion:background ratio, the SUVmax of 64 lesions from 17 patients with biopsy proven lung cancer were corrected. The RCs versus sphere diameters produced characteristic logarithmic curves for each phantom (RCs ranged from 80% to 11%). From a cohort of 17 patients with biopsy proven lung cancer, 64 lesions combined had a mean SUVmax of 7.0 and size of 2.5 cm. After partial volume correction of the SUVmax of each lesion, the average SUVmax increased to 15.5. Hot spheres in a warm background more closely resemble the actual imaging situation in a living subject when compared to hot spheres in a cold background. This method could facilitate generation of equipment specific recovery coefficients for partial volume correction. The clinical implications for the increased accuracy in SUV determination are certainly of potential value in oncologic imaging.
引用
收藏
页码:341 / 348
页数:8
相关论文
共 19 条
[1]   EVALUATION OF BREAST MASSES AND AXILLARY LYMPH-NODES WITH [F-18] 2-DEOXY-2-FLUORO-D-GLUCOSE PET [J].
ADLER, LP ;
CROWE, JP ;
ALKAISI, NK ;
SUNSHINE, JL .
RADIOLOGY, 1993, 187 (03) :743-750
[2]   Positron emission tomography imaging in nonmalignant thoracic disorders [J].
Alavi, A ;
Gupta, N ;
Alberini, JL ;
Hickeson, M ;
Adam, LE ;
Bhargava, P ;
Zhuang, HM .
SEMINARS IN NUCLEAR MEDICINE, 2002, 32 (04) :293-321
[3]  
Boucher L, 2004, J NUCL MED, V45, P214
[4]  
CHAWLUK JB, 1987, J NUCL MED, V28, P431
[5]  
Chen CH, 1999, J NUCL MED, V40, P118
[6]  
Daube-Witherspoon ME, 2002, J NUCL MED, V43, P1398
[7]   Use of a corrected standardized uptake value based on the lesion size on CT permits accurate characterization of lung nodules on FDG-PET [J].
Hickeson, M ;
Yun, MJ ;
Matthies, A ;
Zhuang, HM ;
Adam, LE ;
Lacorte, L ;
Alavi, A .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (12) :1639-1647
[8]   QUANTITATION IN POSITRON EMISSION COMPUTED-TOMOGRAPHY .1. EFFECT OF OBJECT SIZE [J].
HOFFMAN, EJ ;
HUANG, SC ;
PHELPS, ME .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1979, 3 (03) :299-308
[9]   ANALYSIS OF EMISSION TOMOGRAPHIC SCAN DATA - LIMITATIONS IMPOSED BY RESOLUTION AND BACKGROUND [J].
KESSLER, RM ;
ELLIS, JR ;
EDEN, M .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1984, 8 (03) :514-522
[10]   CORRECTION OF PET DATA FOR PARTIAL VOLUME EFFECTS IN HUMAN CEREBRAL-CORTEX BY MR IMAGING [J].
MELTZER, CC ;
LEAL, JP ;
MAYBERG, HS ;
WAGNER, HN ;
FROST, JJ .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1990, 14 (04) :561-570