Large-area phase-contrast X-ray imaging using a two-crystal X-ray interferometer
被引:34
作者:
Yoneyama, A
[1
]
Momose, A
Koyama, I
Seya, E
Takeda, T
Itai, Y
Hirano, K
Hyodo, K
机构:
[1] Hitachi Ltd, Adv Res Lab, Hatoyama, Saitama 3500395, Japan
[2] Univ Tokyo, Dept Appl Phys, Bunkyo Ku, Tokyo 1138656, Japan
[3] Hitachi Ltd, Cent Res Lab, Tokyo 1858601, Japan
[4] Univ Tsukuba, Inst Clin Med, Tsukuba, Ibaraki 3508575, Japan
[5] High Energy Accelerator Res Org, Inst Mat Sci, Tsukuba, Ibaraki 3050801, Japan
关键词:
phase-contrast;
X-ray imaging;
two-crystal X-ray interferometer;
phase-contrast computed tomography;
cancer;
D O I:
10.1107/S090904950201350X
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
Large-area (25 x 20 mm) phase-contrast X-ray imaging was attained by using a skew-symmetric two-crystal X-ray interferometer. The sub-nrad angular control required to operate the X-ray interferometer was achieved with a sleeve bearing and a feedback-positioning system. As a demonstration, measurements of a phase map of a rat's liver and a phase-contrast tomographic three-dimensional image of a piece of a rabbit's liver were performed at the Photon Factory using 0.07 nm synchrotron radiation X-rays.