Late Surgical Bleeding Following Total Artificial Heart Implantation

被引:4
作者
Hermsen, Joshua L. [1 ]
Smith, Jason W. [1 ]
Pal, Jay D. [1 ]
Mahr, Claudius [2 ]
Masri, S. Carolina [2 ]
Dardas, Todd F. [2 ]
Cheng, Richard K. [2 ]
Mokadam, Nahush A. [1 ]
机构
[1] Univ Washington, Med Ctr, Dept Surg, Div Cardiothorac Surg, Seattle, WA 98195 USA
[2] Univ Washington, Med Ctr, Dept Med, Div Cardiol, Seattle, WA 98195 USA
关键词
TRANSPLANTATION; BRIDGE;
D O I
10.1111/jocs.12601
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionMechanical circulatory support for heart failure, including the Total Artificial Heart (TAH, Syncardia, Tucson, AZ, USA) has increased in recent years. This report describes bleeding complications associated with the device. MethodsA single institution prospectively maintained quality improvement database was reviewed encompassing the first year of clinical experience with the TAH. Patients who underwent TAH implantation were identified, and a review of complications and outcomes was undertaken. ResultsTen patients underwent TAH implant. Four patients experienced delayed postoperative bleeding. In three patients the manifestation of bleeding was tamponade and evidenced by TAH decreased cardiac output. In two patients, at postoperative days 31 and 137, there was a partial disruption of the aortic anastomosis along the outer curvature with pseudoaneurysm formation. Both were repaired by primary suture closure, without use of cardiopulmonary bypass. There was no mortality attributable to bleeding. ConclusionsTAH patients are at risk for delayed postoperative bleeding, often manifest as an acute decrease in cardiac output. Due to pulsatility and high dP/dT, bleeding from the aortic anastomosis should be considered in the differential of a patient with low flow and/or tamponade. doi: 10.1111/jocs.12601 (J Card Surg 2015;30:771-774)
引用
收藏
页码:771 / 774
页数:4
相关论文
共 7 条
[1]  
Copeland J G, 2000, Semin Thorac Cardiovasc Surg, V12, P238
[2]   Experience with more than 100 total artificial heart implants [J].
Copeland, Jack G. ;
Copeland, Hannah ;
Gustafson, Monica ;
Mineburg, Nicole ;
Covington, Diane ;
Smith, Richard G. ;
Friedman, Mark .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2012, 143 (03) :727-734
[3]   The CardioWest total artificial heart bridge to transplantation:: 1993 to 1996 national trial [J].
Copeland, JG ;
Arabía, FA ;
Banchy, ME ;
Sethi, GK ;
Foy, B ;
Long, J ;
Kormos, RL ;
Smith, RG .
ANNALS OF THORACIC SURGERY, 1998, 66 (05) :1662-1669
[4]   Cardiac replacement with a total artificial heart as a bridge to transplantation [J].
Copeland, JG ;
Smith, RG ;
Arabia, FA ;
Nolan, PE ;
Sethi, GK ;
Tsau, PH ;
McClellan, D ;
Slepian, MJ ;
Foy, B ;
Long, J ;
Doty, D ;
Tector, A ;
Kormos, R .
NEW ENGLAND JOURNAL OF MEDICINE, 2004, 351 (09) :859-867
[5]  
Ensor CR, 2010, TEX HEART I J, V37, P149
[6]   The SynCardia freedom driver: A portable driver for discharge home with the total artificial heart [J].
Jaroszewski, Dawn E. ;
Anderson, Eric M. ;
Pierce, Christopher N. ;
Arabia, Francisco A. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2011, 30 (07) :844-845
[7]   Persistent Anemia After Implantation of the Total Artificial Heart [J].
Mankad, Anit K. ;
Tang, Daniel G. ;
Clark, William B. ;
Flattery, Maureen ;
Harton, Suzanne ;
Katlaps, Gundars J. ;
Stribling, Warren K. ;
Cooke, Richard H. ;
Hess, Michael L. ;
Kasirajan, Vigneshwar ;
Shah, Keyur B. .
JOURNAL OF CARDIAC FAILURE, 2012, 18 (06) :433-438