Synthetic Transmit Beam Technique in an Aberrating Environment

被引:7
作者
Bjastad, Tore [1 ]
Aase, Svein A. [1 ]
Torp, Hans [1 ]
机构
[1] Norwegian Univ Sci & Technol, Dept Circulat & Med Imaging, Fac Med, N-7034 Trondheim, Norway
关键词
Compensation techniques - Image quality improvements - In-vivo measurement - Parallel beamforming - Performance aspects - Synthetic transmit beams - Tissue mimicking phantom - Ultrasound imaging systems;
D O I
10.1109/TUFFC.2009.1190
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Parallel beamforming is a commonly used method for increasing the frame rate in ultrasound imaging systems. By receiving in several directions for each transmission, the frame rate is increased. However, this method also introduces blocklike artifacts in the B-mode image, due to the reception offsets when compared with the transmission direction. The synthetic transmit beam technique (STB) has been previously proposed as a compensation technique when addressing these artifacts. Previous work by Hergum et al. investigated the performance of this method in regard to the case of 2 parallel beams in tissue mimicking phantoms without aberrations. This study is a continuation of that work in which this method is tested in an aberrating environment using 4 parallel beams. Several quantitative and qualitative performance aspects of this method have been investigated such as lateral shift invariance, beam-to-beam correlation fluctuations, speckle-tracking performance, improvements from higher order STB interpolation and beam profile shape preservation, as well as perceived image quality improvements. The results were obtained from simulations, in vivo measurements, and in vitro measurements. The results showed that aberration amplified the image artifacts for regular parallel beamforming, which resulted in more shift variance, lower beam-to-beam correlation, higher speckle-tracking error, and more variation in beam profile shape. Compared with regular parallel beamforming, STB resulted in a significantly better image quality and a higher score in all measuring methods. The improvements from using STB were largest in cases involving aberration. Using STB., the variation in beam-to-beam correlation was reduced from 30% to 1%, and the standard deviation of the speckle-tracking error was reduced from 8% to 1.5%.
引用
收藏
页码:1340 / 1351
页数:12
相关论文
共 19 条
[1]  
[Anonymous], 2014, Diagnostic Ultrasound Imaging
[2]  
Augustine L. J., 1987, US Patent, Patent No. [4,644,795, 4644795]
[3]   The impact of aberration on high frame rate cardiac B-mode imaging [J].
Bjastad, Tore ;
Aase, Svein A. ;
Torp, Hans .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2007, 54 (01) :32-41
[4]  
D'hooge J., 2007, Myocardial Imaging: Tissue Doppler and Speckle Tracking, chapter Principles and different techniques for speckle tracking, P17
[5]   A parallel tracking method for acoustic radiation force impulse imaging [J].
Dahl, Jeremy J. ;
Pinton, Gianmarco F. ;
Palmeri, Mark L. ;
Agrawal, Vineet ;
Nightingale, Kathryn R. ;
Trahey, Gregg E. .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2007, 54 (02) :301-312
[6]   Parallel beamforming using synthetic transmit beams [J].
Hergum, Torbjorn ;
Bjastad, Tore ;
Kristoffersen, Kjell ;
Torp, Hans .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2007, 54 (02) :271-280
[7]   Measurements of ultrasonic pulse distortion produced by human chest wall [J].
Hinkelman, LM ;
Szabo, TL ;
Waag, RC .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1997, 101 (04) :2365-2373
[8]  
Holley G. L., 1998, US Patent, Patent No. [5,779,640, 5779640]
[9]  
HUNT TJ, 1995, Patent No. 5462057
[10]   CALCULATION OF PRESSURE FIELDS FROM ARBITRARILY SHAPED, APODIZED, AND EXCITED ULTRASOUND TRANSDUCERS [J].
JENSEN, JA ;
SVENDSEN, NB .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1992, 39 (02) :262-267