Two-dimensional EPID dosimetry for an MR-linac: Proof of concept

被引:10
|
作者
Torres-Xirau, Iban [1 ]
Olaciregui-Ruiz, Igor [1 ]
van der Heide, Uulke A. [1 ]
Mans, Anton [1 ]
机构
[1] Antoni van Leeuwenhoek Hosp, Netherlands Canc Inst, Dept Radiat Oncol, Amsterdam, Netherlands
关键词
dose verification; EPID; MR-linac; portal dosimetry; QA; unity; PORTAL IMAGING DEVICE; BACK-PROJECTION ALGORITHM; BODY RADIATION-THERAPY; 3D DOSE RECONSTRUCTION; MODULATED ARC THERAPY; MAGNETIC-RESONANCE; VERIFICATION; IMRT; RADIOTHERAPY; PERFORMANCE;
D O I
10.1002/mp.13664
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose At our institute, in vivo patient dose distributions are reconstructed for all treatments delivered using conventional linacs from electronic portal imaging device (EPID) transit images acquired during treatment using a simple back-projection model. Currently, the clinical implementation of MRI-guided radiotherapy systems, which aims for online and real-time adaptation of the treatment plan, is progressing. In our department, the MR-linac (Unity, Elekta AB, Stockholm, Sweden) is now in clinical use. The aim of this work is to demonstrate the feasibility of two-dimensional (2D) EPID dosimetric verification for the magnetic resonance (MR)-linac by comparing back-projected EPID doses to ionization chamber (IC) array dose distributions. Materials and methods Our conventional back-projection algorithm was adapted for the MR-linac. The most important changes involve modeling of the attenuation by and scatter from the cryostat. The commissioning process involved the acquisition of square field EPID measurements using various phantom setups (varying SSD, phantom thickness, and field size). Commissioning models were created for gantry 0 degrees, 90 degrees, and 180 degrees and verified by comparing EPID-reconstructed 2D dose distributions to measurements made with the OCTAVIUS 1500 IC array (PTW, Freiburg, Germany) for two prostate and one rectum IMRT plans (25 beams total). The average of the gamma parameters (y-mean and y-pass rate) and the dose difference at a reference point were reported. Due to their construction, the attenuation of couch, bridge, and cryostat shows a much stronger dependence on gantry angle in the MR-linac compared to conventional linacs. We present a method to correct for these effects. This method is validated by dose reconstruction of the 25 intensity-modulated radiation therapy beams recorded at a certain gantry angle using the model of another gantry angle, combined with the correction method. Results For dose verification performed at a gantry angle identical to the commissioned model, the average y-mean and y-pass rate values (3% global dose, 2 mm, 10% isodose) were 0.37 +/- 0.07 and 98.1, 95% CI [98.1 +/- 2.4], respectively. The average dose difference at the reference point was -0.5% +/- 1.8%. Verification at gantry angles different from the commissioned model (i.e., using the gantry angle dependent correction) reported 0.39 +/- 0.08 and 97.6, 95% CI [96.9, 98.3] average y-mean and y-pass rate values. The average dose difference at the reference point was -0.1% +/- 1.8%. Conclusion The EPID dosimetry back-projection model was successfully adapted for the MR-linac at gantry 0 degrees, 90 degrees, and 180 degrees, accounting for the presence of the MRI housing between phantom (or patient) and the EPID. A method to account for the gantry angle dependence was also tested reporting similar results.
引用
收藏
页码:4193 / 4203
页数:11
相关论文
共 50 条
  • [1] Characterization of the a-Si EPID in the unity MR-linac for dosimetric applications
    Torres-Xirau, I.
    Olaciregui-Ruiz, I.
    Baldvinsson, G.
    Mijnheer, B. J.
    van der Heide, U. A.
    Mans, A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (02)
  • [2] A back-projection algorithm in the presence of an extra attenuating medium: towards EPID dosimetry for the MR-Linac
    Torres-Xirau, I.
    Olaciregui-Ruiz, I.
    Rozendaal, R. A.
    Gonzalez, P.
    Mijnheer, B. J.
    Sonke, J-J
    van der Heide, U. A.
    Mans, A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (15) : 6322 - 6340
  • [3] A Deep Learning-based correction to EPID dosimetry for attenuation and scatter in the Unity MR-Linac system
    Olaciregui-Ruiz, Igor
    Torres-Xirau, Iban
    Teuwen, Jonas
    van der Heide, Uulke A.
    Mans, Anton
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 71 : 124 - 131
  • [4] 3D dosimetric verification of unity MR-linac treatments by portal dosimetry
    Torres-Xirau, Iban
    Olaciregui-Ruiz, Igor
    Kaas, Jochem
    Nowee, Marlies E.
    van der Heide, Uulke A.
    Mans, Anton
    RADIOTHERAPY AND ONCOLOGY, 2020, 146 : 161 - 166
  • [5] Automatic dosimetric verification of online adapted plans on the Unity MR-Linac using 3D EPID dosimetry
    Olaciregui-Ruiz, Igor
    Vivas-Maiques, Begona
    van der Velden, Sandra
    Nowee, Marlies E.
    Mijnheer, Ben
    Mans, Anton
    RADIOTHERAPY AND ONCOLOGY, 2021, 157 : 241 - 246
  • [6] Elekta Unity MR-linac commissioning: mechanical and dosimetry tests
    Tsuneda, Masato
    Abe, Kota
    Fujita, Yukio
    Ikeda, Yohei
    Furuyama, Yoshinobu
    Uno, Takashi
    JOURNAL OF RADIATION RESEARCH, 2023, 64 (01) : 73 - 84
  • [7] Adaptive radiotherapy: The Elekta Unity MR-linac concept
    Winkel, Dennis
    Bol, Gijsbert H.
    Kroon, Petra S.
    van Asselen, Bram
    Hackett, Sara S.
    Werensteijn-Honingh, Anita M.
    Intven, Martijn P. W.
    Eppinga, Wietse S. C.
    Tijssen, Rob H. N.
    Kerkmeijer, Linda G. W.
    de Boer, Hans C. J.
    Mook, Stella
    Meijer, Gert J.
    Hes, Jochem
    Willemsen-Bosman, Mirjam
    de Groot-van Breugel, Eline N.
    Jurgenliemk-Schulz, Ina M.
    Raaymakers, Bas W.
    CLINICAL AND TRANSLATIONAL RADIATION ONCOLOGY, 2019, 18 : 54 - 59
  • [8] Quantifying uncertainties associated with reference dosimetry in an MR-Linac
    Iakovenko, Viktor
    Keller, Brian
    Malkov, Victor N. N.
    Sahgal, Arjun
    Sarfehnia, Arman
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (11):
  • [9] Gel dosimetry with an anthropomorphic phantom in an Elekta Unity MR-linac
    Ferrer, Carlos
    Huertas, Concepcion
    Navarro, Inma
    Deben, Beatriz
    Morera, Rosa
    Saez, Moises
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3187 - S3189
  • [10] Dosimetry with the TruView gel on a 0.35 T MR-Linac: A feasibility study
    Ermeneux, L.
    Petitfils, A.
    Marage, L.
    Gschwind, R.
    Huet, C.
    RADIATION MEASUREMENTS, 2024, 175