A qualitative and a quantitative analysis of an auto-segmentation module for prostate cancer

被引:58
作者
Huyskens, Dominique P. [1 ,2 ]
Maingon, Philippe [3 ]
Vanuytsel, Luc [4 ]
Remouchamps, Vincent [2 ]
Roques, Tom [5 ]
Dubray, Bernard [6 ]
Haas, Benjamin [7 ]
Kunz, Patrik [7 ]
Coradi, Thomas [7 ]
Buehlman, Rene [7 ]
Reddick, Robin [8 ]
Van Esch, Ann [2 ]
Salamon, Emile [2 ]
机构
[1] 7Sigma, QA Team Radiotherapy Phys, B-3150 Tildonk, Belgium
[2] Clin Ste Elisabeth, Dept Radiotherapy, Namur, Belgium
[3] Ctr F Leclerc, Dept Radiotherapy, Dijon, France
[4] Heilig Hartziekenhuis, Dept Radiotherapy, Roeselare, Belgium
[5] Norwich Univ Hosp, Dept Radiotherapy, Norwich, Norfolk, England
[6] Ctr Henri Becquerel, Dept Radiotherapy, F-76038 Rouen, France
[7] Dept Image Proc & Res, Varian Med Syst Imaging Lab, Baden, Switzerland
[8] Treatment Planning Syst, Varian Med Syst, Las Vegas, NV USA
关键词
Prostate; Automatic segmentation; Treatment planning; COMPUTED-TOMOGRAPHY IMAGES; GUIDED RADIATION-THERAPY; AUTOMATIC SEGMENTATION; CT IMAGES; RADIOTHERAPY; DELINEATION; REGISTRATION; HEAD; KNOWLEDGE; ATLAS;
D O I
10.1016/j.radonc.2008.08.007
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: This work describes the clinical validation Of all automatic segmentation algorithm in CT-based radiotherapy planning for prostate cancer patients. Material and methods: The validated auto-segmentation algorithm (Smart Segmentation, version 1.0.05) is a rule-based algorithm using anatomical reference points and organ-specific segmentation methods, developed by Varian Medical Systems (Varian Medical Systems iLab, Baden, Switzerland). For the qualitative analysis, 39 prostate patients are analysed by six clinicians. Clinicians are asked to rate the auto-segmented organs (prostate, bladder, rectum and femoral heads) and to indicate the number of slices to correct. For the quantitative analysis, seven radiation oncologists are asked to contour seven prostate patients. The individual clinician contour variations are compared to the automatic contours by means of surface and Volume statistics, calculating the relative volume errors and both the Volume and slice-by-slice degree of support, a statistical metric developed for the purposes of this validation. Results: The mean time needed for the automatic module to contour the four structures is about one minute on a standard computer. The qualitative evaluation using a score with four levels ("not acceptable", "acceptable", "good" and "excellent") shows that the mean score for the automatically contoured prostate is "good"; the bladder scores between "excellent" and "good"; the rectum scores between "acceptable" and "not acceptable". Using the concept of surface and volume degree of support, the degree of support given to the automatic module is comparable to the relative agreement among the clinicians for prostate and bladder. The slice-by-slice analysis of the surface degree of support pinpointed the areas of disagreement among the clinicians as well as between the clinicians and the automatic module. Conclusion: The efficiency and the limits of the automatic module are investigated with both a qualitative and a quantitative analysis. In general, with efficient correction tools at hand, the use of this auto-segmentation module will lead to a time gain for the prostate and the bladder; with the present version of the algorithm, modelling of the rectum still needs improvement. For the quantitative validation, the concept of relative volume error and degree of support proved very useful. (C) 2008 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 90 (2009) 337-345
引用
收藏
页码:337 / 345
页数:9
相关论文
共 26 条
  • [11] HAMARNEH G, 2007, APPL BIOMAT MED IMAG, V1, P387
  • [12] A NEW REPRESENTATION OF KNOWLEDGE CONCERNING HUMAN ANATOMY AND FUNCTION
    HOHNE, KH
    PFLESSER, B
    POMMERT, A
    RIEMER, M
    SCHIEMANN, T
    SCHUBERT, R
    TIEDE, U
    [J]. NATURE MEDICINE, 1995, 1 (06) : 506 - 511
  • [13] Evaluation of an atlas-based automatic segmentation software for the delineation of brain organs at risk in a radiation therapy clinical context
    Isambert, Aurelie
    Dhermain, Frederic
    Bidault, Francois
    Commowick, Olivier
    Bondiau, Pierre-Yves
    Malandain, Gregoire
    Lefkopoulos, Dimitri
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 87 (01) : 93 - 99
  • [14] A robust method for extraction and automatic segmentation of brain images
    Kovacevic, N
    Lobaugh, NJ
    Bronskill, MJ
    Levine, B
    Feinstein, A
    Black, SE
    [J]. NEUROIMAGE, 2002, 17 (03) : 1087 - 1100
  • [15] *NAT LIB MED US BO, 1990, NIH PUBL
  • [16] Automatic segmentation of pelvic structures from magnetic resonance images for prostate cancer radiotherapy
    Pasquier, David
    Lacornerie, Thomas
    Vermandel, Maximilien
    Rousseau, Jean
    Lartigau, Eric
    Betrouni, Nacim
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (02): : 592 - 600
  • [17] Automated model-based organ delineation for radiotherapy planning in prostatic region
    Pekar, V
    McNutt, TR
    Kaus, MR
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (03): : 973 - 980
  • [18] Evaluation of a segmentation procedure to delineate organs for use in construction of a radiation therapy planning atlas
    Qatarneh, SM
    Noz, ME
    Hyödynmaa, S
    Maguire, GQ
    Kramer, EL
    Crafoord, J
    [J]. INTERNATIONAL JOURNAL OF MEDICAL INFORMATICS, 2003, 69 (01) : 39 - 55
  • [19] A general methodology for three-dimensional analysis of variation in target volume delineation
    Remeijer, P
    Rasch, C
    Lebesque, JV
    van Herk, M
    [J]. MEDICAL PHYSICS, 1999, 26 (06) : 931 - 940
  • [20] Spline-based elastic image registration: integration of landmark errors and orientation attributes
    Rohr, K
    Fornefett, M
    Stiehl, HS
    [J]. COMPUTER VISION AND IMAGE UNDERSTANDING, 2003, 90 (02) : 153 - 168