PPG Sensor Contact Pressure Should Be Taken Into Account for Cuff-Less Blood Pressure Measurement

被引:79
作者
Chandrasekhar, Anand [1 ]
Yavarimanesh, Mohammad [1 ]
Natarajan, Keerthana [1 ]
Hahn, Jin-Oh [2 ]
Mukkamala, Ramakrishna [1 ]
机构
[1] Michigan State Univ, Dept Elect & Comp Engn, E Lansing, MI 48824 USA
[2] Univ Maryland, Dept Mech Engn, College Pk, MD 20742 USA
关键词
Cuff-less blood pressure; mHealth; photo-plethysmography (PPG); PPG sensor contact pressure; pulse arrival time; pulse transit time (PTT); pulse wave velocity; wearable; PHOTOPLETHYSMOGRAPHY;
D O I
10.1109/TBME.2020.2976989
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: Photo-plethysmography (PPG) sensors are often used to detect pulse transit time (PTT) for potential cuff-less blood pressure (BP) measurement. It is known that the contact pressure (CP) of the PPG sensor markedly alters the PPG waveform amplitude. The objective was to test the hypothesis that PTT detected via PPG sensors is likewise impacted by CP. Methods: A device was built to measure the time delay between ECG and finger PPG waveforms (i.e., pulse arrival time (PAT) - a popular surrogate of PTT) and the PPG sensor CP at different CP levels. These measurements and finger cuff BP were recorded while the CP was slowly varied in 17 healthy subjects. Results: Over a physiologic range of CP, the maximum deviations of PAT detected at the PPG foot and peak were 22 +/- 2 and 40 +/- 7 ms (p<0.05), which translate to similar to 11 and similar to 20 mmHg BP error based on the literature. The curve relating PAT detected at the PPG foot to CP was U-shaped with minimum near finger diastolic BP. A conceptual model accounting for finger artery viscoelasticity and nonlinearity explained this curve. Conclusion: Since the regulatory bias error for BP measurement is limited to 5 mmHg, PPG sensor CP should be taken into account for cuff-less BP measurement via PTT. Significance: This study suggests that widely pursued PPG-based BP measurement devices including those that detect PTT should maintain the CP or include a CP measurement in the calibration equation for deriving BP.
引用
收藏
页码:3134 / 3140
页数:7
相关论文
共 28 条
[1]   Photoplethysmography and its application in clinical physiological measurement [J].
Allen, John .
PHYSIOLOGICAL MEASUREMENT, 2007, 28 (03) :R1-R39
[2]  
Association for the Advancement of Medical Instrumentation, 2009, NON SPHYGM 2
[3]   Cuffless Differential Blood Pressure Estimation Using Smart Phones [J].
Chandrasekaran, Vikram ;
Dantu, Ram ;
Jonnada, Srikanth ;
Thiyagaraja, Shanti ;
Subbu, Kalyan Pathapati .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2013, 60 (04) :1080-1089
[4]  
Chandrasekhar A., 2018, Sci. Rep, V8, P18
[5]   Smartphone-based blood pressure monitoring via the oscillometric finger-pressing method [J].
Chandrasekhar, Anand ;
Kim, Chang-Sei ;
Naji, Mohammed ;
Natarajan, Keerthana ;
Hahn, Jin-Oh ;
Mukkamala, Ramakrishna .
SCIENCE TRANSLATIONAL MEDICINE, 2018, 10 (431)
[6]   Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care [J].
Chung, Ha Uk ;
Kim, Bong Hoon ;
Lee, Jong Yoon ;
Lee, Jungyup ;
Xie, Zhaoqian ;
Ibler, Erin M. ;
Lee, KunHyuck ;
Banks, Anthony ;
Jeong, Ji Yoon ;
Kim, Jongwon ;
Ogle, Christopher ;
Grande, Dominic ;
Yu, Yongjoon ;
Jang, Hokyung ;
Assem, Pourya ;
Ryu, Dennis ;
Kwak, Jean Won ;
Namkoong, Myeong ;
Park, Jun Bin ;
Lee, Yechan ;
Kim, Do Hoon ;
Ryu, Arin ;
Jeong, Jaeseok ;
You, Kevin ;
Ji, Bowen ;
Liu, Zhuangjian ;
Huo, Qingze ;
Feng, Xue ;
Deng, Yujun ;
Xu, Yeshou ;
Jang, Kyung-In ;
Kim, Jeonghyun ;
Zhang, Yihui ;
Ghaffari, Roozbeh ;
Rand, Casey M. ;
Schau, Molly ;
Hamvas, Aaron ;
Weese-Mayer, Debra E. ;
Huang, Yonggang ;
Lee, Seung Min ;
Lee, Chi Hwan ;
Shanbhag, Naresh R. ;
Paller, Amy S. ;
Xu, Shuai ;
Rogers, John A. .
SCIENCE, 2019, 363 (6430) :947-+
[7]   THEORY OF THE OSCILLOMETRIC MAXIMUM AND THE SYSTOLIC AND DIASTOLIC DETECTION RATIOS [J].
DRZEWIECKI, G ;
HOOD, R ;
APPLE, H .
ANNALS OF BIOMEDICAL ENGINEERING, 1994, 22 (01) :88-96
[8]   On the Analysis of Fingertip Photoplethysmogram Signals [J].
Elgendi, Mohamed .
CURRENT CARDIOLOGY REVIEWS, 2012, 8 (01) :14-25
[9]  
Flugge W., 1967, VISCOELASTICITY, VSecond, P4
[10]   Effects of different contacting pressure on the transfer function between finger photoplethysmographic and radial blood pressure waveforms [J].
Hsiu, H. ;
Hsu, C-L ;
Wu, T-L .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2011, 225 (H6) :575-583