摘要
A simulation method is described in this work that aids in quantifying the upper limits of lesion detectability as a function of lesion size, lesion contrast, pixel size, and x-ray exposure for digital x-ray imaging systems. The method entails random lesion placement with subsequent simulated imaging on idealized x-ray detectors with no additive noise and 100% quantum detective efficiency. Lesions of different size and thickness were simulated. Mean (expectation) lesion signal-to-noise ratios (LSNRs) were calculated and receiver operating characteristic (ROC) curves were constructed based on LSNR ensembles. Mean (expectation) values of the areas under the ROC curves were calculated for lesions of varying size on pixel arrays of varying size at different exposures. Analyses were performed across several parameters, including lesion size, pixel size, and exposure levels representative of various areas of radiography. As expected, lesion detectability increased with lesion size, contrast, pixel size, and exposure. The model suggests that lesion detectability is strongly dependent on the relative alignment (phase) of the lesion with the pixel matrix for lesions on the order of the pixel size. (C) 2000 American Association of Physicists in Medicine.
引用
收藏
页码:66 / 74
页数:9
相关论文
共 19 条
  • [1] BEVINGTON PR, 1983, DATA REDUCTION ERROR
  • [2] Spectral modeling and compilation of quantum fluence in radiography and mammography
    Boone, JM
    [J]. PHYSICS OF MEDICAL IMAGING, 1998, 3336 : 592 - 601
  • [3] A NOTE ON THE GENERATION OF RANDOM NORMAL DEVIATES
    BOX, GEP
    MULLER, ME
    [J]. ANNALS OF MATHEMATICAL STATISTICS, 1958, 29 (02): : 610 - 611
  • [4] DIGITAL MAMMOGRAPHY - ROC STUDIES OF THE EFFECTS OF PIXEL SIZE AND UNSHARP-MASK FILTERING ON THE DETECTION OF SUBTLE MICROCALCIFICATIONS
    CHAN, HP
    VYBORNY, CJ
    MACMAHON, H
    METZ, CE
    DOI, K
    SICKLES, EA
    [J]. INVESTIGATIVE RADIOLOGY, 1987, 22 (07) : 581 - 589
  • [5] DIGITIZATION REQUIREMENTS IN MAMMOGRAPHY - EFFECTS ON COMPUTER-AIDED DETECTION OF MICROCALCIFICATIONS
    CHAN, HP
    NIKLASON, LT
    IKEDA, DM
    LAM, KL
    ADLER, DD
    [J]. MEDICAL PHYSICS, 1994, 21 (07) : 1203 - 1211
  • [6] Dainty JC., 1974, IMAGE SCI
  • [7] INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .3. EFFECT OF PIXEL SIZE ON SNR AND THRESHOLD CONTRAST
    GIGER, ML
    DOI, K
    [J]. MEDICAL PHYSICS, 1985, 12 (02) : 201 - 208
  • [8] IMAGE QUALITY OF DIGITAL CHEST RADIOGRAPHY - EVALUATION OF SUBTLE INTERSTITIAL LUNG ABNORMALITIES - (PRELIMINARY-STUDY)
    IKEZOE, J
    KIDO, S
    TAKEUCHI, N
    KONDOH, H
    ARISAWA, J
    KOZUKA, T
    [J]. COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 1994, 43 (1-2) : 107 - 114
  • [9] INTERPRETATION OF SUBTLE INTERSTITIAL CHEST ABNORMALITIES - CONVENTIONAL RADIOGRAPHY VERSUS HIGH-RESOLUTION STORAGE-PHOSPHOR RADIOGRAPHY - A PRELIMINARY-STUDY
    IKEZOE, J
    KOHNO, N
    KIDO, S
    TAKEUCHI, N
    JOHKOH, T
    ARISAWA, J
    KOZUKA, T
    [J]. JOURNAL OF DIGITAL IMAGING, 1995, 8 (01) : 31 - 36
  • [10] JOHNS HE, 1983, PHYSICS RADIOLOGY