Rigid registration of freehand 3D ultrasound and CT-scan kidney images

被引:0
作者
Leroy, A [1 ]
Mozer, P
Payan, Y
Troccaz, J
机构
[1] Fac Med, Lab TIMC, GMCAO, F-38700 La Tronche, France
[2] CHU Pitie Salpetriere, AP HP, Serv Urol & Transplantat Renale, F-75013 Paris, France
来源
MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2004, PT 1, PROCEEDINGS | 2004年 / 3216卷
关键词
D O I
暂无
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
This paper presents a method to register a preoperative CT volume to a sparse set of intraoperative US slices. In the context of percutaneous renal puncture, the aim is to transfer a planning information to an intraoperative coordinate system. The spatial position of the US slices is measured by localizing a calibrated probe. Our method consists in optimizing a rigid 6 degree of freedom (DOF) transform by evaluating at each step the similarity between the set of US images and the CT volume. The images have been preprocessed in order to increase the relationship between CT and US pixels. Correlation Ratio turned out to be the most accurate and appropriate similarity measure to be used in a Powell-Brent minimization scheme. Results are compared to a standard rigid point-to-point registration involving segmentation, and discussed.
引用
收藏
页码:837 / 844
页数:8
相关论文
共 12 条
[1]   Detection of lines and boundaries in speckle images - Application to medical ultrasound [J].
Czerwinski, RN ;
Jones, DL ;
O'Brien, WD .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (02) :126-136
[2]   A global optimisation method for robust affine registration of brain images [J].
Jenkinson, M ;
Smith, S .
MEDICAL IMAGE ANALYSIS, 2001, 5 (02) :143-156
[3]  
LEROY A, 2002, SURGETICA 02, P303
[4]   Multimodality image registration by maximization of mutual information [J].
Maes, F ;
Collignon, A ;
Vandermeulen, D ;
Marchal, G ;
Suetens, P .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) :187-198
[5]  
Maes F, 1999, Med Image Anal, V3, P373, DOI 10.1016/S1361-8415(99)80030-9
[6]   Registration of freehand 3D ultrasound and magnetic resonance liver images [J].
Penney, GP ;
Blackall, JM ;
Hamady, MS ;
Sabharwal, T ;
Adam, A ;
Hawkes, DJ .
MEDICAL IMAGE ANALYSIS, 2004, 8 (01) :81-91
[7]  
Roche A, 1998, LECT NOTES COMPUT SC, V1496, P1115, DOI 10.1007/BFb0056301
[8]   Rigid registration of 3-D ultrasound with MR images: A new approach combining intensity and gradient information [J].
Roche, A ;
Pennec, X ;
Malandain, G ;
Ayache, N .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2001, 20 (10) :1038-1049
[9]  
SARRUT D, 1999, EUR WORKSH CONT BAS, P263
[10]  
Studholme C, 1996, Med Image Anal, V1, P163, DOI 10.1016/S1361-8415(96)80011-9