Performance of electronic portal imaging devices (EPIDs) used in radiotherapy:: Image quality and dose measurements

被引:47
作者
Cremers, F [1 ]
Frenzel, T [1 ]
Kausch, C [1 ]
Albers, D [1 ]
Schönborn, T [1 ]
Schmidt, R [1 ]
机构
[1] Univ Hamburg Hosp, Clin Radiotherapy & Radiooncol, D-20246 Hamburg, Germany
关键词
portal imaging; image quality; amorphous silicon array; fluoroscopic portal imaging system; response; dosimetry; radiation therapy;
D O I
10.1118/1.1688212
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of our Study was to compare the image and dosimetric quality of two different imaging systems. The first one is a fluoroscopic electronic portal imaging device (first generation), while the second is based on an amorphous silicon flat-panel array (second generation). The parameters describing image quality include spatial resolution [modulation transfer function (MTF)], noise [noise power spectrum (NPS)], and signal-to-noise transfer [detective quantum efficiency (DQE)]. The dosimetric measurements were compared with ionization chamber as well as with film measurements. The response of the flat-panel imager and the fluoroscopic-optical device was determined performing a two-step Monte Carlo simulation. All measurements were performed in a 6 MV linear accelerator photon beam. The resolution (MTF) of the fluoroscopic device (f(1/2) = 0.3 mm(-1)) is larger than of the amorphous silicon based system (f(1/2) = 0.21 mm(-1)), which is due to the missing backscattered photons and the smaller pixel size. The noise measurements (NPS) show the correlation of neighboring pixels of the amorphous silicon electronic portal imaging device, whereas the NPS of the fluoroscopic system is frequency independent. At zero spatial frequency the DQE of the flat-panel imager has a value of 0.008 (0.8%). Due to the minor frequency dependency this device may be almost x-ray quantum limited. Monte Carlo simulations verified these characteristics. For the fluoroscopic imaging system the DQE at low frequencies is about 0.0008 (0.08%) and degrades with higher frequencies. Dose measurements with the flat-panel imager revealed that images can only be directly converted to portal dose images, if scatter can be neglected. Thus objects distant to the detector (e.g., inhomogeneous dose distribution generated by a modificator) can be verified dosimetrically, while objects close to a detector (e.g., a patient) cannot be verified directly and must be scatter corrected prior to verification. This is justified by the response of the flat-panel imaging device revealing a strong dependency at low energies. (C) 2004 American Association of PhYsicists in Medicine.
引用
收藏
页码:985 / 996
页数:12
相关论文
共 50 条
  • [21] A quantitative image quality comparison of four different image guided radiotherapy devices
    Stuetzel, Julia
    Oelfke, Uwe
    Nill, Simeon
    RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) : 20 - 24
  • [22] A comprehensive evaluation of angular range and separation on image quality, image registration, and imaging dose for cone beam computed tomography in radiotherapy
    Men, Kuo
    Dai, Jianrong
    MEDICAL DOSIMETRY, 2019, 44 (01) : 67 - 73
  • [23] Electroluminescence Imaging of PV devices: Determining the Image Quality
    Bedrich, K. G.
    Bliss, M.
    Betts, T. R.
    Gottschalg, R.
    2015 IEEE 42ND PHOTOVOLTAIC SPECIALIST CONFERENCE (PVSC), 2015,
  • [24] Initial Clinical Experience with Electronic Portal Imaging Device-Based In Vivo Dosimetry for Radiotherapy in Gynecological Cancers
    Yedekci, Yagiz
    Yuce Sari, Sezin
    Gultekin, Melis
    Yildiz, Ferah
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2022, 37 (04): : 446 - 452
  • [25] The dose response relationship of a liquid-filled electronic portal imaging device
    Boellaard, R
    vanHerk, M
    Mijnheer, BJ
    MEDICAL PHYSICS, 1996, 23 (09) : 1601 - 1611
  • [26] Quality control of dose delivered by in vivo measurements for head and neck radiotherapy
    Farhat, L.
    Besbes, M.
    Bridier, A.
    Kaffel, F.
    Daoud, J.
    CANCER RADIOTHERAPIE, 2010, 14 (01): : 69 - 73
  • [27] A novel water-equivalent electronic portal imaging device for radiotherapy with improved detective quantum efficiency: Proof of concept
    Blake, Samuel J.
    Cheng, Zhangkai
    Atakaramians, Shaghik
    Meikle, Steven
    Lu, Minghui
    Vial, Philip
    Kuncic, Zdenka
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [28] Can CT scan protocols used for radiotherapy treatment planning be adjusted to optimize image quality and patient dose? A systematic review
    Davis, Anne T.
    Palmer, Antony L.
    Nisbet, Andrew
    BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1076)
  • [29] Clinical practice and evaluation of electronic portal imaging device for VMAT quality assurance
    Huang, Yen-Cho
    Yeh, Chien-Yi
    Yeh, Jih-Hsiang
    Lo, Ching-Jung
    Tsai, Ping-Fang
    Hung, Chih-Hung
    Tsai, Chieh-Sheng
    Chen, Chen-Yuan
    MEDICAL DOSIMETRY, 2013, 38 (01) : 35 - 41
  • [30] Clinical Feasibility of Using an Electronic Portal Imaging Device for Position Verification during Conventional Radiotherapy
    Shen, Yuxiaotong
    Zhang, Jie
    Ge, Yun
    Chen, Ying
    Li, Haiwei
    Sun, Wei
    Ji, Mingxi
    Wei, Quanbo
    Cai, Jing
    Li, Bing
    2016 IEEE 10TH INTERNATIONAL CONFERENCE ON NANO/MOLECULAR MEDICINE AND ENGINEERING (NANOMED), 2016, : 99 - 103