Respiratory pattern and rebreathing in the Mapleson A, C and D breathing systems with spontaneous ventilation - A theory

被引:4
作者
Cook, LB
机构
[1] Department of Anaesthesia, Royal Oldham Hospital, Oldham, Lanes, Rochdale Road
关键词
equipment; breathing systems; Mapleson A; C; D; ventilation; spontaneous;
D O I
10.1111/j.1365-2044.1996.tb07752.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A theoretical analysis is presented of the Mapleson A, C and D breathing systems when used in spontaneous respiration. The influence of the respiratory pattern is explained diagrammatically. Simple equations are derived, predicting the fresh gas flow required to prevent rebreathing with different respiratory patterns. Further equations allow the degree of rebreathing caused by inadequate fresh gas flow to be quantified. These are used to examine the effects of different respiratory patterns on the efficiency of the three systems. It is demonstrated that the single most important determinant of efficiency is the duration of the expiratory pause. The nature of the inspiratory and expiratory waveforms is less important and the I:E ratio far less important. The analysis suggests that the Mapleson A system will always be the most efficient of the three systems. The Mapleson C system will be efficient if inspiration is long and the expiratory pause is minimal. The Mapleson D system will be efficient if the expiratory pause is sufficiently long.
引用
收藏
页码:371 / 385
页数:15
相关论文
共 17 条
[1]   CLINICAL COMPARISON OF THE BAIN AND MAGILL ANESTHETIC SYSTEMS DURING SPONTANEOUS RESPIRATION [J].
ALEXANDER, JP .
BRITISH JOURNAL OF ANAESTHESIA, 1982, 54 (10) :1031-1036
[2]  
ANDERSEN PK, 1989, ACTA ANAESTH SCAND, V33, P4339
[3]  
BRAIN JA, 1972, CAN ANAESTH SOC J, V19, P426
[4]   A COMPARISON OF ANESTHETIC BREATHING SYSTEMS DURING SPONTANEOUS VENTILATION - AN INVITRO STUDY USING A LUNG MODEL [J].
CHAN, ASH ;
BRUCE, WE ;
SONI, N .
ANAESTHESIA, 1989, 44 (03) :194-199
[5]   SPONTANEOUS VENTILATION WITH BAIN ANESTHETIC SYSTEM [J].
CONWAY, CM ;
SEELEY, HF ;
BARNES, PK .
BRITISH JOURNAL OF ANAESTHESIA, 1977, 49 (12) :1245-1249
[6]   MINIMUM FRESH GAS-FLOW REQUIREMENTS OF ANESTHETIC BREATHING SYSTEMS DURING SPONTANEOUS VENTILATION - A GRAPHICAL APPROACH [J].
DORRINGTON, KL ;
LEHANE, JR .
ANAESTHESIA, 1987, 42 (07) :732-737
[7]   EFFECT OF INSPIRATION-EXPIRATION RATIO ON REBREATHING WITH THE MAPLESON-D SYSTEM (BAIN MODIFICATION - COAXIAL SYSTEM) [J].
GABRIELSEN, J ;
VANDENBERG, JT ;
DIRKSEN, H ;
RUBEN, H .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1980, 24 (04) :336-338
[8]   THE EFFECT OF THE BAIN CIRCUIT ON GAS-EXCHANGE [J].
JAEGER, MJ ;
SCHULTETUS, RR .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1987, 34 (01) :26-34
[9]   INFLUENCE OF THE RESPIRATORY FLOW PATTERN ON REBREATHING IN MAPLESON-A AND MAPLESON-D CIRCUITS [J].
JONSSON, LO ;
ZETTERSTROM, H .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1987, 31 (02) :174-178
[10]   FRESH GAS-FLOW IN COAXIAL MAPLESON-A AND MAPLESON-D CIRCUITS DURING SPONTANEOUS BREATHING [J].
JONSSON, LO ;
ZETTERSTROM, H .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1986, 30 (07) :588-593