Fiber-Based Clock Synchronization Method for Medical Ultrasound System

被引:0
作者
Zhou, Juan [1 ,2 ]
Li, Jincheng [1 ,2 ]
Zhong, Haibo [1 ,2 ]
Shi, Xudong [1 ,2 ]
Yang, Ge [1 ,2 ]
Huang, Jiqing [1 ,2 ]
Li, Yongchuan [1 ,2 ]
Ma, Teng [1 ,2 ]
Long, Xiaojing [1 ,2 ]
Qiu, Weibao [1 ,2 ]
Zheng, Hairong [1 ,2 ]
机构
[1] Chinese Acad Sci, Paul C Lauterbur Res Ctr Biomed Imaging, Shenzhen Inst Adv Technol, Shenzhen 518055, Peoples R China
[2] Shenzhen Key Lab Ultrasound Imaging & Therapy, Shenzhen 518055, Peoples R China
基金
美国国家科学基金会;
关键词
Clocks; Ultrasonic imaging; Synchronization; Optical fibers; Field programmable gate arrays; Optical transmitters; Magnetic resonance imaging; Brain ultrasound; fiber-based clock synchronization; large-scale array transducer; magnetic resonance imaging (MRI) compatible; ultrasound therapy; BRAIN; TRANSDUCER; DISEASES;
D O I
10.1109/TUFFC.2020.2994097
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Brain ultrasound has attracted great attention recently due to its noninvasive treatment function for brain diseases. However, ultrasound is still difficult to pass through an intact skull. Phase correction is recognized as an effective method for skull compensation. Half-wavelength pitch transducer is important for the phase correction and, hence, thousands of elements array is required to cover large area human tissue. The clock synchronization between elements is crucial for the phase correction; however, the traditional clock scheme which is designed for 128- or 256-element system is not suitable for thousands of elements. In addition, the clock scheme needs to be magnetic resonance imaging (MRI) compatible since MRI-guided intervention is becoming a routine operation for the brain ultrasound. This study is the first to propose an optical fiber-based clock synchronization method for MRI-guided ultrasound array system. The optical fiber not only distributes the clock but also sets up a link to transmit the data for ultrasound beamformer. The link is full-duplex so both the clock and the data can be transmitted and received simultaneously. The precision of clock synchronization is less than 557 ps when using 50 MHz clock, and the period jitter of the clock is less than 10 ps (rms). Multiple 128- or 256-channel ultrasonic systems can be synchronized, and the error between the channels can be less than 10 ns when using 1-MHz ultrasound transducer. The system can work in an MRI scanning room and communicate with a console via only one fiber. In vivo primate animal study has been achieved, and it has been proven that the proposed clock scheme is suitable for MRI-guided large-scale ultrasound array system.
引用
收藏
页码:136 / 142
页数:7
相关论文
共 29 条
  • [1] First-in-human trial of blood-brain barrier opening in amyotrophic lateral sclerosis using MR-guided focused ultrasound
    Abrahao, Agessandro
    Meng, Ying
    Llinas, Maheleth
    Huang, Yuexi
    Hamani, Clement
    Mainprize, Todd
    Aubert, Isabelle
    Heyn, Chinthaka
    Black, Sandra E.
    Hynynen, Kullervo
    Lipsman, Nir
    Zinman, Lorne
    [J]. NATURE COMMUNICATIONS, 2019, 10 (1)
  • [2] Timing and Fast Control for the Upgraded Readout Architecture of the LHCb Experiment at CERN
    Alessio, Federico
    Jacobsson, Richard
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (05) : 3438 - 3445
  • [3] Experimental demonstration of noninvasive transskull adaptive focusing based on prior computed tomography scans
    Aubry, JF
    Tanter, M
    Pernot, M
    Thomas, JL
    Fink, M
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2003, 113 (01) : 84 - 93
  • [4] Bellini R, 2019, 46TH CONGRESSO NAZIONALE SIMFER, P8
  • [5] A Front-End ASIC With Receive Sub-array Beamforming Integrated With a 32 x 32 PZT Matrix Transducer for 3-D Transesophageal Echocardiography
    Chen, Chao
    Chen, Zhao
    Bera, Deep
    Raghunathan, Shreyas B.
    Shabanimotlagh, Maysam
    Noothout, Emile
    Chang, Zu-Yao
    Ponte, Jacco
    Prins, Christian
    Vos, Hendrik J.
    Bosch, Johan G.
    Verweij, Martin D.
    de Jong, Nico
    Pertijs, Michiel A. P.
    [J]. IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2017, 52 (04) : 994 - 1006
  • [6] A non-invasive method for focusing ultrasound through the human skull
    Clement, GT
    Hynynen, K
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (08) : 1219 - 1236
  • [7] Enhanced neuronal activity in mouse motor cortex with microbubbles' oscillations by transcranial focused ultrasound stimulation
    Cui, Zhiwei
    Li, Dapeng
    Feng, Yang
    Xu, Tianqi
    Wu, Shan
    Li, Yibao
    Bouakaz, Ayache
    Wan, Mingxi
    Zhang, Siyuan
    [J]. ULTRASONICS SONOCHEMISTRY, 2019, 59
  • [8] The Cost of Brain Diseases: A Burden or a Challenge?
    DiLuca, Monica
    Olesen, Jes
    [J]. NEURON, 2014, 82 (06) : 1205 - 1208
  • [9] A Pilot Study of Focused Ultrasound Thalamotomy for Essential Tremor
    Elias, W. Jeffrey
    Huss, Diane
    Voss, Tiffini
    Loomba, Johanna
    Khaled, Mohamad
    Zadicario, Eyal
    Frysinger, Robert C.
    Sperling, Scott A.
    Wylie, Scott
    Monteith, Stephen J.
    Druzgal, Jason
    Shah, Binit B.
    Harrison, Madaline
    Wintermark, Max
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2013, 369 (07) : 640 - 648
  • [10] A novel, flat, electronically-steered phased array transducer for tissue ablation: preliminary results
    Ellens, Nicholas P. K.
    Lucht, Benjamin B. C.
    Gunaseelan, Samuel T.
    Hudson, John M.
    Hynynen, Kullervo H.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (06) : 2195 - 2215