A flexible and high throughput liquid chromatography-tandem mass spectrometric assay for the quantitation of telcagepant in human plasma

被引:1
|
作者
Du, Lihong [1 ]
Miller-Stein, Cynthia M. [1 ]
Valesky, Robert J. [1 ]
Martucci, Ashley N. [1 ]
Woolf, Eric J. [1 ]
机构
[1] Merck & Co Inc, Dept Clin PK PD, Merck Res Labs, West Point, PA 19486 USA
关键词
Telcagepant (MK-0974); Protein precipitation; Human plasma; LC-MS/MS; RECEPTOR ANTAGONIST; MIGRAINE; CGRP; MK-0974; ELECTROSPRAY; STRATEGIES; SAMPLES;
D O I
10.1016/j.jchromb.2009.03.019
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Telcagepant (MK-0974) is a novel oral calcitonin gene-related peptide (CGRP) receptor antagonist and is currently under clinical development. Results from phases II and III clinical trials have Suggested that telcagepant is effective for migraine treatment. A reliable and high throughput protein precipitation (PPT) method for determination of telcagepant in human plasma using liquid chromatography coupled with atmospheric pressure chemical ionization (APCI) tandem mass spectrometry has been developed. Clinical samples, internal standard (IS) and acetonitrile are transferred into 96-well plates using a robotic liquid handling system. An aliquot of 10 mu L supernatant is directly injected into the LC-MS/MS system where separation is performed on a FluoPhase RP (150 x 2.1 mm, 5 mu m) column with an isocratic mobile phase (60% acetonitrile with 0.1% formic acid and 40% water with 0.1% formic acid)at 0.2 mL/min. The interfering 3S-diastereomer of telcagepant, which is observed in clinical samples, is chromatographically resolved from telcagepant. The PPT procedure significantly reduces the time required for sample processing and the assay is sufficiently sensitive for detection using both API 4000 and API 3000 mass spectrometers. The linear calibration range is 5-5000 nM using 200 mu L of plasma. Assay intraday validation was conducted using Six calibration Curves derived from six lots of human control plasma. Calibration standard accuracy did not deviate by more than 3% and 6% of nominal values, and precision did not exceed 4% coefficient of variation (CV) and 10% CV, respectively on the Apt 4000 and API 3000. Several clinical phases IIb and III Studies have been successfully supported with this assay. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:1465 / 1471
页数:7
相关论文
共 50 条
  • [2] Liquid Chromatography-Tandem Mass Spectrometric Assay for Determination of Stavudine in Human Plasma
    Jin, Fengdan
    JOURNAL OF SPECTROSCOPY, 2014, 2014
  • [3] Determination of metformin in human plasma by liquid chromatography-tandem mass spectrometric assay
    Wang, J
    Wang, YW
    Gu, JK
    Wu, Y
    Wang, Y
    CHEMICAL RESEARCH IN CHINESE UNIVERSITIES, 2005, 21 (02) : 246 - 250
  • [4] Liquid chromatography-tandem mass spectrometric assay for TPN171 in human plasma
    Zhang, Mengqi
    Liu, Haiming
    He, Jiajin
    Gui, Yuzhou
    Qian, Hongjie
    Jia, Jingying
    Yu, Chen
    Wang, Zhen
    Shen, Jing Shan
    Liu, Gangyi
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2020, 191
  • [5] Rapid and sensitive liquid chromatography-tandem mass spectrometric method for the quantitation of metformin in human plasma
    Wang, YW
    Tang, YB
    Gu, JK
    Fawcett, JP
    Bai, X
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 808 (02): : 215 - 219
  • [6] Validation of a Liquid Chromatography-Tandem Mass Spectrometric Assay for Quantitative Analysis of Lenvatinib in Human Plasma
    Ogawa-Morita, Tomoko
    Sano, Yoshiyuki
    Okano, Tomoka
    Fujii, Hirofumi
    Tahara, Makoto
    Yamaguchi, Masakazu
    Minami, Hironobu
    INTERNATIONAL JOURNAL OF ANALYTICAL CHEMISTRY, 2017, 2017
  • [7] Liquid chromatography-tandem mass spectrometric assay for the multikinase inhibitor regorafenib in plasma
    Luethi, Dino
    Durmus, Selvi
    Schinkel, Alfred H.
    Schellens, Jan H. M.
    Beijnen, Jos H.
    Sparidans, Rolf W.
    BIOMEDICAL CHROMATOGRAPHY, 2014, 28 (10) : 1366 - 1370
  • [8] Liquid chromatography-tandem mass spectrometric assay for the PARP inhibitor rucaparib in plasma
    Sparidans, Rolf W.
    Durmus, Selvi
    Schinkel, Alfred H.
    Schellens, Jan H. M.
    Beijnen, Jos H.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2014, 88 : 626 - 629
  • [9] Liquid chromatography-tandem mass spectrometry assay for the quantitation of β-dihydroartemisinin in rat plasma
    Xing, Jie
    Yan, Hong-Xia
    Wang, Rui-Li
    Zhang, Li-Feng
    Zhang, Shu-Qiu
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 852 (1-2): : 202 - 207
  • [10] Liquid chromatography-tandem mass spectrometric analysis of stobadine in human plasma
    Abdel-Hamid, ME
    Novotny, L
    Stankovicova, M
    Benes, L
    JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2005, 28 (03) : 383 - 393